Impact of defibrillator-can size on defibrillation success with a single-lead unipolar system.

In a study of 11 dogs, we assessed whether the defibrillation energy requirements of a single transvenous right ventricular electrode/defibrillator can system depended on the can size. We compared the defibrillation threshold obtained with 65% fixed-tilt biphasic shocks with 20, 40, and 80 ml surface area defibrillator cans. The energy was delivered between a right ventricular coil inserted through the jugular vein and the can placed in the subcutaneous tissue of the left superior chest wall. The testing order of each can size was randomly determined. Triplicate defibrillation thresholds were obtained with each can. Despite a higher impedance (20 ml 85 +/- 22 ohms vs 80 ml 71 +/- 16 ohms, p < 0.01), the 20 ml can resulted in a similar defibrillation threshold compared with the 80 ml (20 ml 7.6 +/- 2.8 J vs 80 ml 7.5 +/- 3.4 J) and the 40 ml cans (20 ml 7.6 +/- 2.8 J vs 7.5 +/- 3.4 J). In conclusion, with the unipolar lead system the can size does not appear to be a factor limiting defibrillation success. Even a can the size of a pacemaker does not appear to significantly affect the defibrillation efficacy of this lead system.

[1]  D P Zipes,et al.  Termination of ventricular fibrillation in dogs by depolarizing a critical amount of myocardium. , 1975, The American journal of cardiology.

[2]  H. Greene,et al.  Evaluation of electrode polarity on defibrillation efficacy. , 1989, The American journal of cardiology.

[3]  P D Wolf,et al.  Three‐dimensional Potential Gradient Fields Generated by Intracardiac Catheter and Cutaneous Patch Electrodes , 1992, Circulation.

[4]  George A. Johnson,et al.  A Simplified, Single‐Lead Unipolar Transvenous Cardioversion‐Defibrillation System , 1993, Circulation.

[5]  George A. Johnson,et al.  Implantable transvenous cardioverter-defibrillators. , 1993, Circulation.

[6]  George A. Johnson,et al.  An effective and adaptable transvenous defibrillation system using the coronary sinus in humans. , 1990, Journal of the American College of Cardiology.

[7]  R. Fletcher,et al.  Optimization of Biphasic Waveforms for Human Nonthoracotomy Defibrillation , 1993, Circulation.

[8]  J. Schuder,et al.  Ventricular defibrillation in the dog with a bielectrode intravascular catheter. , 1973, Archives of internal medicine.

[9]  P D Wolf,et al.  Improved defibrillation thresholds with large contoured epicardial electrodes and biphasic waveforms. , 1987, Circulation.

[10]  P D Wolf,et al.  Cardiac Potential and Potential Gradient Fields Generated by Single, Combined, and Sequential Shocks During Ventricular Defibrillation , 1992, Circulation.

[11]  B. Lüderitz,et al.  Clinical efficacy of shock waveforms and lead configurations for defibrillation. , 1994, American heart journal.

[12]  R E Ideker,et al.  Ventricular defibrillation using biphasic waveforms: the importance of phasic duration. , 1989, Journal of the American College of Cardiology.

[13]  P. D. Chapman,et al.  Nonthoracotomy internal defibrillation in dogs: threshold reduction using a subcutaneous chest wall electrode with a transvenous catheter electrode. , 1987, Journal of the American College of Cardiology.

[14]  G. Guiraudon,et al.  A permanent transvenous lead system for an implantable pacemaker cardioverter-defibrillator. Nonthoracotomy approach to implantation. , 1992, Circulation.

[15]  P D Wolf,et al.  The potential gradient field created by epicardial defibrillation electrodes in dogs. , 1986, Circulation.

[16]  George A. Johnson,et al.  Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode. , 1994, Circulation.

[17]  B. Wilkoff,et al.  Automatic implantable cardioverter-defibrillator implantation without thoracotomy using an endocardial and submuscular patch system. , 1991, Journal of the American College of Cardiology.

[18]  P. Wolf,et al.  Current Concepts for Selecting the Location, Size and Shape of Defibrillation Electrodes , 1991, Pacing and clinical electrophysiology : PACE.

[19]  P. Kudenchuk,et al.  A Prospective Randomized Comparison in Humans of Biphasic Waveform 60-μF and 120-μF Capacitance Pulses Using A Unipolar Defibrillation System , 1995 .

[20]  George A. Johnson,et al.  Transvenous defibrillation in humans via the coronary sinus. , 1990, Circulation.