The Impact of Positron Range on PET Resolution, Evaluated with Phantoms and PHITS Monte Carlo Simulations for Conventional and Non-conventional Radionuclides

Purpose The increasing interest and availability of non-standard positron-emitting radionuclides has heightened the relevance of radionuclide choice in the development and optimization of new positron emission tomography (PET) imaging procedures, both in preclinical research and clinical practice. Differences in achievable resolution arising from positron range can largely influence application suitability of each radionuclide, especially in small-ring preclinical PET where system blurring factors due to annihilation photon acollinearity and detector geometry are less significant. Some resolution degradation can be mitigated with appropriate range corrections implemented during image reconstruction, the quality of which is contingent on an accurate characterization of positron range. Procedures To address this need, we have characterized the positron range of several standard and non-standard PET radionuclides (As-72, F-18, Ga-68, Mn-52, Y-86, and Zr-89) through imaging of small-animal quality control phantoms on a benchmark preclinical PET scanner. Further, the Particle and Heavy Ion Transport code System (PHITS v3.02) code was utilized for Monte Carlo modeling of positron range-dependent blurring effects. Results Positron range kernels for each radionuclide were derived from simulation of point sources in ICRP reference tissues. PET resolution and quantitative accuracy afforded by various radionuclides in practicable imaging scenarios were characterized using a convolution-based method based on positron annihilation distributions obtained from PHITS. Our imaging and simulation results demonstrate the degradation of small animal PET resolution, and quantitative accuracy correlates with increasing positron energy; however, for a specific “benchmark” preclinical PET scanner and reconstruction workflow, these differences were observed to be minimal given radionuclides with average positron energies below ~ 400 keV. Conclusion Our measurements and simulations of the influence of positron range on PET resolution compare well with previous efforts documented in the literature and provide new data for several radionuclides in increasing clinical and preclinical use. The results will support current and future improvements in methods for positron range corrections in PET imaging.

[1]  Y. Iwamoto,et al.  Benchmark study of the recent version of the PHITS code , 2017 .

[2]  A. Endo,et al.  Compilation of New Nuclear Decay Data Files Used for Dose Calculation , 2001 .

[3]  S. E. Derenzo,et al.  Precision measurement of annihilation point spread distributions for medically important positron emitters , 1979 .

[4]  E. Hoffman,et al.  Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution. , 1999, Physics in medicine and biology.

[5]  S. Larson,et al.  89Zr-DFO-J591 for ImmunoPET of Prostate-Specific Membrane Antigen Expression In Vivo , 2010, The Journal of Nuclear Medicine.

[6]  Pedro Andreo,et al.  Positron flight in human tissues and its influence on PET image spatial resolution , 2003, European Journal of Nuclear Medicine and Molecular Imaging.

[7]  Weibo Cai,et al.  Novel Preparation Methods of (52)Mn for ImmunoPET Imaging. , 2015, Bioconjugate chemistry.

[8]  C. Levin,et al.  Study of PET intrinsic spatial resolution and contrast recovery improvement for PET/MRI systems , 2012, Physics in medicine and biology.

[9]  Overview of particle and heavy ion transport code system PHITS , 2015 .

[10]  Z H Cho,et al.  Positron ranges obtained from biomedically important positron-emitting radionuclides. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[11]  K. Eliceiri,et al.  Development of a novel linearly-filled Derenzo microPET phantom. , 2016, American journal of nuclear medicine and molecular imaging.

[12]  R. Leahy,et al.  Digimouse: a 3D whole body mouse atlas from CT and cryosection data , 2007, Physics in medicine and biology.

[13]  J. Disselhorst,et al.  Spatial resolution of the Inveon small-animal PET scanner for the entire field of view , 2010 .

[14]  S. Jurisson,et al.  Trithiols and their arsenic compounds for potential use in diagnostic and therapeutic radiopharmaceuticals. , 2016, Nuclear medicine and biology.

[15]  C. Champion,et al.  Positron follow-up in liquid water: I. A new Monte Carlo track-structure code , 2006, Physics in medicine and biology.

[16]  M E Phelps,et al.  Effect of positron range on spatial resolution. , 1975, Journal of nuclear medicine : official publication, Society of Nuclear Medicine.

[17]  Chan Hyeong Kim,et al.  Implementation of tetrahedral-mesh geometry in Monte Carlo radiation transport code PHITS , 2017, Physics in medicine and biology.

[18]  A. Martínez-Dávalos,et al.  Positron range in tissue-equivalent materials: experimental microPET studies , 2016, Physics in medicine and biology.

[19]  Takuya Furuta,et al.  Features of Particle and Heavy Ion Transport code System (PHITS) version 3.02 , 2018 .

[20]  Hang Si,et al.  TetGen, a Delaunay-Based Quality Tetrahedral Mesh Generator , 2015, ACM Trans. Math. Softw..

[21]  Jason S. Lewis,et al.  Pretargeting of internalizing trastuzumab and cetuximab with a 18F-tetrazine tracer in xenograft models , 2017, EJNMMI Research.

[22]  J J Vaquero,et al.  Positron range estimations with PeneloPET , 2013, Physics in medicine and biology.

[23]  B. Lewis,et al.  Biodistribution and PET Imaging of pharmacokinetics of manganese in mice using Manganese-52 , 2017, PloS one.

[24]  R Laforest,et al.  Image quality with non-standard nuclides in PET. , 2008, The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the Society of....

[25]  Martin Gotthardt,et al.  Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner , 2008, Journal of Nuclear Medicine.

[26]  A. Sanchez-Crespo,et al.  Comparison of Gallium-68 and Fluorine-18 imaging characteristics in positron emission tomography. , 2013, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.

[27]  C. Bodet-Milin,et al.  A pretargeting system for tumor PET imaging and radioimmunotherapy , 2015, Front. Pharmacol..

[28]  R. Schibli,et al.  Imaging quality of (44)Sc in comparison with five other PET radionuclides using Derenzo phantoms and preclinical PET. , 2016, Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine.