Evaluation of optimal DNA staining for triggering by scanning fluorescence microscopy (SFM)

In imaging and flow cytometry, DNA staining is a common trigger signal for cell identification. Selection of the proper DNA dye is restricted by the hardware configuration of the instrument. The Zeiss Imaging Solutions GmbH (München, Germany) introduced a new automated scanning fluorescence microscope - SFM (Axio Imager.Z1) which combines fluorescence imaging with cytometric parameters measurement. The aim of the study was to select optimal DNA dyes as trigger signal in leukocyte detection and subsequent cytometric analysis of double-labeled leukocytes by SFM. Seven DNA dyes (DAPI, Hoechst 33258, Hoechst 33342, POPO-3, PI, 7-AAD, and TOPRO-3) were tested and found to be suitable for the implemented filtersets (fs) of the SFM (fs: 49, fs: 44, fs: 20). EDTA blood was stained after erythrocyte lysis with DNA dye. Cells were transferred on microscopic slides and embedded in fluorescent mounting medium. Quality of DNA fluorescence signal as well as spillover signals were analyzed by SFM. CD45-APC and CD3-PE as well as CD4-FITC and CD8-APC were selected for immunophenotyping and used in combination with Hoechst. Within the tested DNA dyes DAPI showed relatively low spillover and the best CV value. Due to the low spillover of UV DNA dyes a triple staining of Hoechst and APC and PE (or APC and FITC, respectively) could be analyzed without difficulty. These results were confirmed by FCM measurements. DNA fluorescence is applicable for identifying and triggering leukocytes in SFM analyses. Although some DNA dyes exhibit strong spillover in other fluorescence channels, it was possible to immunophenotype leukocytes. DAPI seems to be best suitable for use in the SFM system and will be used in protocol setups as primary parameter.

[1]  I. Johnson,et al.  The molecular probes handbook : a guide to fluorescent probes and labeling technologies , 2010 .

[2]  P. Chattopadhyay,et al.  Seventeen-colour flow cytometry: unravelling the immune system , 2004, Nature Reviews Immunology.

[3]  R. Haugland The Handbook: A Guide to Fluorescent Probes and Labeling Technologies , 2005 .

[4]  Attila Tárnok,et al.  Polychromatic (eight‐color) slide‐based cytometry for the phenotyping of leukocyte, NK, and NKT subsets , 2005, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[5]  Zsolt Tulassay,et al.  Scanning fluorescent microscopy is an alternative for quantitative fluorescent cell analysis , 2004, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[6]  Attila Tárnok,et al.  Scanning fluorescent microscopy analysis is applicable for absolute and relative cell frequency determinations , 2004, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[7]  Attila Tárnok,et al.  CYTOMICS: AN ENTRY TO BIOMEDICAL CELL SYSTEMS BIOLOGY Cytomics Goes 3D: Toward Tissomics , 2005 .

[8]  Attila Tárnok,et al.  Hyperchromatic cytometry principles for cytomics using slide based cytometry , 2006, Cytometry. Part A : the journal of the International Society for Analytical Cytology.

[9]  Attila Tarnok,et al.  Cytomics goes 3D: Toward tissomics , 2005, Cytometry. Part A : the journal of the International Society for Analytical Cytology.