HEMOLYSIS ANALYSIS OF AXIAL BLOOD PUMPS WITH VARIOUS STRUCTURE IMPELLERS

Low hemolysis is an important factor for axial blood pumps that has been used in patients with heart failure. The structure of impellers plays a key role in the hemolytic properties of axial blood pumps. Axial blood pumps with various structure impellers exhibit different hemolytic characteristic. In the present study, we aimed to investigate the type of impellers structures in axial blood pumps that contain the best low hemolytic properties. Also, it is expensive and time-consuming to validate the axial blood pump's hemolytic property by in vivo experiments. Therefore, in the present study, the numerical method was applied to analyze the hemolytic property in a blood pump. Specifically, the hemolysis of the pump was calculated by using a forward Euler approach based on the changes in shear stress and related exposure times along the particle trace lines. The different vane structures and rotational speed that affect hemolysis were analyzed and compared. The results showed that long–short alternant vanes exhibited the best hemolytic property which could be utilized in the optimization design of axial blood pumps.

[1]  Xinwei Song,et al.  TRANSIENT COMPUTATIONAL FLUID DYNAMICS STUDY OF AN AXIAL BLOOD PUMP , 2003 .

[2]  Xinwei Song,et al.  Inlet and outlet devices for rotary blood pumps. , 2004, Artificial organs.

[3]  Kazumitsu Sekine,et al.  Computational fluid dynamics analysis of an intra-cardiac axial flow pump. , 2003, Artificial organs.

[4]  J D Hellums,et al.  Red blood cell damage by shear stress. , 1972, Biophysical journal.

[5]  Ikuya Nishimura,et al.  An estimation method of hemolysis within an axial flow blood pump by computational fluid dynamics analysis. , 2003, Artificial organs.

[6]  H Reul,et al.  Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics. , 2001, Artificial organs.

[7]  Jiyuan Tu,et al.  Vortical flow analysis , 2010 .

[8]  Marcus Hormes,et al.  A validated computational fluid dynamics model to estimate hemolysis in a rotary blood pump. , 2005, Artificial organs.

[9]  J F Antaki,et al.  A mathematical model for shear-induced hemolysis. , 1995, Artificial organs.

[10]  Yan Zhang,et al.  A novel integrated rotor of axial blood flow pump designed with computational fluid dynamics. , 2007, Artificial organs.

[11]  L. J. Wurzinger,et al.  Mechanical bloodtrauma. An overview , 1986 .

[12]  Jiyuan Tu,et al.  Numerical Simulation and Experimental Validation of Swirling Flow in Spiral Vortex Ventricular Assist Device , 2010, The International journal of artificial organs.

[13]  C. Hall,et al.  DEVELOPMENT OF ARTIFICIAL INTRATHORACIC CIRCULATORY PUMPS. , 1964, American journal of surgery.

[14]  D B Olsen,et al.  The history of continuous-flow blood pumps. , 2000, Artificial organs.

[15]  Giuseppe Pascazio,et al.  Computational prediction of mechanical hemolysis in aortic valved prostheses , 2012 .

[16]  Xinwei Song,et al.  Studies of turbulence models in a computational fluid dynamics model of a blood pump. , 2003, Artificial organs.

[17]  J F Antaki,et al.  Development of an axial flow blood pump LVAS. , 1992, ASAIO journal.

[18]  Derek Abbott,et al.  Cardiac Flow Analysis Applied to Phase Contrast Magnetic Resonance Imaging of the Heart , 2009, Annals of Biomedical Engineering.

[19]  J C ROSE,et al.  The hemodynamic alterations produced by a plastic valvular prosthesis for severe aortic insufficiency in man. , 1954, The Journal of clinical investigation.

[20]  M E Debakey The odyssey of the artificial heart. , 2000, Artificial organs.

[21]  Derek Abbott,et al.  Cardiac flow component analysis. , 2010, Medical engineering & physics.

[22]  M E DeBakey,et al.  Development of a ventricular assist device. , 2008, Artificial organs.

[23]  Jiyuan Tu,et al.  Experimental and numerical study on the hemodynamics of stenosed carotid bifurcation , 2010, Australasian Physical & Engineering Sciences in Medicine.

[24]  Pascal Verdonck,et al.  Red Cell Injury Assessed in a Numeric Model of a Peripheral Dialysis Needle , 1996 .

[25]  C Bludszuweit,et al.  Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump. , 1995, Artificial organs.

[26]  Leok Poh Chua,et al.  NUMERICAL STUDIES OF AN AXIAL FLOW BLOOD PUMP WITH DIFFERENT DIFFUSER DESIGNS , 2013 .

[27]  Klaus Affeld,et al.  Numerical estimation of blood damage in artificial organs. , 2004, Artificial organs.

[28]  Alexandrina Untaroiu,et al.  Numerical and experimental analysis of an axial flow left ventricular assist device: the influence of the diffuser on overall pump performance. , 2005, Artificial organs.

[29]  Derek Abbott,et al.  Medical image diagnostics based on computer-aided flow analysis using magnetic resonance images , 2012, Comput. Medical Imaging Graph..

[30]  F. Arabía,et al.  Ventricular assist devices today and tomorrow. , 2010, Journal of cardiothoracic and vascular anesthesia.

[31]  U Losert,et al.  Mechanical blood traumatization by tubing and throttles in in vitro pump tests: experimental results and implications for hemolysis theory. , 2008, Artificial organs.

[32]  R. Opitz,et al.  A Couette viscometer for short time shearing of blood. , 1980, Biorheology.

[33]  H. Reul,et al.  Estimation of Shear Stress-related Blood Damage in Heart Valve Prostheses - in Vitro Comparison of 25 Aortic Valves , 1990, The International journal of artificial organs.

[34]  Xiaoyu Luo,et al.  Blood flow and damage by the roller pumps during cardiopulmonary bypass , 2005 .

[35]  Zhaohui Yao,et al.  Shape optimization of the diffuser blade of an axial blood pump by computational fluid dynamics. , 2010, Artificial organs.