Hinged external fixation of the elbow: optimal axis alignment to minimize motion resistance.

OBJECTIVE To establish an optimal single hinge axis position for application of hinged external fixation to the elbow joint. DESIGN Cadaveric biomechanical investigation. SETTING A customized motion transducer applied passive elbow motion to six cadaveric upper extremities. The instant rotation axis of the humero-ulnar articulation was determined from three-dimensional kinematic data acquired by an electromagnetic motion tracking system. For each specimen, an optimal fixator hinge position was calculated from these motion data. INTERVENTION A prototype articulated external fixator was applied to the elbow, first with its hinge aligned along the computed optimal position. Then the fixator was mounted in sixteen distinct off-axis positions. MAIN OUTCOME MEASURE Additional resistance to joint motion (in terms of energy) corresponding to deliberately introduced amounts of relative malalignment between the optimal elbow axis and the actual fixator hinge axis. RESULTS Aligning the fixator hinge along the optimized axis position resulted in a minimal amount of energy (0.15 joules) needed to rotate the elbow through a prescribed range of motion. Malpositioning the hinge by ten millimeters caused up to ten times that amount of motion resistance. CONCLUSIONS An optimal fixator hinge position can be determined to minimize the increase in motion resistance due to fixator application. The severely increased motion resistance associated with small amounts of malalignment between the fixator hinge and the anatomic elbow axis suggests the need for highly accurate fixator hinge application.

[1]  K. An,et al.  Biomechanics of the elbow and forearm. , 1994, Hand clinics.

[2]  B. Morrey Distraction arthroplasty. Clinical applications. , 1993, Clinical orthopaedics and related research.

[3]  J. London Kinematics of the elbow. , 1981, The Journal of bone and joint surgery. American volume.

[4]  J. Jupiter,et al.  Non-union of fractures of the distal end of the humerus. , 1988, The Journal of bone and joint surgery. American volume.

[5]  Kai-Nan An,et al.  Articular and ligamentous contributions to the stability of the elbow joint , 1983, The American journal of sports medicine.

[6]  T. Brown,et al.  Assessment of elbow joint kinematics in passive motion by electromagnetic motion tracking , 2000, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[7]  K. An,et al.  Elbow subluxation and dislocation. A spectrum of instability. , 1992, Clinical orthopaedics and related research.

[8]  E. Chao,et al.  A biomechanical study of normal functional elbow motion. , 1981, The Journal of bone and joint surgery. American volume.

[9]  M. V. Volkov,et al.  Restoration of function in the knee and elbow with a hinge-distractor apparatus. , 1975, The Journal of bone and joint surgery. American volume.

[10]  J T Bryant,et al.  Geometry of the humeroulnar joint , 1988, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[11]  J. Bennett,et al.  Surgical management of chronic medial elbow instability. , 1992, Clinical orthopaedics and related research.

[12]  B. Morrey,et al.  Results of treatment of fracture-dislocations of the elbow. , 1987, Clinical orthopaedics and related research.

[13]  A. Weiland,et al.  Valgus stability of the elbow , 1987, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[14]  B F Morrey,et al.  Current concepts in the treatment of fractures of the radial head, the olecranon, and the coronoid. , 1995, Instructional course lectures.

[15]  S. Carmichael,et al.  Anatomy of the ulnar part of the lateral collateral ligament of the elbow , 1992 .

[16]  B. Morrey,et al.  Use of Distraction Arthroplasty in Unstable Fracture Dislocations of the Elbow , 1995, Clinical orthopaedics and related research.

[17]  P S Walker,et al.  Biomechanical basis for elbow hinge-distractor design. , 1987, Clinical orthopaedics and related research.

[18]  H. Hastings,et al.  Post-traumatic contracture of the elbow , 1998 .

[19]  T D Brown,et al.  Factors influencing accuracy of screw displacement axis detection with a D.C.-based electromagnetic tracking system. , 1998, Journal of biomechanical engineering.

[20]  E Y Chao,et al.  The effect of partial removal of proximal ulna on elbow constraint. , 1986, Clinical orthopaedics and related research.

[21]  James A. Johnson,et al.  Accuracy of an electromagnetic tracking device: a study of the optimal range and metal interference. , 1996, Journal of biomechanics.

[22]  B F Morrey,et al.  Valgus stability of the elbow. A definition of primary and secondary constraints. , 1991, Clinical orthopaedics and related research.

[23]  E. Chao,et al.  Biomechanical study of the elbow following excision of the radial head. , 1979, The Journal of bone and joint surgery. American volume.

[24]  B F Morrey,et al.  Functional anatomy of the ligaments of the elbow. , 1985, Clinical orthopaedics and related research.

[25]  B. Morrey Post-traumatic contracture of the elbow. Operative treatment, including distraction arthroplasty. , 1990, The Journal of bone and joint surgery. American volume.

[26]  G. W. Woods,et al.  Biomechanics of elbow instability: the role of the medial collateral ligament. , 1980, Clinical orthopaedics and related research.

[27]  B F Morrey,et al.  Biomechanical study of ligaments around the elbow joint. , 1991, Clinical orthopaedics and related research.

[28]  K. An,et al.  Elbow subluxation and dislocation: A spectrum , 1991 .

[29]  Peter S. Walker,et al.  Biomechanical basis for elbow hinge-distractor design. , 1987 .

[30]  O. Johnell,et al.  Dislocations of the elbow and intraarticular fractures. , 1989, Clinical orthopaedics and related research.

[31]  E Y Chao,et al.  Three-dimensional rotation of the elbow. , 1978, Journal of biomechanics.

[32]  R. Protzman Dislocation of the elbow joint. , 1978, The Journal of bone and joint surgery. American volume.