Are Melodic Medical Equipment Alarms Easily Learned?

BACKGROUND:We tested melodic auditory alarms recommended in the IEC 60601-1-8 standard for medical electrical equipment for ease of learning and discrimination, and for effectiveness during a timeshared task. METHODS:Twenty-two critical care nurses learned the IEC 60601-1-8 melodic alarms over two training sessions more than a week apart, with or without mnemonics suggested in the standard. Subsequently, the nurses identified alarms arriving at quasi random intervals while performing a timeshared arithmetic task. RESULTS:Only one nurse (4.5%) identified the alarms with 100% accuracy after two training sessions. Mnemonics did not improve overall alarm identification accuracy (mnemonic = 56%, nonmnemonic = 55%) but led to a narrower range of confusions between alarms. Nurses responded faster (P < 0.0001) and more accurately (P = 0.032) to medium priority than high priority alarms, despite rating high priority alarms as sounding more urgent (P < 0.0001). Nurses with at least 1 yr of formal musical training identified the alarms much more accurately (musical training = 73%, no musical training = 38%, P < 0.0001), perceived a greater distinction between high and medium priority alarms (P = 0.002), and found identifying the alarms easier overall (P = 0.023). During the timeshared task, nurses’ responses were slower (P = 0.002) and became less accurate (P = 0.02). CONCLUSIONS:The slow rate of learning and persistent confusions suggest that the IEC 60601-1-8 melodic alarms should be redesigned before they are adopted for clinical practice.

[1]  Stephen Brewster,et al.  A Detailed Investigation into the Effectiveness of Earcons , 1997 .

[2]  Sarah Williams,et al.  Measuring the performance of audible alarms for anaesthesia , 2005, Physiological measurement.

[3]  Penelope M. Sanderson,et al.  Auditory alarms, medical standards, and urgency , 2006 .

[4]  Neville Stanton Human factors of alarm design , 1994 .

[5]  M. Weinger Proposed new alarm standards may make a bad situation worse. , 1991, Anesthesiology.

[6]  Clifford Goodman,et al.  American Society of Anesthesiologists , 2020, Definitions.

[7]  J. Edworthy,et al.  Alarms and human behaviour: implications for medical alarms. , 2006, British journal of anaesthesia.

[8]  Penelope M. Sanderson,et al.  Overlapping Melodic Alarms Are Almost Indiscriminable , 2007, Hum. Factors.

[9]  Frank E. Block,et al.  A Proposed New Set of Alarm Sounds which Satisfy Standards and Rationale to Encode Source Information , 2004, Journal of Clinical Monitoring and Computing.

[10]  J. Edworthy,et al.  Fewer but better auditory alarms will improve patient safety , 2005, Quality and Safety in Health Care.

[11]  Judy Edworthy,et al.  Sources of confusion in intensive therapy unit alarms , 1995 .

[12]  P M Sanderson,et al.  Learnability and discriminability of melodic medical equipment alarms * , 2006, Anaesthesia.

[13]  Kerr Jh,et al.  Symposium on anaesthetic equipment. Warning devices. , 1985 .

[14]  Stephen A. Brewster,et al.  Understanding concurrent earcons: Applying auditory scene analysis principles to concurrent earcon recognition , 2004, TAP.