Thermal expansion imaging for monitoring lesion depth using M-mode ultrasound during cardiac RF ablation: in vitro study

PurposeWe demonstrate a novel method for automatic direct lesion depth (LD) tracking during coagulation from time series of a single A-mode ultrasound (US) transducer custom fit at the tip of a RFA catheter. This method is named thermal expansion imaging (TEI).MethodsA total of 35 porcine myocardium samples were ablated (LD 0.5–5 mm) while acquiring US, electrical impedance (EI) and contact force (CF) data. US images are generated in real time in terms of echo intensity (M-mode) and phase (TEI). For TEI, displacements between US time series are estimated with time-domain cross-correlation. A modified least squares strain estimation with temporal and depth smoothing reveals a thermal expansion boundary (TEB)—negative zero-crossing of temporal strain—which is associated to the coagulated tissue front.ResultsM-mode does not reliably delineate RFA lesions. TEI images reveal a traceable TEB with RMSE $$=$$= 0.50 mm and $$R^{2}=0.85$$R2=0.85 with respect to visual observations. The conventional technique, EI, shows lower $$R^{2}=0.7$$R2=0.7 and $$>$$>200 % variations with CF. The discontinuous time progression of the TEB is qualitatively associated to tissue heterogeneity and CF variations, which are directly traceable with TEI. The speed of sound, measured in function of tissue temperature, increases up to a plateau at 55 $$^{\circ }\hbox {C}$$∘C, which does not explain the observed strain bands in the TEB.ConclusionsTEI successfully tracks LD in in vitro experiments based on a single US transducer and is robust to catheter/tissue contact, ablation time and even tissue heterogeneity. The presence of a TEB suggests thermal expansion as the main strain mechanism during coagulation, accompanied by compression of the adjacent non-ablated tissue. The isolation of thermally induced displacements from in vivo motion is a matter of future research. TEI is potentially applicable to other treatments such as percutaneous RFA of liver and high-intensity focused ultrasound.

[1]  Kullervo Hynynen,et al.  Review of ultrasound therapy , 1997, 1997 IEEE Ultrasonics Symposium Proceedings. An International Symposium (Cat. No.97CH36118).

[2]  Indu Khurana,et al.  Textbook of Medical Physiology , 2005 .

[3]  D. Shah,et al.  A critical appraisal of cardiac ablation technology for catheter-based treatment of atrial fibrillation , 2011, Expert review of medical devices.

[4]  Hendrik Lambert,et al.  Area Under the Real‐Time Contact Force Curve (Force–Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile Model , 2010, Journal of cardiovascular electrophysiology.

[5]  John G. Webster,et al.  Using electrical impedance to predict catheter-endocardial contact during RF cardiac ablation , 2002, IEEE Transactions on Biomedical Engineering.

[6]  Christine M. Schubert,et al.  Effects of Gap Geometry on Conduction Through Discontinuous Radiofrequency Lesions , 2006, Circulation.

[7]  Fei Zuo,et al.  Real-time lesion assessment using a novel combined ultrasound and radiofrequency ablation catheter. , 2011, Heart rhythm.

[8]  Elia Biganzoli,et al.  Updated Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial Fibrillation , 2005, Circulation. Arrhythmia and electrophysiology.

[9]  Aravindan Kolandaivelu,et al.  Noninvasive Assessment of Tissue Heating During Cardiac Radiofrequency Ablation Using MRI Thermography , 2008, Circulation. Arrhythmia and electrophysiology.

[10]  LEONARD T. BLOUIN,et al.  Assessment of Effects of a Radiofrequency Energy Field and Thermistor Location in an Electrode Catheter on the Accuracy of Temperature Measurement , 1991, Pacing and clinical electrophysiology : PACE.

[11]  H. Nakagawa,et al.  High Incidence of Thrombus Formation Without Impedance Rise During Radiofrequency Ablation Using Electrode Temperature Control , 2003, Pacing and clinical electrophysiology : PACE.

[12]  Elisa E. Konofagou,et al.  Harmonic Motion Imaging in HIFU Monitoring , 2008 .

[13]  J Clémenty,et al.  Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. , 1998, The New England journal of medicine.

[14]  W. Kannel,et al.  Epidemiologic features of chronic atrial fibrillation: the Framingham study. , 1982, The New England journal of medicine.

[15]  Henry R. Halperin,et al.  Non-invasive Assessment of Tissue Heating During Cardiac Radiofrequency Ablation Using MRI Thermography Running , 2010 .

[16]  Peter R. Hoskins,et al.  Diagnostic ultrasound : physics and equipment , 2010 .

[17]  Yang Li,et al.  An In Vitro Assessment of Acoustic Radiation Force Impulse Imaging for Visualizing Cardiac Radiofrequency Ablation Lesions , 2010, Journal of cardiovascular electrophysiology.

[18]  P. Baki,et al.  Sensorized Cardiac Radiofrequency Ablation System for Lesion Depth Assessment , 2015 .

[19]  R C Chan,et al.  Catheter Tip Orientation Affects Radiofrequency Ablation Lesion Size in the Canine Left Ventricle , 1999, Pacing and clinical electrophysiology : PACE.

[20]  F. Kallel,et al.  A Least-Squares Strain Estimator for Elastography , 1997, Ultrasonic imaging.

[21]  G. Haar,et al.  Ultrasound focal beam surgery. , 1995 .

[22]  R N Hauer,et al.  Control of radiofrequency lesion size by power regulation. , 1989, Circulation.

[23]  G. Székely,et al.  Design and characterization of a novel, robust, tri-axial force sensor , 2013 .

[24]  Chi Hyung Seo,et al.  The feasibility of using thermal strain imaging to regulate energy delivery during intracardiac radio-frequency ablation , 2011, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[25]  D. Haines,et al.  Ablation for the treatment of arrhythmias , 1996, Proc. IEEE.

[26]  G. ter Haar,et al.  Ultrasound focal beam surgery. , 1995, Ultrasound in medicine & biology.

[27]  A. Luna,et al.  Clinical Electrocardiography: A Textbook , 1993 .

[28]  H. Lambert,et al.  Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and Thrombus , 2008, Circulation. Arrhythmia and electrophysiology.

[29]  Sheng-Wen Huang,et al.  Thermal strain imaging: a review , 2011, Interface Focus.

[30]  J A Zagzebski,et al.  Ultrasound monitoring of temperature change during radiofrequency ablation: preliminary in-vivo results. , 2002, Ultrasound in medicine & biology.

[31]  Matthew O'Donnell,et al.  Experimental Studies With a 9F Forward‐Looking Intracardiac Imaging and Ablation Catheter , 2009, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[32]  H Matsuo,et al.  Relationship between myocardial tissue density measured by microgravimetry and sound speed measured by acoustic microscopy. , 1999, Ultrasound in medicine & biology.

[33]  Douglas L Packer,et al.  Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary Vein , 2004, Circulation.

[34]  J C Bamber,et al.  Acoustic properties of lesions generated with an ultrasound therapy system. , 1993, Ultrasound in medicine & biology.

[35]  Daniel L Lustgarten,et al.  Ablation using irrigated radiofrequency: a hands-on guide. , 2008, Heart rhythm.

[36]  Joshua J. E. Blauer,et al.  Identification and Acute Targeting of Gaps in Atrial Ablation Lesion Sets Using a Real-Time Magnetic Resonance Imaging System , 2012, Circulation. Arrhythmia and electrophysiology.

[37]  Y. Okumura,et al.  A Systematical Analysis of In Vivo Contact Forces on Virtual Catheter Tip/Tissue Surface Contact during Cardiac Mapping and Intervention , 2008, Journal of cardiovascular electrophysiology.

[38]  L Solbiati,et al.  Small hepatocellular carcinoma: treatment with radio-frequency ablation versus ethanol injection. , 1999, Radiology.

[39]  Dieter Haemmerich,et al.  Biophysics of radiofrequency ablation. , 2010, Critical reviews in biomedical engineering.

[40]  D. Haines,et al.  Cellular Electrophysiological Effects of Hyperthermia on Isolated Guinea Pig Papillary Muscle Implications for Catheter Ablation , 1993, Circulation.

[41]  Michel Bertrand,et al.  Monitoring the formation of thermal lesions with heat-induced echo-strain imaging: a feasibility study. , 2005, Ultrasound in medicine & biology.

[42]  Cheri X Deng,et al.  High-frequency ultrasound m-mode imaging for identifying lesion and bubble activity during high-intensity focused ultrasound ablation. , 2012, Ultrasound in medicine & biology.

[43]  Leslie Oley,et al.  Use of a Novel Endoscopic Catheter for Direct Visualization and Ablation in an Ovine Model of Chronic Myocardial Infarction , 2012, Circulation.

[44]  D. Haines,et al.  The Biophysics of Radiofrequency Catheter Ablation in the Heart: The Importance of Temperature Monitoring , 1993, Pacing and clinical electrophysiology : PACE.

[45]  F. Epstein,et al.  Electrocardiographic Findings Among the Adult Population of a Total Natural Community, Tecumseh, Michigan , 1965, Circulation.

[46]  Jeffrey C Bamber,et al.  Imaging of temperature-induced echo strain: preliminary in vitro study to assess feasibility for guiding focused ultrasound surgery. , 2004, Ultrasound in medicine & biology.

[47]  L. Sheppard,et al.  In vitro measurement of speed of sound during coagulate tissue heating , 1996, 1996 IEEE Ultrasonics Symposium. Proceedings.