400-ps time resolution with a passively quenched avalanche photodiode.

Avalanche photodiodes (APD's) operated in a single-photon-counting Geiger mode are becoming attractive alternatives to photomultiplier tubes for low-light-level detection and signal timing. By paying careful attention to the design and construction of a simple APD passive quenching circuit to reduce stray capacitances, we directly measured a time resolution of 410 ps FWHM for a commercial APD. A more detailed data analysis shows the actual time resolution to be ~ 390 ps FWHM. This is believed to be the most accurate time response for such a simple, inexpensive, and widely available device achieved to date.