A Biomechanical Study of Anterior Thoracolumbar Screw Fixation

Study Design. The pullout strength of unicortical and bicortical screws in thoracic and lumbar vertebral bodies was measured as a function of bone mineral density. Objectives. To determine the influence of bone mineral density and screw insertion technique on the stability of anterior thoracolumbar spine screw fixation. Summary of Background Data. No previous study has addressed the specific technique of screw insertion or stability of screw fixation in the anterior spine. Methods. Fifty‐one human thoracic vertebral bodies were tested in pullout with 6.55‐mm cancellous screws inserted using unicortical and bicortical techniques. Results. Pullout force increased exponentially with increasing bone mineral density for unicortical and bicortical screws. Bicortical screws were significantly stronger in resisting pullout than unicortical screws. Conclusion. Advancing an anterior vertebral body screw to engage the second cortex increases resistance to pullout by 25‐44%, depending on vertebral bone mineral density. The difference in resistance between unicortical and bicortical techniques was smaller in specimens with low mineral densities.

[1]  P. McAfee,et al.  Anterior spinal fixation after lumbar corpectomy. A study in dogs. , 1991, The Journal of bone and joint surgery. American volume.

[2]  B. Fredrickson,et al.  The Syracuse I‐Plate , 1991, Spine.

[3]  M M Panjabi,et al.  Biomechanical Evaluation of Spinal Fixation Devices: I. A Conceptual Framework , 1988, Spine.

[4]  J. Kostuik Anterior Fixation for Burst Fractures of the Thoracic and Lumbar Spine with or without Neurological Involvement , 1988, Spine.

[5]  C. Hodge,et al.  Early Clinical Experience with the Syracuse I-Plate: An Anterior Spinal Fixation Device , 1988, Spine.

[6]  K. Kaneda,et al.  Burst Fractures with Neurologic Deficits of the Thoracolumbar‐Lumbar Spine Results of Anterior Decompression and Stabilization with Anterior Instrumentation , 1984, Spine.

[7]  H. Dunn Anterior stabilization of thoracolumbar injuries. , 1984, Clinical orthopaedics and related research.

[8]  J. Kostuik Anterior Spinal Cord Decompression for Lesions of the Thoracic and Lumbar Spine, Techniques, New Methods of Internal Fixation Results , 1983, Spine.

[9]  J. C. Flynn,et al.  Anterior fusion of the lumbar spine. End-result study with long-term follow-up. , 1979, The Journal of bone and joint surgery. American volume.

[10]  A. Dwyer Experience of anterior correction of scoliosis. , 1973, Clinical orthopaedics and related research.

[11]  R. Taylor,et al.  Anterior transperitoneal lumbar fusion. , 1971, The Journal of bone and joint surgery. British volume.

[12]  C. J. Campbell FORUM ON FUNDAMENTAL PROBLEMS IN ORTHOPAEDIC SURGERY: Meeting of the American College of Surgeons , 1959 .

[13]  H. Dunn,et al.  Fixation of Dwyer screws for the treatment of scoliosis. A postmortem study. , 1977, The Journal of bone and joint surgery. American volume.