Reduction of procoagulant potential of b-datum leakage jet flow in bileaflet mechanical heart valves via application of vortex generator arrays.

Current designs of bileaflet mechanical heart valves put patients at an increased risk of thromboembolism. In particular, regurgitant flow through the b-datum line is associated with nonphysiologic flow characteristics such as elevated shear stresses, regions of recirculation, and increased mixing, all of which may promote thrombus formation. We have previously shown that passive flow control in the form of vortex generators mounted on the downstream leaflet surfaces can effectively diminish turbulent stresses. The objective of the current work is thus to determine the effect of vortex generators on the thromboembolic potential of the b-datum line leakage jet and to correlate that effect with the vortex generator-induced changes to the flow structure. Flow experiments were performed using a steady model of the transient b-datum line jet. These experiments encompassed flow visualization to gain an overall picture of the flow system, particle image velocimetry to quantify the flow field in detail, and in vitro experiments with human blood to quantify thrombus formation in response to the applied passive flow control. Thrombus formation was quantified over time by an assay for thrombin-antithrombin III (TAT III). In comparing results with and without vortex generators, significantly lower mean TAT III levels were observed at one time point for the case with vortex generators. Also, the TAT III growth rate of the case with vortex generators was significantly lower. While no differences in jet spreading were found with and without vortex generators, lower peak turbulent stresses were observed for the case with vortex generators. The results thus demonstrate the potential of applying passive flow control to cardiovascular hardware in order to mitigate the hemodynamic factors leading to thrombus formation.

[1]  Ajit P Yoganathan,et al.  Flow and thrombosis at orifices simulating mechanical heart valve leakage regions. , 2006, Journal of biomechanical engineering.

[2]  S. Deutsch,et al.  A detailed fluid mechanics study of tilting disk mechanical heart valve closure and the implications to blood damage. , 2008, Journal of biomechanical engineering.

[3]  J. Hathcock Flow effects on coagulation and thrombosis. , 2006, Arteriosclerosis, thrombosis, and vascular biology.

[4]  M. Gharib,et al.  Vortex shedding as a mechanism for free emboli formation in mechanical heart valves. , 2000, Journal of biomechanical engineering.

[5]  V. Walley,et al.  Comparative results with the St. Jude Medical and Medtronic Hall mechanical valves. , 1995, The Journal of thoracic and cardiovascular surgery.

[6]  H. S. Udaykumar,et al.  Two-Dimensional Dynamic Simulation of Platelet Activation During Mechanical Heart Valve Closure , 2006, Annals of Biomedical Engineering.

[7]  S. C. Low,et al.  The flow characteristics of a confined square jet with mixing tabs , 1998 .

[8]  A. Fallon,et al.  The Development of a Novel in vitro Flow System to Evaluate Platelet Activation and Procoagulant Potential Induced by Bileaflet Mechanical Heart Valve Leakage Jets , 2006 .

[9]  J. D. Hellums,et al.  1993 Whitaker lecture: Biorheology in thrombosis research , 1994, Annals of Biomedical Engineering.

[10]  Danny Bluestein,et al.  Fluid mechanics of arterial stenosis: Relationship to the development of mural thrombus , 1997, Annals of Biomedical Engineering.

[11]  S. Sutera,et al.  Flow-induced trauma to blood cells. , 1977, Circulation research.

[12]  L Zuckerman,et al.  Shear-induced activation of platelets. , 1979, Journal of biomechanics.

[13]  H. Pelzer,et al.  Determination of Human Thrombin-Antithrombin III Complex in Plasma with an Enzyme-Linked Immunosorbent Assay , 1988, Thrombosis and Haemostasis.

[14]  L. J. S. Bradbury,et al.  The distortion of a jet by tabs , 1975, Journal of Fluid Mechanics.

[15]  Richard A. Lange,et al.  Prosthetic heart valves. , 1996, The New England journal of medicine.

[16]  Lakshmi Prasad Dasi,et al.  Procoagulant Properties of Flow Fields in Stenotic and Expansive Orifices , 2007, Annals of Biomedical Engineering.

[17]  E. Baudet,et al.  A 5 1/2 year experience with the St. Jude Medical cardiac valve prosthesis. Early and late results of 737 valve replacements in 671 patients. , 1985, The Journal of thoracic and cardiovascular surgery.

[18]  Hélène A. Simon,et al.  Vorticity dynamics of a bileaflet mechanical heart valve in an axisymmetric aorta , 2007 .

[19]  J. Moake,et al.  THE RESPONSE OF HUMAN PLATELETS TO SHEAR STRESS AT SHORT EXPOSURE TIMES , 1977, Transactions - American Society for Artificial Internal Organs.

[20]  Nicholas J. Lawson,et al.  Oscillatory Flow in a Physical Model of a Thin Slab Casting Mould With a Bifurcated Submerged Entry Nozzle , 2002 .

[21]  Jan Vierendeels,et al.  Comparison of the hemodynamic and thrombogenic performance of two bileaflet mechanical heart valves using a CFD/FSI model. , 2007, Journal of biomechanical engineering.

[22]  E. Lefrak,et al.  Cardiac valve prostheses , 1979 .

[23]  M. Sarsam,et al.  The St. Jude Medical prosthesis. A thirteen-year experience. , 1994, The Journal of thoracic and cardiovascular surgery.

[24]  G. Lauder,et al.  Passive and Active Flow Control by Swimming Fishes and Mammals , 2006 .

[25]  J D Hellums,et al.  Red blood cell damage by shear stress. , 1972, Biophysical journal.

[26]  Mark F. Reeder,et al.  Control of an axisymmetric jet using vortex generators , 1994 .

[27]  S. S. Kim,et al.  Long-term results with St. Jude Medical and CarboMedics prosthetic heart valves. , 2001, The Journal of heart valve disease.

[28]  H. Hui,et al.  Changes to the vortical and turbulent structure of jet flows due to mechanical tabs , 1999 .

[29]  J. Mi,et al.  Mixing Characteristics of a Flapping Jet from a Self-Exciting Nozzle , 2001 .

[30]  Mohamed Gad-el-Hak,et al.  Flow Control: Passive, Active, and Reactive Flow Management , 2000 .

[31]  N. Lawson,et al.  Crossflow Characteristics of an Oscillating Jet in a Thin Slab Casting Mould , 1999 .

[32]  Helene A. Simon Influence of the Implant Location on the Hinge and Leakage Flow Fields Through Bileaflet Mechanical Heart Valves , 2004 .

[33]  P. Lu,et al.  A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow. , 2001, Journal of biomechanics.

[34]  A. Yoganathan,et al.  Thrombin formation in vitro in response to shear-induced activation of platelets. , 2007, Thrombosis research.

[35]  A. Yoganathan,et al.  Passive flow control of bileaflet mechanical heart valve leakage flow. , 2008, Journal of biomechanics.

[36]  Nicholas J. Lawson,et al.  SELF-SUSTAINED OSCILLATION OF A SUBMERGED JET IN A THIN RECTANGULAR CAVITY , 2001 .