Robust Longitudinal Ankle Edema Assessment Using Wearable Bioimpedance Spectroscopy

Objective: We present a robust methodology for tracking ankle edema longitudinally based on bioimpedance spectroscopy (BIS). Methods: We designed a miniaturized BIS measurement system and employed a novel calibration method that enables accurate, high-resolution measurements with substantially lower power consumption than conventional approaches. Using this state-of-the-art wearable BIS measurement system, we developed a differential measurement technique for robust assessment of ankle edema. This technique addresses many of the major challenges in longitudinal BIS-based edema assessment, including day-to-day variability in electrode placement, positional/postural variability, and intersubject variability. Results: We first evaluated the hardware in bench-top testing, and determined the error of the bioimpedance measurements to be 0.4 Ω for the real components and 0.54 Ω for the imaginary components with a resolution of 0.2 Ω. We then validated the hardware and differential measurement technique in: 1) an ex vivo, fresh-frozen, cadaveric limb model, and 2) a cohort of 11 human subjects for proof of concept (eight healthy controls and five subjects with recently acquired acute unilateral ankle injury). Conclusion: The hardware design, with novel calibration methodology, and differential measurement technique can potentially enable long-term quantification of ankle edema throughout the course of rehabilitation following acute ankle injuries. Significance: This could lead to better-informed decision making regarding readiness to return to activities and/or tailoring of rehabilitation activities to an individual's changing needs.

[1]  M. Schweitzer,et al.  Fluid in normal and abnormal ankle joints: amount and distribution as seen on MR images. , 1994, AJR. American journal of roentgenology.

[2]  F. Backx,et al.  A Systematic Review on the Treatment of Acute Ankle Sprain , 2011, Sports medicine.

[3]  Dinesh Bhatia,et al.  Portable impedance measurement device for sweat based glucose detection , 2017, 2017 IEEE 14th International Conference on Wearable and Implantable Body Sensor Networks (BSN).

[4]  Sverre Grimnes,et al.  Bioimpedance and Bioelectricity Basics , 2000 .

[5]  Shweta Shah,et al.  Incidence and Cost of Ankle Sprains in United States Emergency Departments , 2016, Sports health.

[6]  Fernando Seoane,et al.  An analog front-end enables electrical impedance spectroscopy system on-chip for biomedical applications , 2008, Physiological measurement.

[7]  G. Chertow,et al.  Multifrequency bioelectrical impedance estimates the distribution of body water. , 1995, Journal of applied physiology.

[8]  Hakan Toreyin,et al.  Wearable Vector Electrical Bioimpedance System to Assess Knee Joint Health , 2017, IEEE Transactions on Biomedical Engineering.

[9]  Mohammed Attia,et al.  A wearable device for monitoring and prevention of repetitive ankle sprain , 2015, 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[10]  B. Koes,et al.  What is the clinical course of acute ankle sprains? A systematic literature review. , 2008, The American journal of medicine.

[11]  Ji-Gou Liu,et al.  Accuracy improvement of impedance measurements by using the self-calibration , 1999 .

[12]  E. M. Avila,et al.  Comparison of Magnetic Resonance Imaging to Physical Examination for Syndesmotic Injury after Lateral Ankle Sprain , 2011, Foot & ankle international.

[13]  New derivation method and simulation of skin effect in biological tissue. , 2015, Bio-medical materials and engineering.

[14]  K. M. Chan,et al.  An epidemiological survey on ankle sprain. , 1994, British journal of sports medicine.

[15]  Jay Hertel,et al.  Rehabilitation of ankle and foot injuries in athletes. , 2010, Clinics in sports medicine.

[16]  Hélène Morel,et al.  Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods. , 2008, Medical engineering & physics.

[17]  Omer T. Inan,et al.  A Robust System for Longitudinal Knee Joint Edema and Blood Flow Assessment Based on Vector Bioimpedance Measurements , 2016, IEEE Transactions on Biomedical Circuits and Systems.

[18]  R. Bragós,et al.  An implantable bioimpedance monitor using 2.45 GHz band for telemetry , 2013, Physiological Measurement.

[19]  Youlian Hong,et al.  A Systematic Review on Ankle Injury and Ankle Sprain in Sports , 2007, Sports medicine.

[20]  Cristian-Gyozo Haba,et al.  Wearable impedance analyzer based on AD5933 , 2014, 2014 International Conference and Exposition on Electrical and Power Engineering (EPE).

[21]  R. Marti,et al.  Physical examination is sufficient for the diagnosis of sprained ankles. , 1996, The Journal of bone and joint surgery. British volume.

[22]  Fernando Seoane,et al.  AD5933-based Spectrometer for Electrical Bioimpedance Applications , 2010 .

[23]  Todd J. Freeborn,et al.  Biceps tissue bioimpedance changes from isotonic exercise-induced fatigue at different intensities , 2018 .

[24]  L Nescolarde,et al.  Localized bioimpedance to assess muscle injury , 2013, Physiological measurement.

[25]  Zonghai Gao,et al.  Accuracy improvement by A three-reference calibration algorithm for a bioimpedance spectrometer , 2010, 2010 International Conference on Bioinformatics and Biomedical Technology.

[26]  A. Anandacoomarasamy,et al.  Long term outcomes of inversion ankle injuries , 2005, British Journal of Sports Medicine.

[27]  Fernando Seoane,et al.  Full Impedance Cardiography Measurement Device Using Raspberry PI3 and System-on-Chip Biomedical Instrumentation Solutions , 2018, IEEE Journal of Biomedical and Health Informatics.

[28]  B. Waterman,et al.  The epidemiology of ankle sprains in the United States. , 2010, The Journal of bone and joint surgery. American volume.

[29]  J. Rosell,et al.  Skin impedance from 1 Hz to 1 MHz , 1988, IEEE Transactions on Biomedical Engineering.

[30]  Jia Li,et al.  Impedance Alterations in Healthy and Diseased Mice During Electrically Induced Muscle Contraction , 2016, IEEE Transactions on Biomedical Engineering.

[31]  M. Morrissey,et al.  Effect of Neuromuscular Electrical Stimulation on Ankle Swelling in the Early Period After Ankle Sprain , 2007, Physical Therapy.

[32]  L Nescolarde,et al.  Effects of muscle injury severity on localized bioimpedance measurements , 2015, Physiological measurement.

[33]  C Margo,et al.  A four-electrode low frequency impedance spectroscopy measurement system using the AD5933 measurement chip , 2013, Physiological measurement.

[34]  P. Bogonez-Franco,et al.  Performance of an implantable impedance spectroscopy monitor using ZigBee , 2010 .