Hi-Res scan mode in clinical MDCT systems: Experimental assessment of spatial resolution performance.

PURPOSE The introduction of a High-Resolution (Hi-Res) scan mode and another associated option that combines Hi-Res mode with the so-called High Definition (HD) reconstruction kernels (referred to as a Hi-Res/HD mode in this paper) in some multi-detector CT (MDCT) systems offers new opportunities to increase spatial resolution for some clinical applications that demand high spatial resolution. The purpose of this work was to quantify the in-plane spatial resolution along both the radial direction and tangential direction for the Hi-Res and Hi-Res/HD scan modes at different off-center positions. METHODS A technique was introduced and validated to address the signal saturation problem encountered in the attempt to quantify spatial resolution for the Hi-Res and Hi-Res/HD scan modes. Using the proposed method, the modulation transfer functions (MTFs) of a 64-slice MDCT system (Discovery CT750 HD, GE Healthcare) equipped with both Hi-Res and Hi-Res/HD modes were measured using a metal bead at nine different off-centered positions (0-16 cm with a step size of 2 cm); at each position, both conventional scans and Hi-Res scans were performed. For each type of scan and position, 80 repeated acquisitions were performed to reduce noise induced uncertainties in the MTF measurements. A total of 15 reconstruction kernels, including eight conventional kernels and seven HD kernels, were used to reconstruct CT images of the bead. An ex vivo animal study consisting of a bone fracture model was performed to corroborate the MTF results, as the detection of this high-contrast and high frequency task is predominantly determined by spatial resolution. Images of this animal model generated by different scan modes and reconstruction kernels were qualitatively compared with the MTF results. RESULTS At the centered position, the use of Hi-Res mode resulted in a slight improvement in the MTF; each HD kernel generated higher spatial resolution than its counterpart conventional kernel. However, the MTF along the tangential direction of the scan field of view (SFOV) was significantly degraded at off-centered positions, yet the combined Hi-Res/HD mode reduced this azimuthal MTF degradation. Images of the animal bone fracture model confirmed the improved spatial resolution at the off-centered positions through the use of the Hi-Res mode and HD kernels. CONCLUSIONS The Hi-Res/HD scan improve spatial resolution of MDCT systems at both centered and off-centered positions.

[1]  Jiang Hsieh,et al.  Reconstruction technique for focal spot wobbling , 1992, Medical Imaging.

[2]  Michael J. Flynn,et al.  Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom , 2006, SPIE Medical Imaging.

[3]  Thomas Toth,et al.  The influence of patient centering on CT dose and image noise. , 2007, Medical physics.

[4]  J. H. Siewerdsen,et al.  Design and optimization of a dedicated cone-beam CT system for musculoskeletal extremities imaging , 2011, Medical Imaging.

[5]  D. Yang,et al.  A dedicated cone-beam CT system for musculoskeletal extremities imaging: design, optimization, and initial performance characterization. , 2011, Medical physics.

[6]  H. Zaidi,et al.  Impact of miscentering on patient dose and image noise in x-ray CT imaging: phantom and clinical studies. , 2012, Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics.

[7]  Ehsan Samei,et al.  Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms. , 2012, Medical physics.

[8]  Ho Kyung Kim,et al.  An experimental study on the shift-variant MTF of CT systems using a simple cylindrical phantom , 2013, Medical Imaging.

[9]  T. Szczykutowicz,et al.  SU-F-18C-03: Loss of Image Sharpness in CT Bone Imaging Due to Positioning Within the Scan Field of View - Possible Solutions. , 2014, Medical physics.

[10]  Guang-Hong Chen,et al.  Statistical model based iterative reconstruction (MBIR) in clinical CT systems. Part II. Experimental assessment of spatial resolution performance. , 2014, Medical physics.

[11]  J. Siewerdsen,et al.  Dedicated cone-beam CT system for extremity imaging. , 2014, Radiology.

[12]  J. Siewerdsen,et al.  Extremity cone-beam CT for evaluation of medial tibiofemoral osteoarthritis: Initial experience in imaging of the weight-bearing and non-weight-bearing knee. , 2015, European journal of radiology.