Dissociation of a Medial Pivot Polyethylene in a Kinematically Aligned Total Knee Arthroplasty

An 81-year-old male patient who underwent a Medacta GMK sphere kinematically aligned (KA) total knee arthroplasty (TKA) for end-stage knee osteoarthritis presented with a dislocated medial pivot (MP) tibia polyethylene (PE) insert on routine six-week postoperative x-rays. The patient presented asymptomatic with a normal range of motion. Dissociation of a fixed-bearing (FB) PE implant is an uncommon complication after TKA. There are only a few cases reported in the literature. We report for the first time a case of non-traumatic dissociation of MP PE from the tibial baseplate in a KA TKA in an asymptomatic patient but identified on routine postoperative radiographs.

[1]  Alessandro Massè,et al.  Medial Pivot in Total Knee Arthroplasty: Literature Review and Our First Experience , 2018, Clinical medicine insights. Arthritis and musculoskeletal disorders.

[2]  Marco Schiraldi,et al.  Mechanical and kinematic alignment in total knee arthroplasty. , 2016, Annals of translational medicine.

[3]  L. Carbó,et al.  Traumatic Dislodgement of Tibial Polyethylene Insert after a High-Flex Posterior-Stabilized Total Knee Replacement , 2015, Case reports in orthopedics.

[4]  J. Goosen,et al.  Insert dissociation after fixed bearing PS constrained Genesis II total knee arthroplasty. A case series of nine patients. , 2015, Acta orthopaedica Belgica.

[5]  Y. In,et al.  Recurrent dissociation of the tibial insert after mini-subvastus posterior-stabilized total knee arthroplasty: a case report. , 2011, The Knee.

[6]  C. Beauchamp,et al.  Five cases of failure of the tibial polyethylene insert locking mechanism in one design of constrained knee arthroplasty. , 2011, The Journal of arthroplasty.

[7]  B. Lanting,et al.  Dislocated polyethylene inserts in fixed-bearing total knee arthroplasty. , 2011, The Journal of arthroplasty.

[8]  M. Hepinstall,et al.  Polyethylene subluxation: a radiographic sign of locking mechanism failure after modular total knee arthroplasty. , 2011, The Journal of arthroplasty.

[9]  R. Janssen,et al.  Spontaneous late dislocation of the high flexion tibial insert after Genesis II total knee arthroplasty. A case report. , 2009, The Knee.

[10]  John A. Anderson,et al.  Spontaneous, recurrent dislodgment of the polyethylene tibial insert after total knee arthroplasty. A case report. , 2007, The Journal of bone and joint surgery. American volume.

[11]  R. Poulter,et al.  A case of dissociation of polyethylene from its metal baseplate in a "one piece" compression-moulded AGC tibial component. , 2005, The Knee.

[12]  M. Ries Dissociation of an ultra-high molecular weight polyethylene insert from the tibial baseplate after total knee arthroplasty. A case report. , 2004, The Journal of bone and joint surgery. American volume.

[13]  B. Waelchli,et al.  Dislocation of the polyethylene inlay due to anterior tibial slope in revision total knee arthroplasty , 2001, Knee Surgery, Sports Traumatology, Arthroscopy.

[14]  U. Hedlundh,et al.  Traumatic late dissociation of the polyethylene articulating surface in a total knee arthroplasty--a case report , 2000, Acta orthopaedica Scandinavica.

[15]  P. F. Davis,et al.  Dissociation of the tibial component in total knee replacements. , 1991, Clinical orthopaedics and related research.