Effects of image artifacts on gray-value density in limited-volume cone-beam computerized tomography.

OBJECTIVE An in vitro study was designed to investigate the influence of projection data discontinuity-related artifacts in limited-volume cone-beam computerized tomography (CBCT) imaging of the jaws. STUDY DESIGN Test objects were positioned in 4 patterns in a water-filled phantom as follows: bimandible and vertebrae, bimandible, left mandible and vertebrae, and left mandible. The CT imaging of the left molar region was performed using image intensifier (II)- and flat panel detector (FPD)-based CBCT scanners. The CT value of the mandible and the adjacent soft tissue region were analyzed for density by means of an 8-bit grayscale. RESULTS The effects of artifacts were scored as the difference in relative density between the lingual and buccal soft tissue. The intensity of artifacts increased when more objects were presented outside the area being imaged. Fewer artifacts were noted in images produced by the particular FPD CBCT used in this investigation. CONCLUSION The CBCT system using an FPD resulted in fewer artifacts than the CBCT system using an II in this particular study.

[1]  S. Dunn,et al.  Feasibility of local CT of dental tissues. , 2003, Dento maxillo facial radiology.

[2]  A Kawamata,et al.  Three-dimensional computed tomography imaging in dentistry. , 2000, Dental clinics of North America.

[3]  Predrag Sukovic,et al.  Accuracy of implant placement with a stereolithographic surgical guide. , 2003, The International journal of oral & maxillofacial implants.

[4]  K Araki,et al.  Characteristics of a newly developed dentomaxillofacial X-ray cone beam CT scanner (CB MercuRay): system configuration and physical properties. , 2004, Dento maxillo facial radiology.

[5]  Max Heiland,et al.  Evaluation of quality of reformatted images from two cone-beam computed tomographic systems. , 2005, Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery.

[6]  Yoshinori Arai,et al.  A comparison of a new limited cone beam computed tomography machine for dental use with a multidetector row helical CT machine. , 2003, Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics.

[7]  Kaoru Kobayashi,et al.  Accuracy in measurement of distance using limited cone-beam computerized tomography. , 2004, The International journal of oral & maxillofacial implants.

[8]  K. Ueda,et al.  Using a flat-panel detector in high resolution cone beam CT for dental imaging. , 2004, Dento maxillo facial radiology.

[9]  J. Ambrose,et al.  Computerized transverse axial scanning (tomography): Part 2. Clinical application* , 1973 .

[10]  Akitoshi Katsumata,et al.  Measurement of mandibles with microfocus x-ray computerized tomography and compact computerized tomography for dental use. , 2004, The International journal of oral & maxillofacial implants.

[11]  G. Hounsfield Computerized transverse axial scanning (tomography): Part I. Description of system. 1973. , 1973, The British journal of radiology.

[12]  K. Bae,et al.  Efficient correction for CT image artifacts caused by objects extending outside the scan field of view. , 2000, Medical physics.

[13]  E Tammisalo,et al.  Development of a compact computed tomographic apparatus for dental use. , 1999, Dento maxillo facial radiology.

[14]  Prasit Aranyarachkul,et al.  Bone density assessments of dental implant sites: 2. Quantitative cone-beam computerized tomography. , 2005, The International journal of oral & maxillofacial implants.

[15]  J. H. Kinsey,et al.  Physics and technical considerations in the design of the DSR: a high temporal resolution volume scanner. , 1980, AJR. American journal of roentgenology.

[16]  B. Wilson,et al.  Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. , 2005, Medical physics.

[17]  A. Katsumata,et al.  Image artifact in dental cone-beam CT. , 2006, Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics.

[18]  P. Westesson,et al.  Ortho cubic super-high resolution computed tomography: a new radiographic technique with application to the temporomandibular joint. , 2001, Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics.

[19]  P. Mozzo,et al.  A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results , 1998, European Radiology.

[20]  Y. Arai,et al.  Diagnostic imaging with newly developed ortho cubic super-high resolution computed tomography (Ortho-CT). , 2000, Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics.