A printed circuit board capacitive sensor for air bubble inside fluidic flow detection

This paper presents a three-electrode capacitive fluidic sensor for detecting an air bubble inside a fluidic channel such as blood vessels, oil or medical liquid channels. The capacitor is designed and fabricated based on a printed circuit board (PCB). The electrodes are fabricated by using copper via structure through top to bottom surface of the PCB. A plastic pipe is layout through the capacitive sensor and perpendicular to the PCB surface. Capacitance of sensor changes when an air bubble inside fluidic flow cross the sensor. The capacitance change can be monitored by using a differential capacitive amplifier, a lock-in amplifier, filter and an NI acquisition card. Signal is processed and calculated on a computer. Air bubble inside the liquid flow are detected by monitor the unbalance signal between the three electrode potential voltages. Output voltage depends on the volume of the air bubble due to dielectric change between capacitor's electrodes. Output voltage is up to 53 mV when an 2.28 mm3 air bubble crosses the sensing channel. Air bubble velocity can be estimated based on the output pulse signal. This proposed fluidic sensor can be used for void fraction detection in medical devices and systems; fluidic characterization; and water---gas, oil---water and oil---water---gas multiphase flows in petroleum technology. That structure also can apply to the micro-size for detecting in microfluidic to monitor and control changes in microfluidic channels.

[1]  W. Heerens,et al.  Capacitance of Kelvin guard‐ring capacitors with modified edge geometry , 1975 .

[2]  Michal Barak,et al.  Microbubbles : Pathophysiology and clinical implications , 2005 .

[3]  Josua P. Meyer,et al.  Horizontal two-phase flow characterization for small diameter tubes with a capacitance sensor , 2007 .

[4]  Ulf Forsberg,et al.  Air bubbles pass the security system of the dialysis device without alarming. , 2007, Artificial organs.

[5]  G. Meijer,et al.  An integrated interface circuit with a capacitance-to-voltage converter as front-end for grounded capacitive sensors , 2008 .

[6]  H. Markus,et al.  Transcranial Doppler detection of circulating cerebral emboli. A review. , 1993, Stroke.

[7]  Daniel Watzenig,et al.  A review of statistical modelling and inference for electrical capacitance tomography , 2009 .

[8]  Kenneth P. Malloy,et al.  Clinical evaluation of an air embolism detection device. , 1980, Cardiovascular diseases.

[9]  Roland Zengerle,et al.  A capacitive sensor for non-contact nanoliter droplet detection , 2009 .

[10]  J. Wei Silicon MEMS for Detection of Liquid and Solid Fronts , 2010 .

[11]  Byong-Jo Yun,et al.  Design of a capacitance sensor for void fraction measurement in annular flows through a vertical pipe , 2012 .

[12]  Emilio Sardini,et al.  Measurement of small capacitance variations , 1990 .

[13]  J. Voldman,et al.  An equilibrium method for continuous-flow cell sorting using dielectrophoresis. , 2008, Analytical chemistry.

[14]  Caglar Elbuken,et al.  Detection of microdroplet size and speed using capacitive sensors , 2011 .

[15]  Shi Liu,et al.  Mass flow measurement of pneumatically conveyed solids using electrical capacitance tomography , 2008 .

[16]  Larry K. Baxter,et al.  Capacitive Sensors: Design and Applications , 1996 .

[17]  B. T. Hjertaker,et al.  Three-phase flow measurement in the petroleum industry , 2012 .

[18]  F. N. Toth A design methodology for low-cost, high-performance capacitive sensors , 1997 .