Systematic Design of a Magnetically Levitated Brushless DC Motor for a Reversible Rotary Intra‐Aortic Blood Pump

The IntraVAD is a miniature intra-aortic ventricular assist device (VAD) designed to work in series with the compromised left ventricle. A reverse-rotation control (RRc) mode has been developed to increase myocardial perfusion and reduce ventricular volume. The RRc mode includes forward rotation in systole and reverse rotation in diastole, which requires the IntraVAD to periodically reverse its rotational direction in synchrony with the cardiac cycle. This periodic reversal leads to changes in pressure force over the impeller, which makes the entire system less stable. To eliminate the mechanical wear of a contact bearing and provide active control over the axial position of the rotor, a miniature magnetically levitated bearing (i.e., the PM-Coil module) composed of two concentric permanent magnetic (PM) rings and a pair of coils-one on each side-was proposed to provide passive radial and active axial rotor stabilization. In the early design stage, the numerical finite element method (FEM) was used to optimize the geometry of the brushless DC (BLDC) motor and the maglev module, but constructing a new model each time certain design parameters were adjusted required substantial computation time. Because the design criteria for the module had to be modified to account for the magnetic force produced by the motor and for the hemodynamic changes associated with pump operation, a simplified analytic expression was derived for the expected magnetic forces. Suitable bearings could then be designed capable of overcoming these forces without repeating the complicated FEM simulation for the motor. Using this method at the initial design stage can inform the design of the miniature maglev BLDC motor for the proposed pulsatile axial-flow VAD.

[1]  William A. Smith,et al.  Short-term in vivo performance of the Cleveland clinic PediPump left ventricular assist device. , 2014, Artificial organs.

[2]  H.-G. Herzog,et al.  System design and comparison of calculated and measured performance of a bearing less BLDC-drive with axial flux path for an implantable blood pump , 2006, International Symposium on Power Electronics, Electrical Drives, Automation and Motion, 2006. SPEEDAM 2006..

[3]  James F. Antaki,et al.  Computational Fluid Dynamics-Based Design Optimization for an Implantable Miniature Maglev Pediatric Ventricular Assist Device , 2012 .

[4]  W. Hofmann,et al.  Analysis of radial forces based on rotor eccentricity of bearingless switched reluctance motors , 2010, The XIX International Conference on Electrical Machines - ICEM 2010.

[5]  Mark S. Slaughter,et al.  Outcomes of Left Ventricular Assist Device Implantation as Destination Therapy in the Post-REMATCH Era: Implications for Patient Selection , 2007, Circulation.

[6]  Jeongho Kim,et al.  PediaFlow™ Maglev Ventricular Assist Device: A Prescriptive Design Approach , 2010, Cardiovascular engineering.

[7]  R. Autschbach,et al.  Past, Present, and Future of Long‐Term Mechanical Cardiac Support in Adults , 2008, Journal of cardiac surgery.

[8]  Daniel Timms,et al.  A review of clinical ventricular assist devices. , 2011, Medical engineering & physics.

[9]  M C Oz,et al.  Long-term use of a left ventricular assist device for end-stage heart failure. , 2001, The New England journal of medicine.

[10]  Daniel L. Timms,et al.  In vitro study of an intra-aortic VAD: Effect of reverse-rotating mode on ventricular recovery , 2015, 2015 37th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[11]  Marvin A. Konstam,et al.  Emerging ventricular assist devices for long-term cardiac support , 2010, Nature Reviews Cardiology.

[12]  C. Dotter,et al.  Aortic Length: Angiocardiographic Measurements , 1950, Circulation.

[13]  Yaxin Wang,et al.  In Vitro Evaluation of the Dual-Diffuser Design for a Reversible Rotary Intra-Aortic Ventricular Assist Device. , 2016, Artificial organs.

[14]  Tetsuya Yano,et al.  Passive magnetic bearing in the 3rd generation miniature axial flow pump-the valvo pump 2 , 2015, Journal of Artificial Organs.

[15]  J.-P. Yonnet,et al.  Permanent magnet bearings and couplings , 1981 .

[16]  G. Filippatos,et al.  Elective bridging to recovery after repair: the surgical approach to ventricular reverse remodeling. , 2008, Artificial organs.

[17]  O H Frazier,et al.  Use of a continuous-flow device in patients awaiting heart transplantation. , 2007, The New England journal of medicine.

[18]  Keith A. Seffen,et al.  Noise and vibration in brushless doubly fed machine and brushless doubly fed reluctance machine , 2014 .

[19]  L. Miller,et al.  Experience with the Levitronix CentriMag circulatory support system as a bridge to decision in patients with refractory acute cardiogenic shock and multisystem organ failure. , 2007, The Journal of thoracic and cardiovascular surgery.

[20]  M. Stone Current Status of Mechanical Circulatory Assistance , 2007, Seminars in cardiothoracic and vascular anesthesia.

[21]  Heinrich Schima,et al.  A passive magnetically and hydrodynamically suspended rotary blood pump. , 2009, Artificial organs.

[22]  Ming-Shi Huang,et al.  Design and Implementation of a Magnetically Levitated Single-Axis Controlled Axial Blood Pump , 2009, IEEE Transactions on Industrial Electronics.