Data consistency based translational motion artifact reduction in fan-beam CT

A basic assumption in the classic computed tomography (CT) theory is that an object remains stationary in an entire scan. In biomedical CT/micro-CT, this assumption is often violated. To produce high-resolution images, such as for our recently proposed clinical micro-CT (CMCT) prototype, it is desirable to develop a precise motion estimation and image reconstruction scheme. In this paper, we first extend the Helgason-Ludwig consistency condition (HLCC) from parallel-beam to fan-beam geometry when an object is subject to a translation. Then, we propose a novel method to estimate the motion parameters only from sinograms based on the HLCC. To reconstruct the moving object, we formulate two generalized fan-beam reconstruction methods, which are in filtered backprojection and backprojection filtering formats, respectively. Furthermore, we present numerical simulation results to show that our approach is accurate and robust

[1]  Pierre Grangeat,et al.  Exact reconstruction in 2D dynamic CT: compensation of time-dependent affine deformations. , 2004, Physics in medicine and biology.

[2]  Yoram Bresler,et al.  Uniqueness of tomography with unknown view angles , 2000, IEEE Trans. Image Process..

[3]  S. Leng,et al.  Fan-beam and cone-beam image reconstruction via filtering the backprojection image of differentiated projection data , 2004, Physics in medicine and biology.

[4]  Hengyong Yu,et al.  Design, analysis and simulation for development of the first clinical micro-CT scanner. , 2005, Academic radiology.

[5]  Pierre Grangeat,et al.  Theoretical framework for a dynamic cone-beam reconstruction algorithm based on a dynamic particle model. , 2002, Physics in medicine and biology.

[6]  Ge Wang,et al.  Preliminary study on helical CT algorithms for patient motion estimation and compensation , 1995, IEEE Trans. Medical Imaging.

[7]  Rolf Clackdoyle,et al.  Cone-beam reconstruction using the backprojection of locally filtered projections , 2005, IEEE Transactions on Medical Imaging.

[8]  Sarah K. Patch,et al.  Computation of unmeasured third-generation VCT views from measured views , 2002, IEEE Transactions on Medical Imaging.

[9]  F. Noo,et al.  Cone-beam reconstruction using 1D filtering along the projection of M-lines , 2005 .

[10]  Y Bresler,et al.  Feasibility of tomography with unknown view angles , 2000, IEEE Trans. Image Process..

[11]  Hengyong Yu,et al.  Exact BPF and FBP algorithms for nonstandard saddle curves. , 2005, Medical physics.

[12]  Hengyong Yu,et al.  A unified framework for exact cone-beam reconstruction formulas. , 2005, Medical physics.

[13]  Yongmin Kim,et al.  Correction of computed tomography motion artifacts using pixel-specific back-projection , 1996, IEEE Trans. Medical Imaging.

[14]  Pierre Grangeat,et al.  Dynamic X-ray computed tomography , 2003, Proc. IEEE.

[15]  Norma C. Linney,et al.  Organ motion detection in CT images using opposite rays in fan-beam projection systems , 2001, IEEE Transactions on Medical Imaging.

[16]  Weiguo Lu,et al.  Tomographic motion detection and correction directly in sinogram space. , 2002, Physics in medicine and biology.

[17]  Alexander Katsevich,et al.  An improved exact filtered backprojection algorithm for spiral computed tomography , 2004, Adv. Appl. Math..

[18]  Cameron J. Ritchie,et al.  Respiratory compensation in projection imaging using a magnification and displacement model , 1996, IEEE Trans. Medical Imaging.

[19]  F. Natterer The Mathematics of Computerized Tomography , 1986 .

[20]  Hengyong Yu,et al.  A differentiable Shepp–Logan phantom and its applications in exact cone-beam CT , 2005, Physics in medicine and biology.

[21]  Hengyong Yu,et al.  Feldkamp-type VOI reconstruction from super-short-scan cone-beam data. , 2004, Medical physics.

[22]  Hengyong Yu,et al.  Design, analysis and simulation for development of the first clinical micro-CT scanner1 , 2005 .

[23]  K F King,et al.  Computed tomography scanning with simultaneous patient translation. , 1990, Medical physics.

[24]  S. Patch,et al.  Consistency conditions upon 3D CT data and the wave equation. , 2002, Physics in medicine and biology.

[25]  Y. Liu,et al.  Half-scan cone-beam CT fluoroscopy with multiple x-ray sources. , 2001, Medical physics.

[26]  Hengyong Yu,et al.  Local ROI Reconstruction via Generalized FBP and BPF Algorithms along More Flexible Curves , 2006, Int. J. Biomed. Imaging.

[27]  Hiroyuki Kudo,et al.  Image reconstruction from fan-beam projections on less than a short scan , 2002, Physics in medicine and biology.

[28]  Ge Wang,et al.  Filtered backprojection formula for exact image reconstruction from cone-beam data along a general scanning curve. , 2004, Medical physics.

[29]  Hengyong Yu,et al.  A backprojection-filtration algorithm for nonstandard spiral cone-beam CT with an n-PI-window , 2005, Physics in medicine and biology.

[30]  Hengyong Yu,et al.  A general exact reconstruction for cone-beam CT via backprojection-filtration , 2005, IEEE Transactions on Medical Imaging.

[31]  Alexander Katsevich,et al.  Theoretically Exact Filtered Backprojection-Type Inversion Algorithm for Spiral CT , 2002, SIAM J. Appl. Math..

[32]  Hengyong Yu,et al.  Exact reconstruction for cone-beam scanning along nonstandard spirals and other curves , 2004, SPIE Optics + Photonics.

[33]  Hengyong Yu,et al.  Studies on implementation of the Katsevich algorithm for spiral cone-beam CT , 2004 .

[34]  Guang-Hong Chen,et al.  A new data consistency condition for fan-beam projection data. , 2005, Medical physics.

[35]  Jiang Hsieh,et al.  Computed Tomography: Principles, Design, Artifacts, and Recent Advances, Fourth Edition , 2022 .

[36]  Xiaochuan Pan,et al.  Exact image reconstruction on PI-lines from minimum data in helical cone-beam CT. , 2004, Physics in medicine and biology.

[37]  Franz Kainberger,et al.  Quantification and Clinical Relevance of Head Motion During Computed Tomography , 2003, Investigative radiology.