Three-Degree-of-Freedom MR-Compatible Multisegment Cardiac Catheter Steering Mechanism

This paper presents a novel MR-compatible 3-DOF cardiac catheter steering mechanism. The catheter's steerable structure is tendon driven and consists of miniature deflectable, helical segments created by a precise rapid prototyping technique. The created catheter prototype has an outer diameter of 9 Fr (3 mm) and a steerable distal end that can be deflected in a 3-D space via four braided high-tensile Spectra fiber tendons. Any longitudinal twist commonly observed in helical structures is compensated for by employing clockwise (CW) and counter clockwise (CCW) helical segments in an alternating fashion. A 280 μm flexible carbon fiber rod is used as a backbone in a central channel to improve the structure's steering and positioning repeatability. In addition to the backbone, a carbon fiber tube can be inserted into the structure to a varying amount capable of changing the structure's forcibility and, thus, providing a means to change the curvature and to modify the deflectable length of the catheter leading to an extension of reachable points in the catheter-tip workspace. A unique feature of this helical segment structure is that the stiffness can be further adjusted by appropriately tensioning tendons simultaneously. An experimental study has been conducted examining the catheter-tip trajectory in a 3-D space and its positioning repeatability using a 5-DOF magnetic coil tracking system. Furthermore, MRI experiments in a 1.5-T scanner confirmed the MR-compatibility of the catheter prototype. The study shows that the proposed concept for catheter steering has great potential to be employed for robotically steered and MR-guided cardiac catheterization.

[1]  Aravinda Thiagalingam,et al.  Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force‐Sensing Catheter , 2010, Journal of cardiovascular electrophysiology.

[2]  J. Driller,et al.  Kinetics of magnetically guided catheters , 1970 .

[3]  Jan Peirs,et al.  A micro optical force sensor for force feedback during minimally invasive robotic surgery , 2003 .

[4]  Reza Razavi,et al.  In vivo evaluation and proof of radiofrequency safety of a novel diagnostic MR‐electrophysiology catheter , 2011, Magnetic resonance in medicine.

[5]  Tobias Schaeffter,et al.  Performance Evaluation of an MRI-Compatible Pre-Clinical PET System Using Long Optical Fibers , 2010, IEEE Transactions on Nuclear Science.

[6]  M. Bock,et al.  MR‐guided intravascular interventions: Techniques and applications , 2008, Journal of magnetic resonance imaging : JMRI.

[7]  D. Ortendahl,et al.  Measuring signal-to-noise ratios in MR imaging. , 1989, Radiology.

[8]  Reza Razavi,et al.  MR imaging-guided cardiovascular interventions in young children. , 2012, Magnetic resonance imaging clinics of North America.

[9]  Nicholas S. Peters,et al.  Experience of robotic catheter ablation in humans using a novel remotely steerable catheter sheath , 2008, Journal of Interventional Cardiac Electrophysiology.

[10]  Maggie Fung,et al.  Electroanatomic Mapping of the Left Ventricle in a Porcine Model of Chronic Myocardial Infarction With Magnetic Resonance–Based Catheter Tracking , 2008, Circulation.

[11]  Pierre E. Dupont,et al.  Design and Control of Concentric-Tube Robots , 2010, IEEE Transactions on Robotics.

[12]  C J Bakker,et al.  Heating Around Intravascular Guidewires by Resonating RF Waves , 2000, Journal of magnetic resonance imaging : JMRI.

[13]  R. Lederman Cardiovascular Interventional Magnetic Resonance Imaging , 2005, Circulation.

[14]  D. Hill,et al.  Cardiac catheterisation guided by MRI in children and adults with congenital heart disease , 2003, The Lancet.

[15]  Nadir Saoudi,et al.  The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. , 2012, Heart rhythm.

[16]  Tobias Schaeffter,et al.  Magnetic resonance imaging and spectroscopy. , 2008, Handbook of experimental pharmacology.

[17]  Kaspar Althoefer,et al.  MRI-Compatible Intensity-Modulated Force Sensor for Cardiac Catheterization Procedures , 2011, IEEE Transactions on Biomedical Engineering.

[18]  Rajiv Gupta,et al.  Multi-turn, tension-stiffening catheter navigation system , 2010, 2010 IEEE International Conference on Robotics and Automation.