Detection of (Reversible) Myocardial Ischemic Injury by Means of Electrical Bioimpedance

The scope of this paper was to determine whether ischemic and reperfusion damage in cardiac surgery can be detected by measurement of electrical bioimpedance (EBI). Conventional pacing wires were replaced by pacing wires with sputtered iridium coating in order to reduce polarization associated with two-electrode impedance measurements. A custom-built bioimpedance analyzer (Osypka Medical GmbH, Berlin, Germany) measured the real part of impedance Re( Z) and the phase (φ) at three frequencies (1, 10, and 1000 kHz) and determined an extracellular space index (EZRI) as the quotient of Re( Z) at 1000 kHz and Re(Z) at 1 kHz. Our study included six patients (conventional coronary artery bypass graft, age 68.1 ± 8.3 years) subject to routine cardioplegic ischemia and reperfusion. Preischemic bioimpedance measurements were not impaired by interference of the beating heart. Intraischemically, bioimpedance at 1 kHz and phase at 10 kHz increased until opening of a bypass graft, which is probably induced by closure of gap junctions and cell swelling processes. After cross clamping, EZRI slowly decreased as an effect of mild cell swelling. After ischemia, values returned almost to baseline measurements, indicating sufficient reperfusion processes. Measurement of EBI correlates with myocardial ischemic injury and is applicable in a two-electrode setup providing low-polarization pacing wires.

[1]  A. Welz,et al.  Evaluation of Fractal‐Coated Temporary Pacing Leads in the Early Postoperative Course , 2008, Pacing and clinical electrophysiology : PACE.

[2]  R. Dzwonczyk,et al.  An Evaluation of a New Two-Electrode Myocardial Electrical Impedance Monitor for Detecting Myocardial Ischemia , 2001, Anesthesia and analgesia.

[3]  T Dudykevych,et al.  Impedance analyser module for EIT and spectroscopy using undersampling. , 2001, Physiological measurement.

[4]  J. Lowe,et al.  On-line detection of reversible myocardial ischemic injury by measurement of myocardial electrical impedance. , 1987, The Annals of thoracic surgery.

[5]  E. Gersing,et al.  Messung der elektrischen Impedanz von Organen — Apparative Ausrüstung für Forschung und klinische Anwendung - Measurement of Electrical Impedance in Organs – Measuring Equipment for Research and Clinical Applications , 1991 .

[6]  R. Villa,et al.  Minimally invasive silicon probe for electrical impedance measurements in small animals. , 2003, Biosensors & bioelectronics.

[7]  Roger Dzwonczyk,et al.  Early time course of myocardial electrical impedance during acute coronary artery occlusion in pigs, dogs, and humans. , 2005, Journal of applied physiology.

[8]  H. Fricke,et al.  A Mathematical Treatment of the Electric Conductivity and Capacity of Disperse Systems ii. The Capacity of a Suspension of Conducting Spheroids Surrounded by a Non-Conducting Membrane for a Current of Low Frequency , 1925 .

[9]  E. Gersing,et al.  Tissue Capacitance Measurements at 10 Hz for Detection of Cellular Uncoupling , 1999 .

[10]  H. Fricke,et al.  THE ELECTRIC RESISTANCE AND CAPACITY OF BLOOD FOR FREQUENCIES BETWEEN 800 AND 4½ MILLION CYCLES , 1925, The Journal of general physiology.

[11]  R. Eisenberg,et al.  Impedance Measurements as Estimators of the Properties of the Extracellular Space , 1986, Annals of the New York Academy of Sciences.

[12]  J. Cinca,et al.  Changes in myocardial electrical impedance induced by coronary artery occlusion in pigs with and without preconditioning: correlation with local ST-segment potential and ventricular arrhythmias. , 1997, Circulation.

[13]  K R Foster,et al.  Myocardial Electrical Impedance Mapping of Ischemic Sheep Hearts and Healing Aneurysms , 1993, Circulation.

[14]  M M Gebhard,et al.  The complex dielectric spectrum of heart tissue during ischemia. , 2002, Bioelectrochemistry.

[15]  E Gersing,et al.  Impedance Spectroscopy: a Method for Surveillance of Ischemia Tolerance of the Heart , 1987, The Thoracic and cardiovascular surgeon.

[16]  W. Cascio,et al.  Electrical Properties and Conduction in Reperfused Papillary Muscle , 2001, Circulation research.

[17]  E. Gersing Impedance spectroscopy on living tissue for determination of the state of organs , 1998 .

[18]  Carlos del-Rio,et al.  Myocardial electrical impedance responds to ischemia and reperfusion in humans , 2004, IEEE Transactions on Biomedical Engineering.

[19]  E Gersing [Measuring electrical impedance of organs--instrumental equipment for research and clinical use]. , 1991, Biomedizinische Technik. Biomedical engineering.