Temperature changes caused by MR imaging of the brain with a head coil.

Tissue heating caused by exposure to RF radiation is a primary safety concern in MR imaging. Therefore, to determine temperature changes caused by high field strength MR imaging of the brain with a head coil, we measured body and skin temperatures in 35 patients immediately before and after clinical MR imaging. MR imaging was performed with a 1.5 T MR system using a 28-cm, open-bore RF transmit/receive head coil specifically designed for examinations of the brain. The average body temperature was 36.6 +/- 0.2 degrees C before MR imaging and 36.6 +/- 0.2 degrees C immediately afterward (mean +/- SD, p = not significant). The average forehead skin temperature increased from 32.6 +/- 0.6 to 32.8 +/- 0.5 degrees C (p less than .01), and the average outer canthus skin temperature increased from 32.1 +/- 0.6 to 32.7 +/- 0.6 degrees C (p less than .01) after MR imaging. The highest skin temperature recorded was 34.2 degrees C, and the largest temperature change was +2.1 degrees There were no statistically significant changes in the average skin temperatures of the upper arm and hand. We conclude that patients undergoing MR imaging of the brain with a head coil at the RF radiation exposure we studied experience no significant changes in average body temperature and statistically significant increases in local (i.e., areas within the head coil) skin temperatures. The observed elevations in skin temperatures were physiologically inconsequential.

[1]  J. Hickam,et al.  CARDIOVASCULAR REACTIONS TO EMOTIONAL STIMULI. EFFECT ON THE CARDIAC OUTPUT, ARTERIOVENOUS OXYGEN DIFFERENCE, ARTERIAL PRESSURE, AND PERIPHERAL RESISTANCE. , 1971, The Journal of clinical investigation.

[2]  H. Wolff,et al.  Pain Sensations and Reactions , 1953, Neurology.

[3]  W. I. Cranston,et al.  Oral, rectal, and oesophageal temperatures in relation to central temperature control in man. , 1954, Clinical science.

[4]  B. J. Winer Statistical Principles in Experimental Design , 1992 .

[5]  Beischer De,et al.  INFLUENCE OF STRONG MAGNETIC FIELDS ON THE ELECTROCARDIOGRAM OF SQUIRREL MONKEYS (SAIMIRI SCIUREUS). PROJECT MR005.13-9010, SUBTASK 1, REP NO. 8. , 1964 .

[6]  INFLUENCE OF STRONG MAGNETIC FIELDS ON THE ELECTROCARDIOGRAM OF SQUIRREL MONKEYS (SAIMIRI SCIUREUS). , 1964, Aerospace medicine.

[7]  V A Vardanian,et al.  [Effect of a magnetic field on blood flow]. , 1973, Biofizika.

[8]  J. Bligh Temperature regulation in mammals and other vertebrates , 1973 .

[9]  Vardanian Va Effect of a magnetic field on blood flow , 1973 .

[10]  R Reis,et al.  Potential interference with medical electronic devices. , 1979, Bulletin of the New York Academy of Medicine.

[11]  R. Spiegel,et al.  A thermal model of the human body exposed to an electromagnetic field. , 1980, Bioelectromagnetics.

[12]  T F Budinger,et al.  Nuclear Magnetic Resonance (NMR) In Vivo Studies: Known Thresholds for Health Effects , 1981, Journal of computer assisted tomography.

[13]  L. Rowell,et al.  Cardiovascular aspects of human thermoregulation. , 1983, Circulation research.

[14]  R. D. Saunders,et al.  Safety aspects of NMR clinical imaging. , 1984, British medical bulletin.

[15]  E. Adair,et al.  On the thermoregulatory consequences of NMR imaging. , 1986, Magnetic resonance imaging.

[16]  Shellock Fg,et al.  Monitoring during MRI. An evaluation of the effect of high-field MRI on various patient monitors. , 1986 .

[17]  D B Plewes,et al.  Physiologic changes during high field strength MR imaging. , 1987, AJR. American journal of roentgenology.

[18]  F. Shellock,et al.  Temperature, heart rate, and blood pressure changes associated with clinical MR imaging at 1.5 T. , 1987, Radiology.

[19]  F G Shellock,et al.  Biological effects and safety aspects of magnetic resonance imaging. , 1989, Magnetic resonance quarterly.