Parallel coil resonators for time-domain radiofrequency electron paramagnetic resonance imaging of biological objects.

Resonators suitable for time-domain electron paramagnetic resonance spectroscopy and imaging at a radiofrequency capable of accommodating experimental animals such as mice are described. Design considerations included B(1) field homogeneity, optimal Q, spectral bandwidth, resonator ring-down, and sensitivity. Typically, a resonator with 25-mm diameter and 25-mm length was constructed by coupling 11 single loops in parallel with a separation of 2.5 mm. To minimize the resonator ring-down time and provide the necessary spectral bandwidth for in vivo imaging experiments, the Q was reduced predominantly by overcoupling. Capacitative coupling was utilized to minimize microphonic effects. The B(1) field in the resonator was mapped both radially and axially and found to be uniform and adequate for imaging studies. Imaging studies with phantom objects containing a narrow-line spin probe as well as in vivo objects administered with the spin probe show the suitability of these resonators for valid reproduction of the spin probe distribution in three dimensions. The fabrication of such resonators is simple and can be scaled up with relative ease to accommodate larger objects as well.

[1]  H J Halpern,et al.  Oxymetry deep in tissues with low-frequency electron paramagnetic resonance. , 1994, Proceedings of the National Academy of Sciences of the United States of America.

[2]  P. Lauterbur,et al.  Image Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic Resonance , 1973, Nature.

[3]  M. Bowman,et al.  Modern pulsed and continuous-wave electron spin resonance , 1990 .

[4]  J. S. Hyde,et al.  ESR studies of O2 uptake by Chinese hamster ovary cells during the cell cycle. , 1982, Proceedings of the National Academy of Sciences of the United States of America.

[5]  Sankaran Subramanian,et al.  Radiofrequency FT EPR Spectroscopy and Imaging , 1993 .

[6]  H. Hirata,et al.  A surface-coil-type resonator for in vivo ESR measurements. , 1994, Journal of magnetic resonance. Series B.

[7]  James B. Mitchell,et al.  A broadband pulsed radio frequency electron paramagnetic resonance spectrometer for biological applications , 1998 .

[8]  Alecci,et al.  A Radiofrequency (220-MHz) Fourier Transform EPR Spectrometer , 1998, Journal of magnetic resonance.

[9]  J. Zweier,et al.  High resolution electron paramagnetic resonance imaging of biological samples with a single line paramagnetic label , 1997, Magnetic resonance in medicine.

[10]  H. Togashi,et al.  Spatiotemporal measurement of free radical elimination in the abdomen using an in vivo ESR-CT imaging system. , 1998, Free radical biology & medicine.

[11]  J S Petersson,et al.  EPR and DNP properties of certain novel single electron contrast agents intended for oximetric imaging. , 1998, Journal of magnetic resonance.

[12]  Giuseppe Placidi,et al.  A Submicrosecond Resonator and Receiver System for Pulsed Magnetic Resonance with Large Samples , 1998 .

[13]  G. Chingas Overcoupling NMR probes to improve transient response , 1983 .

[14]  J. Freed,et al.  Two‐dimensional Fourier transform ESR correlation spectroscopy , 1988 .

[15]  J. Hyde,et al.  The loop-gap resonator: a new microwave lumped circuit ESR sample structure , 1982 .

[16]  J. Zweier,et al.  Three-dimensional spectral-spatial EPR imaging of free radicals in the heart: a technique for imaging tissue metabolism and oxygenation. , 1994, Proceedings of the National Academy of Sciences of the United States of America.

[17]  J. Berliner,et al.  Magnetic resonance imaging of biological specimens by electron paramagnetic resonance of nitroxide spin labels. , 1985, Science.

[18]  J. Zweier,et al.  Development of a highly sensitive probe for measuring oxygen in biological tissues. , 1994, Journal of magnetic resonance. Series B.

[19]  R Murugesan,et al.  High-speed data acquisition system and receiver configurations for time-domain radiofrequency electron paramagnetic resonance spectroscopy and imaging. , 1999, Journal of magnetic resonance.

[20]  H. Utsumi,et al.  In vivo ESR measurement of free radicals in whole mice. , 1990, Biochemical and biophysical research communications.

[21]  Howard J. Halpern,et al.  Imaging radio frequency electron‐spin‐resonance spectrometer with high resolution and sensitivity for in vivo measurements , 1989 .

[22]  H. Swartz,et al.  Lithium phthalocyanine: a probe for electron paramagnetic resonance oximetry in viable biological systems. , 1993, Proceedings of the National Academy of Sciences of the United States of America.

[23]  NMR coils with segments in parallel to achieve higher frequencies or larger sample volumes , 1984 .

[24]  Marcello Alecci,et al.  Electron paramagnetic resonance spectrometer for three-dimensional in vivo imaging at very low frequency , 1992 .

[25]  P. Callaghan,et al.  Three-Dimensional Pulsed ESR Imaging , 1996 .

[26]  R Murugesan,et al.  In vivo imaging of a stable paramagnetic probe by pulsed‐radiofrequency electron paramagnetic resonance spectroscopy , 1997, Magnetic resonance in medicine.

[27]  B. Halliwell,et al.  Free radicals in biology and medicine , 1985 .

[28]  W. Froncisz,et al.  Relative Benefits of Overcoupled Resonators vs Inherently Low-Q Resonators for Pulsed Magnetic Resonance , 1994 .

[29]  R. Demeure,et al.  Small particles of fusinite and carbohydrate chars coated with aqueous soluble polymers: Preparation and applications for in Vivo EPR oximetry , 1998, Magnetic resonance in medicine.

[30]  G. Ehnholm,et al.  Overhauser-enhanced MR imaging (OMRI) , 1998, Acta radiologica.

[31]  J. Hyde,et al.  General method for adjusting the quality factor of EPR resonators , 1995 .

[32]  J. Dunn,et al.  Changes of oxygen tension in experimental tumors after a single dose of X-ray irradiation. , 1995, Cancer research.