Technical Report: TG-142 compliant and comprehensive quality assurance tests for respiratory gating.

PURPOSE To develop and establish a comprehensive gating commissioning and quality assurance procedure in compliance with TG-142. METHODS Eight Varian TrueBeam Linacs were used for this study. Gating commissioning included an end-to-end test and baseline establishment. The end-to-end test was performed using a CIRS dynamic thoracic phantom with a moving cylinder inside the lung, which was used for carrying both optically simulated luminescence detectors (OSLDs) and Gafchromic EBT2 films while the target is moving, for a point dose check and 2D profile check. In addition, baselines were established for beam-on temporal delay and calibration of the surrogate, for both megavoltage (MV) and kilovoltage (kV) beams. A motion simulation device (MotionSim) was used to provide periodic motion on a platform, in synchronizing with a surrogate motion. The overall accuracy and uncertainties were analyzed and compared. RESULTS The OSLD readings were within 5% compared to the planned dose (within measurement uncertainty) for both phase and amplitude gated deliveries. Film results showed less than 3% agreement to the predicted dose with a standard sinusoid motion. The gate-on temporal accuracy was averaged at 139±10 ms for MV beams and 92±11 ms for kV beams. The temporal delay of the surrogate motion depends on the motion speed and was averaged at 54.6±3.1 ms for slow, 24.9±2.9 ms for intermediate, and 23.0±20.1 ms for fast speed. CONCLUSIONS A comprehensive gating commissioning procedure was introduced for verifying the output accuracy and establishing the temporal accuracy baselines with respiratory gating. The baselines are needed for routine quality assurance tests, as suggested by TG-142.

[1]  J. Palta,et al.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. , 1994, Medical physics.

[2]  C. Ramsey,et al.  Clinical efficacy of respiratory gated conformal radiation therapy. , 1999, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[3]  Fang-Fang Yin,et al.  Time delay measurement for linac based treatment delivery in synchronized respiratory gating radiotherapy. , 2005, Medical physics.

[4]  Wendy L Smith,et al.  Measurement of time delays in gated radiotherapy for realistic respiratory motions. , 2014, Medical physics.

[5]  R Wiersma,et al.  SU-E-J-126: Respiratory Gating Quality Assurance: A Simple Method to Achieve Millisecond Temporal Resolution. , 2014, Medical physics.

[6]  Fang-Fang Yin,et al.  Task Group 142 report: quality assurance of medical accelerators. , 2009, Medical physics.

[7]  Lei Xing,et al.  Dose verification for respiratory-gated volumetric modulated arc therapy. , 2011, Physics in medicine and biology.

[8]  Steve B. Jiang,et al.  Quality assurance challenges for motion-adaptive radiation therapy: gating, breath holding, and four-dimensional computed tomography. , 2008, International journal of radiation oncology, biology, physics.

[9]  Hiroki Shirato,et al.  Dynamic gating window for compensation of baseline shift in respiratory-gated radiation therapy. , 2011, Medical physics.

[10]  R Mohan,et al.  Determining parameters for respiration-gated radiotherapy. , 2001, Medical physics.

[11]  Steve B. Jiang,et al.  Technical aspects of image-guided respiration-gated radiation therapy. , 2006, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[12]  Wendy L. Smith,et al.  Time delays and margins in gated radiotherapy , 2009, Journal of applied clinical medical physics.

[13]  Paul Keall,et al.  The clinical implementation of respiratory-gated intensity-modulated radiotherapy. , 2006, Medical dosimetry : official journal of the American Association of Medical Dosimetrists.

[14]  John E. Bayouth,et al.  TU‐C‐ValB‐07: Evaluation of 4DRT: CT Acquisition and Gated Delivery System , 2006 .

[15]  T Haberer,et al.  Gating delays for two respiratory motion sensors in scanned particle radiation therapy , 2013, Physics in medicine and biology.

[16]  Arpita Agarwal,et al.  Performance evaluation of respiratory motion‐synchronized dynamic IMRT delivery , 2013, Journal of applied clinical medical physics.

[17]  Steve B. Jiang,et al.  The management of respiratory motion in radiation oncology report of AAPM Task Group 76. , 2006, Medical physics.

[18]  P. Jursinic,et al.  Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements. , 2007, Medical physics.

[19]  Paul J Keall,et al.  First demonstration of combined kV/MV image-guided real-time dynamic multileaf-collimator target tracking. , 2009, International journal of radiation oncology, biology, physics.

[20]  Steve B Jiang,et al.  A respiratory-gated treatment system for proton therapy. , 2007 .