Multichannel Reflective PPG Earpiece Sensor With Passive Motion Cancellation

This paper addresses the design considerations of a novel earpiece photoplethymograph (PPG) sensor and its in-situ evaluation results. The device is encapsulated with multiple LEDs and photodiodes based on a reflective PPG design. A compact and low power circuitry was developed for signal control and conditioning. PPG signals with an averaged ac/dc ratio of 0.001-0.01 and 10% relative strength (compared to finger-based approach) were recorded from the superior and posterior auricular skins. The integrity of PPG signal and accuracy of heart rate detection were evaluated and the results showed that with adequate optical shielding and the proposed passive motion cancellation, the device was able to reliably detect heart rate both during rest and moderate exercise. The proposed sensor design is low power, easy to wear compared to conventional earlobe PPG devices.

[1]  Y. Mendelson,et al.  Signal quality and power consumption of a new prototype reflectance pulse oximeter sensor , 2005, Proceedings of the IEEE 31st Annual Northeast Bioengineering Conference, 2005..

[2]  P. Gibbs,et al.  Reducing motion artifact in wearable bio-sensors using MEMS accelerometers for active noise cancellation , 2005, Proceedings of the 2005, American Control Conference, 2005..

[3]  P. Kyriacou Pulse oximetry in the oesophagus , 2006, Physiological measurement.

[4]  H.H. Asada,et al.  Active Motion Artifact Reduction for Wearable Sensors Using Laguerre Expansion and Signal Separation , 2005, 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference.

[5]  Philippe Renevey,et al.  Wrist-located pulse detection using IR signals, activity and nonlinear artifact cancellation , 2001, 2001 Conference Proceedings of the 23rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[6]  Y. Mendelson,et al.  Measurement site and photodetector size considerations in optimizing power consumption of a wearable reflectance pulse oximeter , 2003, Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (IEEE Cat. No.03CH37439).

[7]  M. Nogawa,et al.  Development of a tissue oxygen consumption measurement method based on near-infrared photoplethysmography , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[8]  Liliana Grajales,et al.  Wearable multisensor heart rate monitor , 2006, International Workshop on Wearable and Implantable Body Sensor Networks (BSN'06).

[9]  A. B. Hertzman THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH , 1938 .

[10]  Adam W. M. Mitchell,et al.  Gray's Anatomy for Students , 2004 .

[11]  Yuan-ting Zhang,et al.  Reduction of motion artifact in pulse oximetry by smoothed pseudo Wigner-Ville distribution , 2005, Journal of NeuroEngineering and Rehabilitation.

[12]  Seungjoon Lee,et al.  Processing of pulse oximeter data using discrete wavelet analysis , 2005, IEEE Transactions on Biomedical Engineering.

[13]  James L. Reuss,et al.  Multilayer modeling of reflectance pulse oximetry , 2005, IEEE Transactions on Biomedical Engineering.

[14]  H. Harry Asada,et al.  Artifact-resistant power-efficient design of finger-ring plethysmographic sensors , 2001, IEEE Transactions on Biomedical Engineering.

[15]  Tomas E. Ward,et al.  The Design And Clinical Use Of A Reflective Brachial Photoplethysmograph , 2002 .

[16]  M. J. Hayes,et al.  Artefact reduction in photoplethysmography , 1998 .

[17]  Katsunori Ikoma,et al.  Obituary: Yukio Mano (1943–2004) , 2005, Journal of NeuroEngineering and Rehabilitation.

[18]  J. Spigulis Optical noninvasive monitoring of skin blood pulsations , 2005 .

[19]  John G. Webster,et al.  Design of Pulse Oximeters , 1997 .

[20]  Youngil Kim,et al.  Design of filter to reject motion artifact of pulse oximetry , 2004, Comput. Stand. Interfaces.

[21]  Yuan-Ting Zhang,et al.  A near-infrared heart rate measurement IC with very low cutoff frequency using current steering technique , 2005, IEEE Transactions on Circuits and Systems I: Regular Papers.