The effect of refraction and assumed speeds of sound in tissue and blood on Doppler ultrasound blood velocity measurements.

The combined effect of three assumptions relating to refraction, the speed of sound in tissue and the speed of sound in blood on the accuracy of Doppler ultrasound blood velocity measurements has been investigated. A theoretical relationship giving the net velocity measurement error introduced by these three assumptions has derived using a model in which tissue and blood layers are separated by straight, parallel boundaries. This net error is dependent on the assumed and actual speed of sound in tissue, the assumed speed of sound in blood and the Doppler angle, but is effectively independent of the actual speed of sound in blood. For clinical blood velocity measurements, the net error is estimated to be as much as an 8% overestimation of the actual velocity, higher than previously predicted for any of the factors individually. The relationship also predicts a net velocity measurement error in experimental flow systems and string phantoms which is dependent on the speed of sound in the liquid bath. A water bath at room temperature will give an overestimation of approximately 2%. Experimental investigations using conventional and modified string phantoms and a 5-MHz linear phased array system support these conclusions. The effect of perturbing the layers from their parallel orientation has also been considered theoretically and has provided additional support for the above conclusions. These results may help more accurate Doppler velocity measurements in both experimental and clinical settings.

[1]  C. Oates The Doppler shift & speed of sound in blood , 1989 .

[2]  R. Gill Pulsed Doppler with B-mode imaging for quantitative blood flow measurement. , 1979, Ultrasound in medicine & biology.

[3]  F. Foster,et al.  Quantitative contrast measurements in B-mode images comparison between experiment and theory. , 1986, Ultrasound in medicine & biology.

[4]  R. Cobbold,et al.  Quantitative study of steady flow using color Doppler ultrasound. , 1991, Ultrasound in medicine & biology.

[5]  A Goldstein,et al.  Ethylene glycol‐water mixtures for use in ultrasound test objects , 1979, Journal of clinical ultrasound : JCU.

[6]  J. Liu,et al.  The effect of frequency dependent scattering and attenuation on the estimation of blood velocity using ultrasound , 1992, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[7]  A Goldstein Effect of tank liquid acoustic velocity on Doppler string phantom measurements. , 1991, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[8]  Vernon L. Newhouse,et al.  Analysis of Transit Time Effects on Doppler Flow Measurement , 1976, IEEE Transactions on Biomedical Engineering.

[9]  P. Veltink,et al.  Influence of Beam Profile and Degree of Insonation on the CW Doppler Ultrasound Spectrum and Mean Velocity , 1983, IEEE Transactions on Sonics and Ultrasonics.

[10]  F. Foster,et al.  Ultrasonic characterization of selected renal tissues. , 1989, Ultrasound in medicine & biology.

[11]  C. Oates,et al.  The use of a Diasonics DRF400 duplex ultrasound scanner to measure volume flow in arterio-venous fistulae in patients undergoing haemodialysis: an analysis of measurement uncertainties. , 1990, Ultrasound in medicine & biology.

[12]  R S Cobbold,et al.  Influence of spectral broadening on continuous wave Doppler ultrasound spectra: a geometric approach. , 1986, Ultrasound in medicine & biology.

[13]  Stelios C. Orphanoudakis,et al.  Frequency-Dependent Attenuation Effects in Pulsed Doppler Ultrasound: Experimental Results , 1984, IEEE Transactions on Biomedical Engineering.

[14]  D H Evans,et al.  Some aspects of the relationship between instantaneous volumetric blood flow and continuous wave Doppler ultrasound recordings--I. The effect of ultrasonic beam width on the output of maximum, mean and RMS frequency processors. , 1982, Ultrasound in medicine & biology.

[15]  W N McDicken,et al.  A versatile test-object for the calibration of ultrasonic Doppler flow instruments. , 1986, Ultrasound in medicine & biology.

[16]  R W Gill,et al.  Performance of the mean frequency Doppler modulator. , 1979, Ultrasound in medicine & biology.

[17]  D. Evans Doppler Ultrasound: Physics Instrumentation and Clinical Applications , 1989 .

[18]  W. McDicken,et al.  Refraction in Doppler ultrasound. , 1993, Ultrasound in medicine & biology.

[19]  R. Gill Measurement of blood flow by ultrasound: accuracy and sources of error. , 1985, Ultrasound in medicine & biology.

[20]  L. William Varner,et al.  Geometrical Spectrum Broadening in Ultrasonic Doppler Systems , 1977, IEEE Transactions on Biomedical Engineering.

[21]  Jaffe Cc,et al.  Quantitative flow measurements with Doppler ultrasound: techniques, accuracy, and limitations. , 1985 .

[22]  D H Evans,et al.  On the measurement of the mean velocity of blood flow over the cardiac cycle using Doppler ultrasound. , 1985, Ultrasound in medicine & biology.

[23]  A Goldstein,et al.  Performance tests of Doppler ultrasound equipment with a string phantom. , 1991, Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine.

[24]  J. Woodcock,et al.  Doppler ultrasound and its use in clinical measurement , 1983 .

[25]  W.D. O'Brien,et al.  Pulsed Doppler accuracy assessment due to frequency-dependent attenuation and Rayleigh scattering error sources , 1990, IEEE Transactions on Biomedical Engineering.

[26]  F. Kremkau Doppler angle error due to refraction. , 1990, Ultrasound in Medicine and Biology.