Neuronal current detection with low-field magnetic resonance: simulations and methods.

The noninvasive detection of neuronal currents in active brain networks [or direct neuronal imaging (DNI)] by means of nuclear magnetic resonance (NMR) remains a scientific challenge. Many different attempts using NMR scanners with magnetic fields >1 T (high-field NMR) have been made in the past years to detect phase shifts or magnitude changes in the NMR signals. However, the many physiological (i.e., the contemporarily BOLD effect, the weakness of the neuronal-induced magnetic field, etc.) and technical limitations (e.g., the spatial resolution) in observing the weak signals have led to some contradicting results. In contrast, only a few attempts have been made using low-field NMR techniques. As such, this paper was aimed at reviewing two recent developments in this front. The detection schemes discussed in this manuscript, the resonant mechanism (RM) and the DC method, are specific to NMR instrumentations with main fields below the earth magnetic field (50 microT), while some even below a few microteslas (ULF-NMR). However, the experimental validation for both techniques, with differentiating sensitivity to the various neuronal activities at specific temporal and spatial resolutions, is still in progress and requires carefully designed magnetic field sensor technology. Additional care should be taken to ensure a stringent magnetic shield from the ambient magnetic field fluctuations. In this review, we discuss the characteristics and prospect of these two methods in detecting neuronal currents, along with the technical requirements on the instrumentation.

[1]  N. Tepley,et al.  Magnetic fields associated with spreading depression in anaesthetised rabbits , 1991, Brain Research.

[2]  Bruno Maraviglia,et al.  Challenges for detection of neuronal currents by MRI. , 2006, Magnetic resonance imaging.

[3]  Michele Migliore,et al.  Realistic simulations of neuronal activity: A contribution to the debate on direct detection of neuronal currents by MRI , 2008, NeuroImage.

[4]  A. Matlachov,et al.  Simultaneously detected biomagnetic signals and NMR. , 2004, Neurology & clinical neurophysiology : NCN.

[5]  Heidrun Wabnitz,et al.  Dynamics of cortical neurovascular coupling analyzed by simultaneous DC-magnetoencephalography and time-resolved near-infrared spectroscopy , 2008, NeuroImage.

[6]  K. Welch,et al.  Magnetic fields associated with anoxic depolarization in anesthetized rats , 1991, Brain Research.

[7]  Dietmar Drung,et al.  dc Magnetoencephalography: Direct measurement in a magnetically extremely-well shielded room , 2004 .

[8]  Gabriel Curio,et al.  Non-invasive long-term recordings of cortical ‘direct current’ (DC–) activity in humans using magnetoencephalography , 1999, Neuroscience Letters.

[9]  L. Trahms,et al.  SQUID Systems Adapted to Record Nuclear Magnetism in Low Magnetic Fields , 2007, IEEE Transactions on Applied Superconductivity.

[10]  M. Hatridge,et al.  SQUID-detected in vivo MRI at microtesla magnetic fields , 2005, IEEE Transactions on Applied Superconductivity.

[11]  Bernd Lütkenhöner,et al.  Near-DC magnetic fields following a periodic presentation of long-duration tonebursts , 2001, Clinical Neurophysiology.

[12]  Bruno Maraviglia,et al.  Microscopic investigation of the resonant mechanism for the implementation of nc-MRI at ultra-low field MRI , 2008, NeuroImage.

[13]  D. Tank,et al.  Brain magnetic resonance imaging with contrast dependent on blood oxygenation. , 1990, Proceedings of the National Academy of Sciences of the United States of America.

[14]  R. Edelman,et al.  Magnetic resonance imaging (2) , 1993, The New England journal of medicine.

[15]  Michelle A. Espy,et al.  Toward direct neural current imaging by resonant mechanisms at ultra-low field , 2008, NeuroImage.

[16]  John Clarke,et al.  Calculated signal-to-noise ratio of MRI detected with SQUIDs and Faraday detectors in fields from 10 microT to 1.5 T. , 2007, Journal of magnetic resonance.

[17]  N. Logothetis The neural basis of the blood-oxygen-level-dependent functional magnetic resonance imaging signal. , 2002, Philosophical transactions of the Royal Society of London. Series B, Biological sciences.

[18]  W. Haberkorn,et al.  Neuromagnetic Inverse Solutions in Terms of Multipoles , 2001 .

[19]  Andrew R. Mayer,et al.  Modelling the magnetic signature of neuronal tissue , 2007, NeuroImage.

[20]  Dmitry Budker,et al.  Magnetic resonance imaging with an optical atomic magnetometer , 2006, Proceedings of the National Academy of Sciences.

[21]  L. Trahms,et al.  High-resolution magnetoneurography , 1999 .

[22]  J. Bodurka,et al.  Direct detection of neuronal activity with MRI: Fantasy, possibility, or reality? , 2005 .

[23]  Gabriel Curio,et al.  Tonic neuronal activation during simple and complex finger movements analyzed by DC-magnetoencephalography , 2006, Neuroscience Letters.