Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers

IntroductionTo assess potential metabolic and microcirculatory alterations in critically ill patients, near-infrared spectroscopy (NIRS) has been used, in combination with a vascular occlusion test (VOT), for the non-invasive measurement of tissue oxygen saturation (StO2), oxygen consumption, and microvascular reperfusion and reactivity. The methodologies for assessing StO2 during a VOT, however, are very inconsistent in the literature and, consequently, results vary from study to study, making data comparison difficult and potentially inadequate. Two major aspects concerning the inconsistent methodology are measurement site and probe spacing. To address these issues, we investigated the effects of probe spacing and measurement site using 15 mm and 25 mm probe spacings on the thenar and the forearm in healthy volunteers and quantified baseline, ischemic, reperfusion, and hyperemic VOT-derived StO2 variables.MethodsStO2 was non-invasively measured in the forearm and thenar in eight healthy volunteers during 3-minute VOTs using two InSpectra tissue spectrometers equipped with a 15 mm probe or a 25 mm probe. VOT-derived StO2 traces were analyzed for base-line, ischemic, reperfusion, and hyperemic parameters. Data were categorized into four groups: 15 mm probe on the forearm (F15 mm), 25 mm probe on the forearm (F25 mm), 15 mm probe on the thenar (T15 mm), and 25 mm probe on the thenar (T25 mm).ResultsAlthough not apparent at baseline, probe spacing and measurement site significantly influenced VOT-derived StO2 variables. For F15 mm, F25 mm, T15 mm, and T25 mm, StO2 ownslope was -6.4 ± 1.7%/minute, -10.0 ± 3.2%/minute, -12.5 ± 3.0%/minute, and -36.7 ± 4.6%/minute, respectively. StO2 upslope was 105 ± 34%/minute, 158 ± 55%/minute, 226 ± 41%/minute, and 713 ± 101%/minute, and the area under the hyperemic curve was 7.4 ± 3.8%·minute, 10.1 ± 4.9%·minute, 12.6 ± 4.4%·minute, and 21.2 ± 2.7%·minute in these groups, respectively. Furthermore, the StO2 parameters of the hyperemic phase of the VOT, such as the area under the curve, significantly correlated to the minimum StO2 during ischemia.ConclusionsNIRS measurements in combination with a VOT are measurement site-dependent and probe-dependent. Whether this dependence is anatomy-, physiology-, or perhaps technology-related remains to be elucidated. Our study also indicated that reactive hyperemia depends on the extent of ischemic insult.

[1]  Sven Zenker,et al.  Use of non-invasive NIRS during a vascular occlusion test to assess dynamic tissue O2 saturation response , 2008, Intensive Care Medicine.

[2]  M. Podbregar,et al.  Measurement of skeletal muscle tissue oxygenation in the critically ill , 2008 .

[3]  M. Podbregar,et al.  Changes in muscle tissue oxygenation during stagnant ischemia in septic patients , 2005, Intensive Care Medicine.

[4]  Jean-Charles Preiser,et al.  Microvascular blood flow is altered in patients with sepsis. , 2002, American journal of respiratory and critical care medicine.

[5]  Dean E. Myers,et al.  Noninvasive method for measuring local hemoglobin oxygen saturation in tissue using wide gap second derivative near-infrared spectroscopy. , 2005, Journal of biomedical optics.

[6]  M. Fink Cytopathic hypoxia. Mitochondrial dysfunction as mechanism contributing to organ dysfunction in sepsis. , 2001, Critical care clinics.

[7]  Dean E. Myers,et al.  DYNAMIC NEAR-INFRARED SPECTROSCOPY MEASUREMENTS IN PATIENTS WITH SEVERE SEPSIS , 2007, Shock.

[8]  Can Ince,et al.  The microcirculation is the motor of sepsis , 2005, Critical care.

[9]  J. Vincent,et al.  The prognostic value of muscle StO2 in septic patients , 2007, Intensive Care Medicine.

[10]  E. Rackow,et al.  Microvascular function and rheologic changes in hyperdynamic sepsis. , 1995, Critical care medicine.

[11]  C. Ince,et al.  Microcirculatory oxygenation and shunting in sepsis and shock. , 1999, Critical care medicine.

[12]  C. Roussos,et al.  Acute effects of smoking on skeletal muscle microcirculation monitored by near-infrared spectroscopy. , 2007, Chest.

[13]  C. Pierrakos,et al.  Inotropic agents improve the peripheral microcirculation of patients with end-stage chronic heart failure. , 2008, Journal of cardiac failure.

[14]  J. Parrillo,et al.  Early increases in microcirculatory perfusion during protocol-directed resuscitation are associated with reduced multi-organ failure at 24 h in patients with sepsis , 2008, Intensive Care Medicine.

[15]  Marco Mercieri,et al.  Microvascular dysfunction and skeletal muscle oxygenation assessed by phase-modulation near-infrared spectroscopy in patients with septic shock , 2005, Intensive Care Medicine.

[16]  G. Schmidt,et al.  Impairments in microvascular reactivity are related to organ failure in human sepsis. , 2007, American journal of physiology. Heart and circulatory physiology.

[17]  E. Rackow,et al.  Microvascular response in patients with cardiogenic shock , 2000, Critical care medicine.

[18]  Can Ince,et al.  Quantifying bedside-derived imaging of microcirculatory abnormalities in septic patients: a prospective validation study , 2005, Critical care.