Frequency response of real-time digital oscilloscope with timeinterleaving architecture

This paper describes in detail how to characterize the response of a commercial real-time digital oscilloscope (RTDO) with traceability. The commercial RTDO usually employs a time interleaving ADC (TIADC) structure. Thus, aliasing is produced when the measurement is separated according to each ADC. As a result, it is difficult to obtain an accurate frequency response for each ADC. In this paper, we propose a method to avoid under-sampling issue, and it can achieve high sampling rate because it is similar to the equivalent sampling method.

[1]  W. Black,et al.  Time interleaved converter arrays , 1980, 1980 IEEE International Solid-State Circuits Conference. Digest of Technical Papers.

[2]  Dylan F. Williams,et al.  Terminology for high-speed sampling-oscilloscope calibration , 2006, 2006 68th ARFTG Conference: Microwave Measurement.

[3]  Irshaad Fatadin,et al.  Calibration of Wideband Digital Real-Time Oscilloscopes , 2015, IEEE Transactions on Instrumentation and Measurement.

[4]  Joo-Gwang Lee,et al.  Calibration of channel mismatch in time-interleaved real-time digital oscilloscopes , 2015, 2015 85th Microwave Measurement Conference (ARFTG).

[5]  U. Madhow,et al.  Comprehensive digital correction of mismatch errors for a 400-msamples/s 80-dB SFDR time-interleaved analog-to-digital converter , 2005, IEEE Transactions on Microwave Theory and Techniques.

[6]  Jeffrey A. Jargon,et al.  Calibration of Time-Interleaved Errors in Digital Real-Time Oscilloscopes , 2016, IEEE Transactions on Microwave Theory and Techniques.

[7]  Koji Asami An Algorithm to Improve the Performance of M-Channel Time-Interleaved A-D Converters , 2007, IEICE Trans. Fundam. Electron. Commun. Comput. Sci..

[8]  K.A. Remley,et al.  Systematic Error of the Nose-to-Nose Sampling-Oscilloscope Calibration , 2007, IEEE Transactions on Microwave Theory and Techniques.

[9]  Dong-Joon Lee,et al.  Traceable calibration for a digital real-time oscilloscope with time interleaving architecture , 2017 .

[10]  Dylan F. Williams,et al.  Covariance-Based Vector-Network-Analyzer Uncertainty Analysis for Time- and Frequency-Domain Measurements , 2010, IEEE Transactions on Microwave Theory and Techniques.

[11]  Dylan F. Williams,et al.  Physical models for 2.4 mm and 3.5 mm coaxial VNA calibration kits developed within the NIST microwave uncertainty framework , 2015, 2015 85th Microwave Measurement Conference (ARFTG).

[12]  P. Hale,et al.  Covariance-based uncertainty analysis of the NIST electrooptic sampling system , 2006, IEEE transactions on microwave theory and techniques.