Comparison of metal release from various metallic biomaterials in vitro.

To investigate the metal release of each base and alloying elements in vitro, SUS316L stainless steel, Co-Cr-Mo casting alloy, commercially pure Ti grade 2, and Ti-6Al-4V, V-free Ti-6Al-7Nb and Ti-15Zr-4Nb-4Ta alloys were immersed in various solutions, namely, alpha-medium, PBS(-), calf serum, 0.9% NaCl, artificial saliva, 1.2 mass% L-cysteine, 1 mass% lactic acid and 0.01 mass% HCl for 7d. The difference in the quantity of Co released from the Co-Cr-Mo casting alloy was relatively small in all the solutions. The quantities of Ti released into alpha-medium, PBS(-), calf serum, 0.9% NaCl and artificial saliva were much lower than those released into 1.2% L-cysteine, 1% lactic acid and 0.01% HCl. The quantity of Fe released from SUS316L stainless steel decreased linearly with increasing pH. On the other hand, the quantity of Ti released from Ti materials increased with decreasing pH, and it markedly attenuated at pHs of approximately 4 and higher. The quantity of Ni released from stainless steel gradually decreased with increasing pH. The quantities of Al released from the Ti-6Al-4V and Ti-6Al-7Nb alloys gradually decreased with increasing pH. A small V release was observed in calf serum, PBS(-), artificial saliva, 1% lactic acid, 1.2% l-cysteine and 0.01% HCl. The quantity of Ti released from the Ti-15Zr-4Nb-4Ta alloy was smaller than those released from the Ti-6Al-4V and Ti-6Al-7Nb alloys in all the solutions. In particular, it was approximately 30% or smaller in 1% lactic acid, 1.2% L-cysteine and 0.01% HCl. The quantity of (Zr + Nb + Ta) released was also considerably lower than that of (Al + Nb) or (Al + V) released. Therefore, the Ti-15Zr-4Nb-4Ta alloy with its low metal release in vitro is considered advantageous for long-term implants.

[1]  P. Gregson,et al.  Surface modification of titanium alloy implants. , 1994, Biomaterials.

[2]  K. Schaller,et al.  Nickel-, chrom- and cobalt-concentrations in human tissue and body fluids of hip prosthesis patients. , 1992, Journal of trace elements and electrolytes in health and disease.

[3]  S. Steinemann,et al.  Compatibility of Titanium in Soft and Hard Tissue – The Ultimate is Osseointegration , 2005 .

[4]  Y. Okazaki,et al.  Corrosion Resistance of Implant Alloys in Pseudo Physiological Solution and Role of Alloying Elements in Passive Films , 1995 .

[5]  Y. Okazaki,et al.  Effects of Mo and Pd on Corrosion Resistance of V-Free Titanium Alloys for Medical Implants , 1997 .

[6]  J. Davidson,et al.  Mechanical and Tribological Properties and Biocompatibility of Diffusion Hardened Ti-13Nb-13Zr — A New Titanium Alloy for Surgical Implants , 1996 .

[7]  D Buch,et al.  In vitro corrosion of titanium. , 1998, Biomaterials.

[8]  M. Reich,et al.  Systemic effects of implanted prostheses made of cobalt-chromium alloys , 2004, Archives of Orthopaedic and Trauma Surgery.

[9]  K. Wapner,et al.  Implications of metallic corrosion in total knee arthroplasty. , 1991, Clinical orthopaedics and related research.

[10]  D. Williams,et al.  In-vitro corrosion and wear of titanium alloys in the biological environment. , 1996, Biomaterials.

[11]  Y. Okazaki,et al.  Effects of Ti, Al and V Concentrations on Cell Viability , 1998 .

[12]  J. Galante,et al.  Metal release and excretion from cementless titanium alloy total knee replacements. , 1999, Clinical orthopaedics and related research.

[13]  M. Koike,et al.  In Vitro Assessment of Release from Titanium by Immersion Tests. , 1997 .

[14]  Y. Okazaki Effect of friction on anodic polarization properties of metallic biomaterials. , 2002, Biomaterials.

[15]  Y. Okazaki,et al.  In vitro biocompatibility, mechanical properties, and corrosion resistance of Ti-Zr-Nb-Ta-Pd and Ti-Sn-Nb-Ta-Pd alloys. , 1995, Journal of biomedical materials research.

[16]  T. Hurlen,et al.  Passive behaviour of titanium , 1986 .

[17]  M. Imam,et al.  Titanium Alloys as Implant Materials , 1996 .

[18]  J. Dumbleton,et al.  Microstructure and Properties of a New Beta Titanium Alloy, Ti-12Mo-6Zr-2Fe, Developed for Surgical Implants , 1996 .

[19]  Y. Okazaki,et al.  Corrosion resistance, mechanical properties, corrosion fatigue strength and cytocompatibility of new Ti alloys without Al and V. , 1998, Biomaterials.

[20]  R. Sodhi,et al.  Nitric acid passivation of Ti6Al4V reduces thickness of surface oxide layer and increases trace element release. , 1995, Journal of biomedical materials research.

[21]  Y. Okazaki A New Ti–15Zr–4Nb–4Ta alloy for medical applications , 2001 .

[22]  L. Zardiackas,et al.  Characterization of Ti-15Mo Beta Titanium Alloy for Orthopaedic Implant Applications , 1996 .

[23]  J. Helsen,et al.  In vitro simulation of biocompatibility of Ti-Al-V. , 1988, Journal of biomedical materials research.

[24]  H. W. Ruediger,et al.  Increased blood cobalt and chromium after total hip replacement. , 1999, Journal of toxicology. Clinical toxicology.

[25]  E. Collings,et al.  Materials Properties Handbook: Titanium Alloys , 1994 .

[26]  P. Ducheyne,et al.  Passive Dissolution of Titanium in Biological Environments , 1996 .

[27]  Y. Okazaki,et al.  Effect of Metal Released from Ti Alloy Wear Powder on Cell Viability , 2000 .

[28]  S. Hopfer,et al.  Cobalt, chromium, and nickel concentrations in body fluids of patients with porous‐coated knee or hip prostheses , 1989, Journal of orthopaedic research : official publication of the Orthopaedic Research Society.

[29]  Y. Okazaki,et al.  New Titanium Alloys to be Considered for Medical Implants , 1996 .

[30]  S. G. Steinemann,et al.  Corrosion of Surgical Implants-in vivo and in vitro Tests , 1980 .

[31]  J. Galante,et al.  Metal Release in Patients Who Have Had a Primary Total Hip Arthroplasty. A Prospective, Controlled, Longitudinal Study* , 1998, The Journal of bone and joint surgery. American volume.

[32]  P. Ducheyne,et al.  The mechanisms of passive dissolution of titanium in a model physiological environment. , 1992, Journal of biomedical materials research.

[33]  J. Davidson,et al.  Chemical and Electrochemical Aspects of the Biocompatibility of Titanium and its Alloys , 1996 .

[34]  A. Kaider,et al.  Elevated Serum Cobalt With Metal-on-metal Articulating Surfaces , 1997 .

[35]  P. D. Wilson,et al.  Metallic wear in failed titanium-alloy total hip replacements. A histological and quantitative analysis. , 1988, The Journal of bone and joint surgery. American volume.

[36]  J. Galante,et al.  Release and excretion of metal in patients who have a total hip-replacement component made of titanium-base alloy. , 1991, The Journal of bone and joint surgery. American volume.

[37]  Y. Okazaki,et al.  Effects of Metallic Concentrations Other Than Ti, Al and V on Cell Viability , 1998 .

[38]  C. Minoia,et al.  Metal determination in organic fluids of patients with stainless steel hip arthroplasty. , 1983, Acta orthopaedica Scandinavica.

[39]  N. Gjerdet,et al.  Metal release from heat-treated orthodontic archwires. , 1987, Acta odontologica Scandinavica.

[40]  Shih-Ching Wu,et al.  Ion release from NiTi orthodontic wires in artificial saliva with various acidities. , 2003, Biomaterials.

[41]  P J Gregson,et al.  Effect of surface treatment on the dissolution of titanium-based implant materials. , 1991, Biomaterials.

[42]  C. Minoia,et al.  Metal ions in body fluids after arthroplasty. , 1986, Acta orthopaedica Scandinavica.