In Vivo Impedance Characterization of Cortical Recording Electrodes Shows Dependence on Electrode Location and Size

Objective: Neural prostheses are improving the quality of life for those suffering from neurological impairments. Electrocorticography electrodes located in subdural, epidural, and intravascular positions show promise as long-term neural prostheses. However, chronic implantation affects the electrochemical environments of these arrays. Methods: In the present work, the effect of electrode location on the electrochemical properties of the interface was compared. The impedances of the electrode arrays were measured using electrochemical impedance spectroscopy in vitro in saline and in vivo four-week postimplantation. Results: There was not a significant effect of electrode location (subdural, intravascular, or epidural) on the impedance magnitude, and the effect of the electrode size on the impedance magnitude was frequency dependent. There was a frequency-dependent statistically significant effect of electrode location and electrode size on the phase angles of the three arrays. The subdural and epidural arrays showed phase shifts closer to –90° indicating the capacitive nature of the interface in these locations. The impact of placing electrodes within a blood vessel and adjacent to the blood vessel wall was most obvious when looking at the phase responses at frequencies below 10 kHz. Conclusion: Our results show that intravascular electrodes, like those in subdural and epidural positions, show electrical properties that are suitable for recording. These results provide support for the use of intravascular arrays in clinically relevant neural prostheses and diagnostic devices. Significance: Comparison of electrochemical impedance of the epidural, intravascular, and subdural electrode array showed that all three locations are possible placement options, since impedances are in comparable ranges.

[1]  S. Cogan Neural stimulation and recording electrodes. , 2008, Annual review of biomedical engineering.

[2]  Leonid Churilov,et al.  Minimally invasive endovascular stent-electrode array for high-fidelity, chronic recordings of cortical neural activity , 2016, Nature Biotechnology.

[3]  Thomas Stieglitz,et al.  Biomedical Microdevices for Neural Implants , 2006 .

[4]  Krishna V. Shenoy,et al.  Challenges and Opportunities for Next-Generation Intracortically Based Neural Prostheses , 2011, IEEE Transactions on Biomedical Engineering.

[5]  Joaquin Andres Hoffer,et al.  Diaphragm Activation in Ventilated Patients Using a Novel Transvenous Phrenic Nerve Pacing Catheter , 2017, Critical care medicine.

[6]  David B. Grayden,et al.  Feasibility of a chronic, minimally invasive endovascular neural interface , 2016, 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[7]  David B. Grayden,et al.  7T-fMRI: Faster temporal resolution yields optimal BOLD sensitivity for functional network imaging specifically at high spatial resolution , 2018, NeuroImage.

[8]  Matthew D. Johnson,et al.  In vivo impedance spectroscopy of deep brain stimulation electrodes , 2009, Journal of neural engineering.

[9]  David Daomin Zhou,et al.  Implantable Neural Prostheses 2 , 2010 .

[10]  Andrian Sue,et al.  Influence of blood vessel conductivity in cochlear implant stimulation using a finite element head model , 2013, 2013 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC).

[11]  Rodney C. Samaco,et al.  Forniceal deep brain stimulation rescues hippocampal memory in Rett syndrome mice , 2015, Nature.

[12]  C. Gabriel Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies. , 1996 .

[13]  J. Tait,et al.  Challenges and opportunities. , 1996, Journal of psychiatric and mental health nursing.

[14]  Graeme M. Clark,et al.  Testing of thin-film electrode arrays for cochlear implants of the future , 1999 .

[15]  Heather L. Benz,et al.  Neuroprosthetics and the science of patient input , 2017, Experimental Neurology.

[16]  Thomas Stieglitz,et al.  Mechanical characterization of neural electrodes based on PDMS-parylene C-PDMS sandwiched system , 2011, 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[17]  M G Bleichner,et al.  The effects of blood vessels on electrocorticography , 2011, Journal of neural engineering.

[18]  Andrew B Schwartz,et al.  Cortical neural prosthetics. , 2004, Annual review of neuroscience.

[19]  Robert E. Hampson,et al.  A cognitive prosthesis for memory facilitation by closed-loop functional ensemble stimulation of hippocampal neurons in primate brain , 2017, Experimental Neurology.

[20]  Terence J. O'Brien,et al.  Micro-CT and Histological Evaluation of an Neural Interface Implanted Within a Blood Vessel , 2017, IEEE Transactions on Biomedical Engineering.

[21]  Nicholas G Hatsopoulos,et al.  The science of neural interface systems. , 2009, Annual review of neuroscience.

[22]  G J Suaning,et al.  Fabrication of implantable microelectrode arrays by laser cutting of silicone rubber and platinum foil , 2005, Journal of neural engineering.

[23]  B. Roth The Electrical Conductivity of Tissues , 1999 .

[24]  Terence J. O'Brien,et al.  Chronic impedance spectroscopy of an endovascular stent-electrode array , 2016, Journal of neural engineering.

[25]  Brett C. Hannigan,et al.  Mitigation of Ventilator‐induced Diaphragm Atrophy by Transvenous Phrenic Nerve Stimulation , 2017, American journal of respiratory and critical care medicine.

[26]  Robert Oostenveld,et al.  Gain of the human dura in vivo and its effects on invasive brain signal feature detection , 2010, Journal of Neuroscience Methods.

[27]  A. Burkitt,et al.  Electrical probing of cortical excitability in patients with epilepsy , 2011, Epilepsy & Behavior.

[28]  L. Miller,et al.  Optimal spacing of surface electrode arrays for brain–machine interface applications , 2010, Journal of neural engineering.

[29]  Daryl R Kipke,et al.  Complex impedance spectroscopy for monitoring tissue responses to inserted neural implants , 2007, Journal of neural engineering.

[30]  C. Gabriel,et al.  Electrical conductivity of tissue at frequencies below 1 MHz , 2009, Physics in medicine and biology.