Modelling and Calibration Technique of Laser Triangulation Sensors for Integration in Robot Arms and Articulated Arm Coordinate Measuring Machines

A technique for intrinsic and extrinsic calibration of a laser triangulation sensor (LTS) integrated in an articulated arm coordinate measuring machine (AACMM) is presented in this paper. After applying a novel approach to the AACMM kinematic parameter identification problem, by means of a single calibration gauge object, a one-step calibration method to obtain both intrinsic—laser plane, CCD sensor and camera geometry—and extrinsic parameters related to the AACMM main frame has been developed. This allows the integration of LTS and AACMM mathematical models without the need of additional optimization methods after the prior sensor calibration, usually done in a coordinate measuring machine (CMM) before the assembly of the sensor in the arm. The experimental tests results for accuracy and repeatability show the suitable performance of this technique, resulting in a reliable, quick and friendly calibration method for the AACMM final user. The presented method is also valid for sensor integration in robot arms and CMMs.

[1]  Jorge Santolaria,et al.  Kinematic parameter estimation technique for calibration and repeatability improvement of articulated arm coordinate measuring machines , 2008 .

[2]  Xin Li,et al.  Modelling and calibration of the laser beam-scanning triangulation measurement system , 2002, Robotics Auton. Syst..

[3]  Zhang Guoxiong,et al.  Modeling and calibration of a structured-light-sensor-based five-axis scanning system , 2004 .

[4]  Lijiang Wang,et al.  A study of the precision factors of large-scale object surface profile laser scanning measurement , 2002 .

[5]  Guangjun Zhang,et al.  A novel calibration approach to structured light 3D vision inspection , 2002 .

[6]  Xie Zexiao,et al.  A simplified method for the extrinsic calibration of structured-light sensors using a single-ball target , 2004 .

[7]  Jun Ni,et al.  A ball-target-based extrinsic calibration technique for high-accuracy 3-D metrology using off-the-shelf laser-stripe sensors , 2000 .

[8]  Sharon S. Welch Effects of window size and shape on accuracy of subpixel centroid estimation of target images , 1993 .

[9]  Giovanna Sansoni,et al.  State-of-The-Art and Applications of 3D Imaging Sensors in Industry, Cultural Heritage, Medicine, and Criminal Investigation , 2009, Sensors.

[10]  D. Marquardt An Algorithm for Least-Squares Estimation of Nonlinear Parameters , 1963 .

[11]  Hsi-Yung Feng,et al.  Analysis of digitizing errors of a laser scanning system , 2001 .

[12]  Kenneth Levenberg A METHOD FOR THE SOLUTION OF CERTAIN NON – LINEAR PROBLEMS IN LEAST SQUARES , 1944 .

[13]  H. S. Wolff,et al.  iRun: Horizontal and Vertical Shape of a Region-Based Graph Compression , 2022, Sensors.

[14]  J. Denavit,et al.  A kinematic notation for lower pair mechanisms based on matrices , 1955 .

[15]  Avinash C. Kak,et al.  Modeling and calibration of a structured light scanner for 3-D robot vision , 1987, Proceedings. 1987 IEEE International Conference on Robotics and Automation.

[16]  Jorge Santolaria,et al.  A one-step intrinsic and extrinsic calibration method for laser line scanner operation in coordinate measuring machines , 2009 .

[17]  Alexandar Djordjevich,et al.  Study on laser stripe sensor , 1999 .

[18]  Wang Jianguo,et al.  Study on a full field of view laser scanning system , 2007 .