Damping Force and Loading Position Dependence of Mass Sensitivity of Magnetoelastic Biosensors in Viscous Liquid

We established the vibration governing equation for a magnetoelastic (ME) biosensor with target loading in liquid. Based on the equation, a numerical simulation approach was used to determine the effect of the target loading position and viscous damping coefficient on the node ("blind points") and mass sensitivity (Sm) of an ME biosensor under different order resonances. The results indicate that viscous damping force causes the specific nodes shift but does not affect the overall variation trend of Sm as the change of target loading position and the effect on Sm gradually reduces when the target approaches to the node. In addition, Sm decreases with the increase of viscous damping coefficient but the tendency becomes weak at high-order resonance. Moreover, the effect of target loading position on Sm decreases with the increase of viscous damping coefficient. Finally, the results provide certain guidance on improving the mass sensitivity of an ME biosensor in liquid by controlling the target loading position.

[1]  Mikio Muraoka Sensitivity-enhanced atomic force acoustic microscopy with concentrated-mass cantilevers , 2005 .

[2]  Lin Zhang,et al.  Mass Load Distribution Dependence of Mass Sensitivity of Magnetoelastic Sensors under Different Resonance Modes , 2015, Sensors.

[3]  Bryan A. Chin,et al.  Phage coated magnetoelastic micro-biosensors for real-time detection of Bacillus anthracis spores , 2009 .

[4]  Shin Horikawa,et al.  Direct detection of Salmonella typhimurium on fresh produce using phage-based magnetoelastic biosensors. , 2010, Biosensors & bioelectronics.

[5]  Bin Zhou,et al.  A wireless magnetoelastic sensor for uranyl using DNAzyme–graphene oxide and gold nanoparticles-based amplification , 2013 .

[6]  Yating Chai,et al.  Rapid and sensitive detection of Salmonella Typhimurium on eggshells by using wireless biosensors. , 2012, Journal of food protection.

[7]  V. Petrenko,et al.  A pulse system for spectrum analysis of magnetoelastic biosensors , 2010 .

[8]  Craig A. Grimes,et al.  Magnetoelastic sensors for remote query environmental monitoring , 1999 .

[9]  Bryan A. Chin,et al.  Development of highly sensitive handheld device for real-time detection of bacteria in food , 2010, Defense + Commercial Sensing.

[10]  Aoqun Jian,et al.  Detection system based on magnetoelastic sensor for classical swine fever virus. , 2016, Biosensors & bioelectronics.

[11]  J. A. Jendrzejczyk,et al.  Added Mass and Damping of a Vibrating Rod in Confined Viscous Fluids , 1976 .

[12]  Kewei Zhang,et al.  Magnetostrictive particle based biosensors for in situ and real‐time detection of pathogens in water , 2014, Biotechnology and bioengineering.

[13]  Bryan A. Chin,et al.  Evaluation of phage-based magnetoelastic biosensors for direct detection of Salmonella Typhimurium on spinach leaves , 2013 .

[14]  Liling Fu,et al.  Magnetostrictive resonators as sensors and actuators , 2013 .

[15]  Yating Chai,et al.  A pulsed wave excitation system to characterize micron-scale magnetoelastic biosensors , 2014 .

[16]  Craig A. Grimes,et al.  Quantification of Staphylococcus epidermidis using a wireless, mass-responsive sensor , 2010 .

[17]  Yating Chai,et al.  A surface-scanning coil detector for real-time, in-situ detection of bacteria on fresh food surfaces. , 2013, Biosensors & bioelectronics.

[18]  Kewei Zhang,et al.  Study of “blind point” and mass sensitivity of a magnetostrictive biosensor with asymmetric mass loading , 2014 .

[19]  C A Grimes,et al.  A remote query magnetostrictive viscosity sensor. , 2000, Sensors and actuators. A, Physical.

[20]  V A Petrenko,et al.  Sequential detection of Salmonella typhimurium and Bacillus anthracis spores using magnetoelastic biosensors. , 2009, Biosensors & bioelectronics.

[21]  Yong-Sheng Wang,et al.  Streptavidin and gold nanoparticles-based dual signal amplification for sensitive magnetoelastic sensing of mercury using a specific aptamer probe , 2016 .

[22]  Bryan A. Chin,et al.  Detection of methicillin-resistant Staphylococcus aureus using novel lytic phage-based magnetoelastic biosensors , 2015 .

[23]  Suiqiong Li,et al.  Nonuniform mass detection using magnetostrictive biosensors operating under multiple harmonic resonance modes , 2010 .

[24]  Z.-Y. Cheng,et al.  Location Dependence of Mass Sensitivity for Acoustic Wave Devices , 2015, Sensors.

[25]  Qiang Zhang,et al.  A bovine serum albumin-coated magnetoelastic biosensor for the wireless detection of heavy metal ions , 2018 .