Breathing resistance in automated metabolic systems is high in comparison with the Douglas Bag method and previous recommendations

The purpose of this study was to investigate the resistance to breathing in metabolic systems used for the distribution and measurement of pulmonary gas exchange. A mechanical lung simulator was used to standardize selected air flow rates ( V · , L/s). The delta pressure (Δp, Pa) between the ambient air and the air inside the equipment was measured in the breathing valve’s mouthpiece adapter for four metabolic systems and four types of breathing valves. Resistance for the inspiratory and expiratory sides was calculated as RES = (Δp/ V · ) Pa/L/s. The results for resistance showed significant (p < 0.05) between-group variance among the tested metabolic systems, breathing valves, and between most of the completed V · . The lowest resistance among the metabolic systems was found for a Douglas Bag system which had approximately half of the resistance compared to the automated metabolic systems. The automated systems were found to have higher resistance even at low V · in comparison with previous findings and recommendations. For the hardware components, the highest resistance was found for the breathing valves, while the lowest resistance was found for the hoses. The results showed that resistance in metabolic systems can be minimized through conscious choices of system design and hardware components.

[1]  Roy J. Shephard,et al.  Measurement of human energy expenditure, with particular reference to field studies: an historical perspective , 2011, European Journal of Applied Physiology.

[2]  C. Gore,et al.  Validity and reliability of the Cortex MetaMax3B portable metabolic system , 2010, Journal of sports sciences.

[3]  P. Schantz,et al.  Evaluation of the Oxycon Mobile metabolic system against the Douglas bag method , 2010, European Journal of Applied Physiology.

[4]  Roy J. Shephard,et al.  Open-circuit respirometry: a brief historical review of the use of Douglas bags and chemical analyzers , 2017, European Journal of Applied Physiology.

[5]  C J Gore,et al.  Automated VO2max calibrator for open-circuit indirect calorimetry systems. , 1997, Medicine and science in sports and exercise.

[6]  C. J. Gore,et al.  Physiological tests for elite athletes: Australian Sports Commission. , 2000 .

[7]  Frank I. Katch,et al.  Exercise physiology : nutrition, energy and human performance , 2015 .

[8]  A Huszczuk,et al.  A respiratory gas exchange simulator for routine calibration in metabolic studies. , 1990, The European respiratory journal.

[9]  Mats Ainegren,et al.  Skiing Economy and Efficiency in Recreational and Elite Cross-Country Skiers , 2013, Journal of strength and conditioning research.

[10]  Fredrik Andersson,et al.  A Portable Douglas Bag System , 2015 .

[11]  H. Rosdahl,et al.  The Moxus Modular metabolic system evaluated with two sensors for ventilation against the Douglas bag method , 2013, European Journal of Applied Physiology.

[12]  G. J. van Ingen Schenau,et al.  A portable lightweight Douglas bag instrument for use during various types of exercise. , 1983, International journal of sports medicine.

[13]  B. Dietrich Textbook of Work Physiology: Physiological Bases of Exercise , 2004 .

[14]  D. Macfarlane,et al.  Validity, reliability and stability of the portable Cortex Metamax 3B gas analysis system , 2011, European Journal of Applied Physiology.

[15]  C. G. Douglas A method for determinng the total respiratory exchange in man. , 1911 .

[16]  Niels H. Secher,et al.  Influence of body mass on maximal oxygen uptake: effect of sample size , 2001, European Journal of Applied Physiology.

[17]  B. Saltin,et al.  Maximal oxygen uptake in athletes. , 1967, Journal of applied physiology.

[18]  Lars Johansen,et al.  A metabolic cart for measurement of oxygen uptake during human exercise using inspiratory flow rate , 2002, European Journal of Applied Physiology.

[19]  Frank I. Katch,et al.  Exercise Physiology: Energy, Nutrition, and Human Performance , 2006 .

[20]  D A Brodie,et al.  Validity and reliability of selected commercially available metabolic analyzer systems , 2005, Scandinavian journal of medicine & science in sports.

[21]  The validity of the Moxus Modular metabolic system during incremental exercise tests: impacts on detection of small changes in oxygen consumption , 2014, European Journal of Applied Physiology.

[22]  D J Macfarlane,et al.  Automated Metabolic Gas Analysis Systems , 2001, Sports medicine.