Photon count rates estimated from 1980 clinical CT scans

Abstract Background All photon counting detectors have a characteristic count rate over which their performance degrades. Degradation in the clinical setting takes the form of increased noise, reduced material quantification accuracy, and image artifacts. Count rate is a function of patient attenuation, beam filtration, scanner geometry, and X‐ray technique. Purpose To guide protocol and technology development in the photon counting space, knowledge of clinical count rates spanning the complete range of clinical indications and patient sizes is needed. In this paper, we use clinical data to characterize the range of computed tomography (CT) count rates. Methods We retrospectively gathered 1980 patient exams spanning the entire body (head/neck/chest/abdomen/extremity) and sampled 36 951 axial image slices. We assigned the tissue labels air/lung/fat/soft tissue/bone to each voxel for each slice using CT number thresholds. We then modeled four different bowtie filters, 70/80/100/120/140 kV spectra, and a range of mA values. We forward‐projected each slice to obtain detector‐incident count rates, using the geometry of a GE Revolution Apex scanner. Our analysis divided the detector into thirds: the central one‐third, one‐third of the detector split into two equal regions adjacent to the central third, and the final one‐third divided equally between the outer detector edges. We report the 99th percentile of counts to mimic the upper limits of count rates making passing through a patient as a function of patient water equivalent diameter. We also report the percentage of patient scans, by body region, over different count rate thresholds for all combinations of bowtie and beam energy. Results For routine exam types, we recorded count rates of approximately 3.5 × 108 counts/mm2/s in the torso, extremities, and brain. For neck scans, we observed count rates near 6 × 108 counts/mm2/s. Our simulations of 1000 mA, appropriately mimicking the mA needs for fast pediatric, fast thoracic, and cardiac scanning, resulted in count rates of over 10 × 108 counts/mm2/s for the torso, extremities, and brain. At 1000 mA, for the neck region, we observed count rates close to 2 × 109 counts/mm2/s. Importantly, we saw only a small change in maximum count rate needs over patient size, which we attribute to patient mis‐positioning with respect to the bowtie filters. As expected, combinations of kV and bowtie filter with higher beam energies and wider/less attenuating bowtie fluence profiles lead to higher count rates relative to lower energies. The 99th–50th percentile count rate changed the most for the torso region, with a maximum variation of 3.9 × 108 to 1.2 × 107 counts/mm2/s. The head/neck/extremity regions had less than a 50% change in count rate from the 99th to 50th percentiles. Conclusions Our results are the first to use a large patient cohort spanning all body regions to characterize count rates in CT. Our results should be useful in helping researchers understand count rates as a function of body region and mA for various combinations of bowtie filter designs and beam energies. Our results indicate clinical rates >1 × 109 counts/mm2/s, but they do not predict the image quality impact of using a detector with lower characteristic count rates.

[1]  Jiang Hsieh,et al.  Computed Tomography: Principles, Design, Artifacts, and Recent Advances, Fourth Edition , 2022 .

[2]  Kris Iniewski,et al.  Improving Paralysis Compensation in Photon Counting Detectors , 2020, IEEE Transactions on Medical Imaging.

[3]  Shuai Leng,et al.  Photon-counting Detector CT: System Design and Clinical Applications of an Emerging Technology. , 2019, Radiographics : a review publication of the Radiological Society of North America, Inc.

[4]  Amir Pourmorteza,et al.  Photon-counting CT: Technical Principles and Clinical Prospects. , 2018, Radiology.

[5]  P. Pickhardt,et al.  Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCT. , 2017, AJR. American journal of roentgenology.

[6]  Xuejin Liu,et al.  Count rate performance of a silicon-strip detector for photon-counting spectral CT , 2016 .

[7]  Mats Danielsson,et al.  Upper limits of the photon fluence rate on CT detectors: Case study on a commercial scanner. , 2016, Medical physics.

[8]  Ewald Roessl,et al.  A Fourier approach to pulse pile-up in photon-counting x-ray detectors. , 2016, Medical physics.

[9]  Kai Mei,et al.  Reduction of Metal Artifact in Single Photon-Counting Computed Tomography by Spectral-Driven Iterative Reconstruction Technique , 2015, PloS one.

[10]  X Duan,et al.  Initial results from a prototype whole-body photon-counting computed tomography system , 2015, Medical Imaging.

[11]  Rochester,et al.  Use of Water Equivalent Diameter for Calculating Patient Size and Size-Specific Dose Estimates (SSDE) in CT: The Report of AAPM Task Group 220. , 2014, AAPM report.

[12]  N. Pelc,et al.  The piecewise-linear dynamic attenuator reduces the impact of count rate loss with photon-counting detectors , 2014, Physics in medicine and biology.

[13]  K. Taguchi,et al.  Vision 20/20: Single photon counting x-ray detectors in medical imaging. , 2013, Medical physics.

[14]  C. Svensson,et al.  Evaluation of a Second-Generation Ultra-Fast Energy-Resolved ASIC for Photon-Counting Spectral CT , 2013, IEEE Transactions on Nuclear Science.

[15]  Adam S Wang,et al.  Pulse pileup statistics for energy discriminating photon counting x-ray detectors. , 2011, Medical physics.

[16]  Edgar Kraft,et al.  A research prototype system for quantum-counting clinical CT , 2010, Medical Imaging.

[17]  Thomas Toth,et al.  The influence of patient centering on CT dose and image noise. , 2007, Medical physics.

[18]  J H Siewerdsen,et al.  Spektr: a computational tool for x-ray spectral analysis and imaging system optimization. , 2004, Medical physics.

[19]  尚弘 島影 National Institute of Standards and Technologyにおける超伝導研究及び生活 , 2001 .

[20]  J. H. Hubbell,et al.  Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients 1 keV to 20 MeV for Elements Z = 1 to 92 and 48 Additional Substances of Dosimetric Interest , 1995 .