Evaluation of a photon counting Medipix3RX cadmium zinc telluride spectral x-ray detector

Abstract. We assess the performance of a cadmium zinc telluride (CZT)-based Medipix3RX energy-resolving and photon-counting x-ray detector as a candidate for spectral microcomputed tomography (micro-CT) imaging. It features an array of 128  ×  128, 110-μm2 pixels, each with four simultaneous threshold counters that utilize real-time charge summing. Each pixel’s response is assessed by imaging with a range of incident x-ray intensities and detector integration times. Energy-related assessments are made by exposing the detector to the emission from an I-125 radioisotope brachytherapy seed. Long-term stability is assessed by repeating identical exposures over the course of 1 h. The high yield of properly functioning pixels (98.8%), long-term stability (linear regression of whole-chip response over 1 h of acquisitions: y  =    −  0.0038x  +  2284; standard deviation: 3.7 counts), and energy resolution [2.5 keV full-width half-maximum (FWHM) (single pixel), 3.7 keV FWHM (across the full image)] make this device suitable for spectral micro-CT.

[1]  Michael Campbell,et al.  How spectroscopic x-ray imaging benefits from inter-pixel communication. , 2014, Physics in medicine and biology.

[2]  E. Nygard,et al.  Photon counting energy dispersive detector arrays for x-ray imaging , 2007, 2007 IEEE Nuclear Science Symposium Conference Record.

[3]  Thomas Koenig,et al.  Performance of a Medipix3RX Spectroscopic Pixel Detector With a High Resistivity Gallium Arsenide Sensor , 2015, IEEE Transactions on Medical Imaging.

[4]  Sabee Molloi,et al.  Image-based spectral distortion correction for photon-counting x-ray detectors. , 2012, Medical physics.

[5]  Michael Campbell,et al.  Charge Summing in Spectroscopic X-Ray Detectors With High-Z Sensors , 2013, IEEE Transactions on Nuclear Science.

[6]  Erik Fredenberg,et al.  Contrast-enhanced spectral mammography with a photon-counting detector. , 2010, Medical physics.

[7]  Nate D. Tang,et al.  Using algebraic reconstruction in computed tomography , 2012, IVCNZ '12.

[8]  Srinidhi Bheesette,et al.  Per-pixel energy calibration of photon counting detectors , 2017 .

[9]  Anthony P. H. Butler,et al.  Energy Calibration of the Pixels of Spectral X-ray Detectors , 2015, IEEE Transactions on Medical Imaging.

[10]  A. Butler,et al.  Spectral CT data acquisition with Medipix3.1 , 2013 .

[11]  Tyler E. Curtis,et al.  Effects of calibration methods on quantitative material decomposition in photon‐counting spectral computed tomography using a maximum a posteriori estimator , 2017, Medical physics.

[12]  Michael Campbell,et al.  The Medipix3RX: a high resolution, zero dead-time pixel detector readout chip allowing spectroscopic imaging , 2013 .

[13]  K. Taguchi,et al.  Vision 20/20: Single photon counting x-ray detectors in medical imaging. , 2013, Medical physics.

[14]  Srinidhi Bheesette,et al.  Beam profile assessment in spectral CT scanners , 2018, Journal of applied clinical medical physics.

[15]  Giovanni Mettivier,et al.  First images of a digital autoradiography system based on a Medipix2 hybrid silicon pixel detector. , 2003, Physics in medicine and biology.

[16]  Ewald Roessl,et al.  Evaluation of spectral photon counting computed tomography K-edge imaging for determination of gold nanoparticle biodistribution in vivo. , 2017, Nanoscale.

[17]  A. Butler,et al.  Spectroscopic (multi-energy) CT distinguishes iodine and barium contrast material in MICE , 2010, European Radiology.

[18]  Michael Campbell,et al.  Simulations of charge summing and threshold dispersion effects in Medipix3 , 2011 .

[19]  Cynthia H McCollough,et al.  Evaluation of a photon counting Medipix3RX CZT spectral x-ray detector , 2016, Optical Engineering + Applications.

[20]  R. Ballabriga,et al.  The Design and Implementation in $0.13\mu m$ CMOS of an Algorithm Permitting Spectroscopic Imaging with High Spatial Resolution for Hybrid Pixel Detectors , 2009 .

[21]  Ryan K. Roeder,et al.  Nanoparticle imaging probes for molecular imaging with computed tomography and application to cancer imaging , 2017, Medical Imaging.

[22]  Robert Redus Application Note ANCZT-2 Rev . 3 Charge Trapping in XR-100 T-CdTe and-CZT Detectors , 2008 .