A scatter correction method for contrast-enhanced dual-energy digital breast tomosynthesis

Contrast-enhanced dual energy digital breast tomosynthesis (CE-DE-DBT) is designed to image iodinated masses while suppressing breast anatomical background. Scatter is a problem, especially for high energy acquisition, in that it causes severe cupping artifact and iodine quantitation errors. We propose a patient specific scatter correction (SC) algorithm for CE-DE-DBT. The empirical algorithm works by interpolating scatter data outside the breast shadow into an estimate within the breast shadow. The interpolated estimate is further improved by operations that use an easily obtainable (from phantoms) table of scatter-to-primary-ratios (SPR)--a single SPR value for each breast thickness and acquisition angle. We validated our SC algorithm for two breast emulating phantoms by comparing SPR from our SC algorithm to that measured using a beam-passing pinhole array plate. The error in our SC computed SPR, averaged over acquisition angle and image location, was about 5%, with slightly worse errors for thicker phantoms. The SC projection data, reconstructed using OS-SART, showed a large degree of decupping. We also observed that SC removed the dependence of iodine quantitation on phantom thickness. We applied the SC algorithm to a CE-DE-mammographic patient image with a biopsy confirmed tumor at the breast periphery. In the image without SC, the contrast enhanced tumor was masked by the cupping artifact. With our SC, the tumor was easily visible. An interpolation-based SC was proposed by (Siewerdsen et al 2006 Med. Phys. 33 187-97) for cone-beam CT (CBCT), but our algorithm and application differ in several respects. Other relevant SC techniques include Monte-Carlo and convolution-based methods for CBCT, storage of a precomputed library of scatter maps for DBT, and patient acquisition with a beam-passing pinhole array for breast CT. Our SC algorithm can be accomplished in clinically acceptable times, requires no additional imaging hardware or extra patient dose and is easily transportable between sites.

[1]  Serge Muller,et al.  Evaluation of tumor angiogenesis of breast carcinoma using contrast-enhanced digital mammography. , 2006, AJR. American journal of roentgenology.

[2]  Serge Muller,et al.  Dual-energy contrast enhanced digital mammography using a new approach for breast tissue canceling , 2007, SPIE Medical Imaging.

[3]  John M Boone,et al.  Evaluation of scatter effects on image quality for breast tomosynthesis , 2007, SPIE Medical Imaging.

[4]  C. D'Orsi,et al.  Scatter radiation in digital tomosynthesis of the breast. , 2007, Medical physics.

[5]  M. Jiang,et al.  Ordered-subset simultaneous algebraic reconstruction techniques (OS-SART) , 2004 .

[6]  Ann-Katherine Carton,et al.  Quantification for contrast-enhanced digital breast tomosynthesis , 2006, SPIE Medical Imaging.

[7]  C. D'Orsi,et al.  X-ray scatter correction in breast tomosynthesis with a precomputed scatter map library. , 2014, Medical physics.

[8]  D R Dance,et al.  Estimation of scattered radiation in digital breast tomosynthesis , 2014, Physics in medicine and biology.

[9]  Martin J Yaffe,et al.  Contrast-enhanced digital mammography: initial clinical experience. , 2003, Radiology.

[10]  Lei Zhu,et al.  Scatter correction for full-fan volumetric CT using a stationary beam blocker in a single full scan. , 2011, Medical physics.

[11]  Gene Gindi,et al.  Impact of subtraction and reconstruction strategies on dual-energy contrast enhanced breast tomosynthesis with interleaved acquisition , 2013, Medical Imaging.

[12]  K. Ogawa,et al.  A practical method for position-dependent Compton-scatter correction in single photon emission CT. , 1991, IEEE transactions on medical imaging.

[13]  I. Sechopoulos,et al.  A software-based x-ray scatter correction method for breast tomosynthesis. , 2011, Medical physics.

[14]  John M Lewin,et al.  Dual-energy contrast-enhanced digital subtraction mammography: feasibility. , 2003, Radiology.

[15]  Serge Muller,et al.  Dual-energy contrast enhanced digital breast tomosynthesis: concept, method, and evaluation on phantoms , 2007, SPIE Medical Imaging.

[16]  Otto Zhou,et al.  Demonstration of a scatter correction technique in digital breast tomosynthesis , 2013, Medical Imaging.

[17]  Serge Muller,et al.  Development of contrast digital mammography. , 2002, Medical physics.

[18]  Gene Gindi,et al.  Application of the ordered-subsets transmission reconstruction algorithm to contrast-enhanced dual-energy digital breast tomosynthesis , 2014, 2014 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC).

[19]  Wei Zhao,et al.  A simple scatter correction method for dual energy contrast-enhanced digital breast tomosynthesis , 2014, Medical Imaging.

[20]  Gyula Faigel,et al.  X-Ray Holography , 1999 .

[21]  Hilde Bosmans,et al.  Quantification of scattered radiation in projection mammography: four practical methods compared. , 2012, Medical physics.

[22]  Kai Yang,et al.  A breast-specific, negligible-dose scatter correction technique for dedicated cone-beam breast CT: a physics-based approach to improve Hounsfield Unit accuracy , 2014, Physics in medicine and biology.

[23]  Nicolas Duchateau,et al.  Optimization of beam parameters and iodine quantification in dual-energy contrast enhanced digital breast tomosynthesis , 2008, SPIE Medical Imaging.

[24]  S. Velan,et al.  A Practical Method , 1936 .

[25]  Ehsan Samei,et al.  Dual-energy contrast-enhanced breast tomosynthesis: optimization of beam quality for dose and image quality , 2011, Physics in medicine and biology.

[26]  Wei Zhao,et al.  Amorphous selenium flat panel detectors for digital mammography: validation of a NPWE model observer with CDMAM observer performance experiments. , 2006, Medical physics.

[27]  J M Boone,et al.  Scatter/primary in mammography: Monte Carlo validation. , 2000, Medical physics.

[28]  Wei Zhao,et al.  Patient-specific scatter correction for flat-panel detector-based cone-beam CT imaging , 2015, Physics in medicine and biology.

[29]  D Lazos,et al.  Impact of flat panel-imager veiling glare on scatter-estimation accuracy and image quality of a commercial on-board cone-beam CT imaging system. , 2012, Medical physics.

[30]  S. Richard,et al.  A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. , 2005, Medical physics.

[31]  Beverly A. Lau Modeling digital breast tomosynthesis imaging systems for optimization studies , 2012 .

[32]  David A. Scaduto,et al.  Optimization of clinical protocols for contrast enhanced breast imaging , 2013, Medical Imaging.

[33]  Felix Diekmann,et al.  Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study , 2012, Breast Cancer Research.

[34]  Robert M. Nishikawa,et al.  Algorithmic scatter correction in dual-energy digital mammography for calcification imaging , 2012, Medical Imaging.

[35]  J Yorkston,et al.  High-fidelity artifact correction for cone-beam CT imaging of the brain , 2015, Physics in medicine and biology.