Wireless Power Transfer to Miniature Implants: Transmitter Optimization

This paper examines transmitter optimization for wirelessly powering a small implant embedded in tissue. The wireless link between the transmitter and receiver is first modeled as a two-port network and an expression for the power transfer efficiency derived. For a given small receiver in a multilayer tissue model, the transmitter is abstracted as a sheet of magnetic current density for which the optimal distribution is analytically found. The optimal transmitter is compared to the point and uniform source across a range of frequencies. At higher frequencies, the optimal current distribution is shown to induce fields that exhibit focusing. The effects of constructive and destructive interference substantially improves the power transfer efficiency and reinforces operation in the low GHz-range. The optimal transmitter establishes an upper bound on the power transfer efficiency for a given implant and provides insight on the design of the optimal transmit antenna.

[1]  Teresa H. Y. Meng,et al.  A mm-sized implantable power receiver with adaptive link compensation , 2009, 2009 IEEE International Solid-State Circuits Conference - Digest of Technical Papers.

[2]  Erwin S. Hochmair,et al.  System Optimization for Improved Accuracy in Transcutaneous Signal and Power Transmission , 1984, IEEE Transactions on Biomedical Engineering.

[3]  R. W. Lau,et al.  The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues. , 1996, Physics in medicine and biology.

[4]  Shahriar Mirabbasi,et al.  Design and Optimization of Resonance-Based Efficient Wireless Power Delivery Systems for Biomedical Implants , 2011, IEEE Transactions on Biomedical Circuits and Systems.

[5]  W. Chew Waves and Fields in Inhomogeneous Media , 1990 .

[6]  Ieee Standards Board IEEE standard for safety levels with respect to human exposure to radio frequency electromagnetic fields, 3kHz to 300 GHz , 1992 .

[7]  Ada S. Y. Poon,et al.  Optimal transmit dimension for wireless powering of miniature implants , 2011, 2011 IEEE International Symposium on Antennas and Propagation (APSURSI).

[8]  Teresa H. Y. Meng,et al.  A mm-Sized Wirelessly Powered and Remotely Controlled Locomotive Implant , 2012, IEEE Transactions on Biomedical Circuits and Systems.

[9]  Maysam Ghovanloo,et al.  Design and Optimization of Printed Spiral Coils for Efficient Transcutaneous Inductive Power Transmission , 2007, IEEE Transactions on Biomedical Circuits and Systems.

[10]  R. Harrington Time-Harmonic Electromagnetic Fields , 1961 .

[11]  Teresa H. Y. Meng,et al.  A mm-sized wirelessly powered and remotely controlled locomotive implantable device , 2012, 2012 IEEE International Solid-State Circuits Conference.

[12]  N. de N. Donaldson,et al.  Analysis of resonant coupled coils in the design of radio frequency transcutaneous links , 1983, Medical and Biological Engineering and Computing.

[13]  T. Meng,et al.  Optimal Frequency for Wireless Power Transmission Into Dispersive Tissue , 2010, IEEE Transactions on Antennas and Propagation.

[14]  I C Forster,et al.  Theoretical design and implementation of a transcutaneous, multichannel stimulator for neural prosthesis applications. , 1981, Journal of biomedical engineering.