Clinical Study of the Use of the Laser Fluorescence Device DIAGNOdent for Detection of Occlusal Caries in Children

Diagnosis of non-cavitated occlusal caries is generally considered problematic. The aim of this study was to evaluate the utility of the laser fluorescence device DIAGNOdent as part of routine dental check-ups in children. A total of 613 occlusal surfaces on permanent molars and 436 surfaces on primary molars in children aged 7–8 (n = 55) and 13–14 years (n = 54) were examined visually, using DIAGNOdent and for most 13- to 14-year-olds also using radiographs. The teeth were not cleaned professionally before examination. To provide a gold standard for carious teeth, the depth of the caries lesion was determined in those lesions that were judged visually to need opening by drilling (51 teeth). The higher the visual score, the higher were the mean DIAGNOdent values, but in each visual category the variation among values was large. The presence of a clear sealant did not affect DIAGNOdent measurements. The values for permanent and primary molars differed slightly. For permanent teeth, Youden’s index value was highest (60%) at a cut-off value of 30, which indicates that the overall performance for DIAGNOdent in detecting dentinal caries was best at this point. Using this cut-off, sensitivity was 92% and specificity 69% with visual examination for validation, while sensitivity was 92% and specificity 82% with validation by fissure opening. Of the three methods, radiographic examination was the least accurate. In routine dental check-ups of children, DIAGNOdent appears to be useful as an adjunct to visual examination.

[1]  A. Lussi,et al.  Clinical performance of a laser fluorescence device for detection of occlusal caries lesions. , 2001, European journal of oral sciences.

[2]  D. Attrill,et al.  Diagnostics: Occlusal caries detection in primary teeth: a comparison of DIAGNOdent with conventional methods , 2001, British Dental Journal.

[3]  U. Welander,et al.  Occlusal Caries Detection with KaVo DIAGNOdent and Radiography: An in vitro Comparison , 2000, Caries Research.

[4]  A. Lussi,et al.  Performance and Reproducibility of a Laser Fluorescence System for Detection of Occlusal Caries in vitro , 1999, Caries Research.

[5]  W. Youden,et al.  Index for rating diagnostic tests , 1950, Cancer.

[6]  D. Beighton,et al.  Comparison between Visual Examination and a Laser Fluorescence System for in vivo Diagnosis of Occlusal Caries , 2001, Caries Research.

[7]  E. Kidd,et al.  Reproducibility and accuracy of three methods for assessment of demineralization depth of the occlusal surface: an in vitro examination. , 1997, Caries research.

[8]  K Takamori,et al.  Detection of occlusal caries under sealants by use of a laser fluorescence system. , 2001, Journal of clinical laser medicine & surgery.

[9]  E. Kidd,et al.  Detection, Diagnosing, Monitoring and Logical Treatment of Occlusal Caries in Relation to Lesion Activity and Severity: An in vivo Examination with Histological Validation , 1998, Caries Research.

[10]  C. Pinelli,et al.  Validity and Reproducibility of a Laser Fluorescence System for Detecting the Activity of White-Spot Lesions on Free Smooth Surfaces in vivo , 2002, Caries Research.