Error Mechanisms of the Oscillometric Fixed-Ratio Blood Pressure Measurement Method

The oscillometric fixed-ratio method is widely employed for non-invasive measurement of systolic and diastolic pressures (SP and DP) but is heuristic and prone to error. We investigated the accuracy of this method using an established mathematical model of oscillometry. First, to determine which factors materially affect the errors of the method, we applied a thorough parametric sensitivity analysis to the model. Then, to assess the impact of the significant parameters, we examined the errors over a physiologically relevant range of those parameters. The main findings of this model-based error analysis of the fixed-ratio method are that: (1) SP and DP errors drastically increase as the brachial artery stiffens over the zero trans-mural pressure regime; (2) SP and DP become overestimated and underestimated, respectively, as pulse pressure (PP) declines; (3) the impact of PP on SP and DP errors is more obvious as the brachial artery stiffens over the zero trans-mural pressure regime; and (4) SP and DP errors can be as large as 58 mmHg. Our final and main contribution is a comprehensive explanation of the mechanisms for these errors. This study may have important implications when using the fixed-ratio method, particularly in subjects with arterial disease.

[1]  G. Drzewiecki,et al.  Theory of the oscillometric maximum and the systolic and diastolic detection ratios , 2006, Annals of Biomedical Engineering.

[2]  Myriam Oróstegui,et al.  Precisão do aparelho Omron HEM-705 CP na medida de pressão arterial em grandes estudos epidemiológicos , 2011 .

[3]  L A Geddes,et al.  The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure--part ii. , 1980, Journal of biomechanical engineering.

[4]  L. Bautista,et al.  Accuracy of the Omron HEM-705 CP for blood pressure measurement in large epidemiologic studies. , 2011, Arquivos brasileiros de cardiologia.

[5]  Andrew H. Shennan,et al.  Accuracy of inflationary versus deflationary oscillometry in pregnancy and preeclampsia: OMRON-MIT versus OMRON-M7 , 2009, Blood pressure monitoring.

[6]  A K Ream,et al.  Indirect measurement of mean blood pressure in the anesthetized patient. , 1979, Anesthesiology.

[7]  S. Moebus,et al.  Algorithms for converting random-zero to automated oscillometric blood pressure values, and vice versa. , 2006, American journal of epidemiology.

[8]  Jana Kivastik,et al.  Errors of oscillometric blood pressure measurement as predicted by simulation , 2011, Blood pressure monitoring.

[9]  Andrew Shennan,et al.  Validation of the Omron MX3 Plus oscillometric blood pressure monitoring device according to the European Society of Hypertension international protocol , 2005, Blood pressure monitoring.

[10]  Cristina Cristalli,et al.  Influence of arm soft tissue on non-invasive blood pressure measurements: An experimental and mathematical study , 1995 .

[11]  L. Geddes,et al.  Characterization of the oscillometric method for measuring indirect blood pressure , 2006, Annals of Biomedical Engineering.

[12]  K. Nicolaides,et al.  Maternal arterial stiffness in pregnancies affected by preeclampsia. , 2009, American journal of physiology. Heart and circulatory physiology.

[13]  M. Ursino,et al.  A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement , 1996, IEEE Transactions on Biomedical Engineering.

[14]  C. Mittermayer,et al.  Volume elasticity, modulus of elasticity and compliance of normal and arteriosclerotic human aorta. , 1984, Biorheology.

[15]  Paolo Palatini,et al.  Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension protocol , 2003, Blood pressure monitoring.