Demonstration of iodine K-edge imaging by use of an energy-discrimination X-ray computed tomography system with a cadmium telluride detector

An energy-discrimination K-edge X-ray computed tomography (CT) system is useful for increasing the contrast resolution of a target region by utilizing contrast media. The CT system has a cadmium telluride (CdTe) detector, and a projection curve is obtained by linear scanning with use of the CdTe detector in conjunction with an X-stage. An object is rotated by a rotation step angle with use of a turntable between the linear scans. Thus, CT is carried out by repetition of the linear scanning and the rotation of an object. Penetrating X-ray photons from the object are detected by the CdTe detector, and event signals of X-ray photons are produced with use of charge-sensitive and shaping amplifiers. Both the photon energy and the energy width are selected by use of a multi-channel analyzer, and the number of photons is counted by a counter card. For performing energy discrimination, a low-dose-rate X-ray generator for photon counting was developed; the maximum tube voltage and the minimum tube current were 110 kV and 1.0 μA, respectively. In energy-discrimination CT, the tube voltage and the current were 60 kV and 20.0 μA, respectively, and the X-ray intensity was 0.735 μGy/s at 1.0 m from the source and with a tube voltage of 60 kV. Demonstration of enhanced iodine K-edge X-ray CT was carried out by selection of photons with energies just beyond the iodine K-edge energy of 33.2 keV.

[1]  Eiichi Sato,et al.  X-ray Spectra from a Brass-target Plasma Triode , 2007 .

[2]  H. Hosaka,et al.  Small-vessel radiography in situ with monochromatic synchrotron radiation. , 1996, Radiology.

[3]  Toru Aoki,et al.  Application of CdTe photon-counting x-ray imager to material discriminated x-ray CT , 2007, SPIE Optical Engineering + Applications.

[4]  S. G. Buffey,et al.  A new macromolecular crystallography Station (9.5) on the SRS wiggler beam line for very rapid Laue and rapidly tunable monochromatic measurements: Commissioning and first results , 1992 .

[5]  Eiichi Sato,et al.  Enhanced K-edge Angiography Utilizing Tantalum Plasma X-ray Generator in Conjunction with Gadolinium-Based Contrast Media , 2005 .

[6]  Eiichi Sato,et al.  Repetitive flash x‐ray generator having a high‐durability diode driven by a two‐cable‐type line pulser , 1994 .

[7]  Eiichi Sato,et al.  K-edge angiography utilizing a tungsten plasma X-ray generator in conjunction with gadolinium-based contrast media , 2006 .

[8]  Eiichi Sato,et al.  Variations in Cerium X-ray Spectra and Enhanced K-Edge Angiography , 2005 .

[9]  T Takeda,et al.  Development of a two-dimensional imaging system for clinical applications of intravenous coronary angiography using intense synchrotron radiation produced by a multipole wiggler. , 1998, Journal of synchrotron radiation.

[10]  Eiichi Sato,et al.  Quasi-Monochromatic Flash X-Ray Generator Utilizing Disk-Cathode Molybdenum Tube , 2004 .

[11]  Eiichi Sato,et al.  Energy discriminating x-ray camera utilizing a cadmium telluride detector , 2009 .

[12]  Eiichi Sato,et al.  Monochromatic flash x-ray generator utilizing a disk-cathode silver tube , 2005 .

[13]  Eiichi Sato,et al.  Fundamental study on a long-duration flash x-ray generator with a surface-discharge triode , 1994 .

[14]  William Thomlinson,et al.  First operation of the medical research facility at the NSLS for coronary angiography , 1992 .

[15]  Eiichi Sato,et al.  Clean monochromatic x-ray irradiation from weakly ionized linear copper plasma , 2005 .

[16]  Eiichi Sato,et al.  X-ray fluorescence camera for imaging of iodine media in vivo , 2009, Radiological physics and technology.

[17]  R. Germer,et al.  X-ray flash techniques , 1979 .

[18]  Eiichi Sato,et al.  Demonstration of enhanced K-edge angiography using a cerium target x-ray generator. , 2004, Medical physics.

[19]  Eiichi Sato,et al.  Characteristic X-ray Generator Utilizing Angle Dependence of Bremsstrahlung X-ray Distribution , 2006 .

[20]  Frank J. Londy,et al.  Gadolinium-enhanced MR angiography (Gd-MRA) of thoracic vasculature in an animal model using double-dose gadolinium and quiet breathing , 2001, Pediatric Radiology.

[21]  Eiichi Sato,et al.  X-ray Spectra from Weakly Ionized Linear Copper Plasma , 2006 .