Experimental verification of self-calibration radiometer based on spontaneous parametric downconversion

Based on spontaneous parametric downconversion process, we propose a novel self-calibration radiometer scheme which can self-calibrate the degradation of its own response and ultimately monitor the fluctuation of a target radiation. Monitor results were independent of its degradation and not linked to the primary standard detector scale. The principle and feasibility of the proposed scheme were verified by observing bromine–tungsten lamp. A relative standard deviation of 0.39 % was obtained for stable bromine–tungsten lamp. Results show that the proposed scheme is advanced of its principle. The proposed scheme could make a significant breakthrough in the self-calibration issue on the space platform.

[1]  Greg McKee,et al.  Design and characterization of a large area uniform radiance source for calibration of a remote sensing imaging system , 2007, SPIE Optical Engineering + Applications.

[2]  D. Klyshko,et al.  Use of two-photon light for absolute calibration of photoelectric detectors , 1980 .

[3]  Xiaosong Ma,et al.  Quantum teleportation over 143 kilometres using active feed-forward , 2012, Nature.

[4]  Giorgio Brida,et al.  Towards an uncertainty budget in quantum-efficiency measurements with parametric fluorescence , 2000 .

[5]  Nipa Phojanamongkolkij,et al.  Achieving Climate Change Absolute Accuracy in Orbit , 2013 .

[6]  Ginger Drake,et al.  A hyperspectral imager for high radiometric accuracy Earth climate studies , 2011, Optical Engineering + Applications.

[7]  Alan L Migdall,et al.  High accuracy verification of a correlated-photon- based method for determining photoncounting detection efficiency. , 2007, Optics express.

[8]  Rainer Winkler,et al.  Accurate radiometry from space: an essential tool for climate studies , 2011, Philosophical Transactions of the Royal Society A: Mathematical, Physical and Engineering Sciences.

[9]  Bruce A. Wielicki,et al.  Satellite Instrument Calibration for Measuring Global Climate Change: Report of a Workshop , 2004 .

[10]  G. Brida,et al.  Experimental realization of sub-shot-noise quantum imaging , 2010 .

[11]  Sergey V. Polyakov,et al.  Quantum radiometry | NIST , 2009 .

[12]  Ingmar Müller,et al.  Metrology of single-photon sources and detectors: a review , 2014 .