Surface properties of pulsed laser deposited Ir, Rh, and Ir0.9Rh0.1 thin films for use as microelectrode arrays in electroanalytical heavy metal trace sensors

Abstract Pulsed laser deposition (PLD) of iridium, rhodium, and Ir 0.9 Rh 0.1 thin films onto a-SiC:H-coated silicon substrates for use in heavy metal sensors has been achieved by ablating iridium and/or rhodium targets with a KrF excimer laser. The deposited films are polycrystalline and exhibited very smooth surfaces with an average roughness value ( R a ) of ∼1 nm. The wettability by mercury of their surface was investigated by means of the contact angle measurement technique, and compared to that of Ag and Pt films. The PLD films (whether Ir, Rh, or Ir 0.9 Rh 0.1 ) were found to present an almost identical Hg-wetting behavior, which is characterized by a high static contact angle of 132±2°. Moreover, in contrast to the cases of Pt or Ag films, where amalgamation with Hg rapidly occurs during the contact angle measurements, no evident interaction of Hg with the surface of the PLD films was observed. Microelectrode arrays of each of the three films (Ir, Rh and Ir 0.9 Rh 0.1 ) were fabricated and used as a conducting base onto which Hg microdrops are electroplated. Reproducible and uniform Hg deposits on the microelectrode arrays were obtained. The electroanalytical performance of these Hg-electroplated microelectrode arrays based sensors was then evaluated by means of Square Wave Anodic Stripping Voltammetry (SWASV), in synthetic solutions containing Zn, Cd and Pb ion traces. Over a concentration range as wide as 0.2–20 ppb, the detected signals are found to exhibit a strong linear correlation with ion concentration. For a preconcentration time of only 5 min, detection limits as low as 0.2 ppb for both Cd and Pb, and 0.5 ppb for Zn were achieved.

[1]  S. Kounaves,et al.  An indium based mercury ultramicroelectrode: Fabrication and characterization , 1991 .

[2]  M. I. Montenegro,et al.  Microelectrodes : theory and applications , 1991 .

[3]  Gregory T. A. Kovacs,et al.  Microfabricated electrochemical analysis system for heavy metal detection , 1996 .

[4]  N. Parthasarathy,et al.  Reproducible, reliable and rugged Hg-plated Ir-based microelectrode for in situ measurements in natural waters , 1995 .

[5]  Drew Myers,et al.  Surfaces, Interfaces, and Colloids: Principles and Applications , 1991 .

[6]  M. Khakani,et al.  Development of Hg-electroplated-iridium based microelectrode arrays for heavy metal traces analysis , 1999 .

[7]  J. Osteryoung,et al.  Iridium-based small mercury electrodes , 1987 .

[8]  R. Hintsche,et al.  Highly sensitive heavy metal analysis on platinum- and gold-ultramicroelectrode arrays , 1997 .

[9]  W. Stickle,et al.  Handbook of X-Ray Photoelectron Spectroscopy , 1992 .

[10]  R. Wightman,et al.  Voltammetry with Microscopic Electrodes in New Domains , 1988, Science.

[11]  Milena Koudelka-Hep,et al.  Mercury-Plated Iridium-Based Microelectrode Arrays for Trace Metals Detection by Voltammetry: Optimum Conditions and Reliability , 1996 .

[12]  A. Boronin,et al.  An XPS study of the composition of iridium films obtained by MO CVD , 1992 .

[13]  J. Buffle,et al.  An iridium-based mercury-film electrode: Part I. Selection of substrate and preparation , 1987 .

[14]  S. Kounaves,et al.  Determination of Selenium(IV) at a Microfabricated Gold Ultramicroelectrode Array Using Square Wave Anodic Stripping Voltammetry , 1998 .

[15]  R. C. Weast CRC Handbook of Chemistry and Physics , 1973 .

[16]  Gregory T. A. Kovacs,et al.  Iridium-Based Ultramicroelectrode Array Fabricated by Microlithography , 1994 .

[17]  C. Enke,et al.  Investigation of the compound formed at a platinum-mercury interface , 1969 .