Multi-channel whole-head OPM-MEG: Helmet design and a comparison with a conventional system

Magnetoencephalography (MEG) is a powerful technique for functional neuroimaging, offering a non-invasive window on brain electrophysiology. MEG systems have traditionally been based on cryogenic sensors which detect the small extracranial magnetic fields generated by synchronised current in neuronal assemblies, however, such systems have fundamental limitations. In recent years, non-cryogenic quantum-enabled sensors, called optically-pumped magnetometers (OPMs), in combination with novel techniques for accurate background magnetic field control, have promised to lift those restrictions offering an adaptable, motion-robust MEG system, with improved data quality, at reduced cost. However, OPM-MEG remains a nascent technology, and whilst viable systems exist, most employ small numbers of sensors sited above targeted brain regions. Here, building on previous work, we construct a wearable OPM-MEG system with 'whole-head' coverage based upon commercially available OPMs, and test its capabilities to measure alpha, beta and gamma oscillations. We design two methods for OPM mounting; a flexible (EEG-like) cap and rigid (additively-manufactured) helmet. Whilst both designs allow for high quality data to be collected, we argue that the rigid helmet offers a more robust option with significant advantages for reconstruction of field data into 3D images of changes in neuronal current. Using repeat measurements in two participants, we show signal detection for our device to be highly robust. Moreover, via application of source-space modelling, we show that, despite having 5 times fewer sensors, our system exhibits comparable performance to an established cryogenic MEG device. While significant challenges still remain, these developments provide further evidence that OPM-MEG is likely to facilitate a step change for functional neuroimaging.

[1]  Robert Oostenveld,et al.  Localizing human visual gamma-band activity in frequency, time and space , 2006, NeuroImage.

[2]  R. Bowtell,et al.  Using optically-pumped magnetometers to measure magnetoencephalographic signals in the human cerebellum , 2018, bioRxiv.

[3]  D. Cohen Magnetoencephalography: Detection of the Brain's Electrical Activity with a Superconducting Magnetometer , 1972, Science.

[4]  Lauri Parkkonen,et al.  Potential of on‐scalp MEG: Robust detection of human visual gamma‐band responses , 2019, Human brain mapping.

[5]  J. Kitching,et al.  Chip-scale atomic magnetometer , 2004 .

[6]  Matthew J. Brookes,et al.  Towards OPM-MEG in a virtual reality environment , 2019, NeuroImage.

[7]  T. Roberts,et al.  Delayed M50/M100 evoked response component latency in minimally verbal/nonverbal children who have autism spectrum disorder , 2019, Molecular Autism.

[8]  Sylvain Baillet,et al.  Magnetoencephalography for brain electrophysiology and imaging , 2017, Nature Neuroscience.

[9]  Julia M Stephen,et al.  Non-Invasive Functional-Brain-Imaging with an OPM-based Magnetoencephalography System , 2020, PloS one.

[10]  Alexander D Shaw,et al.  Peak visual gamma frequency is modified across the healthy menstrual cycle , 2018, Human brain mapping.

[11]  Matti Stenroos,et al.  Sampling theory for spatial field sensing: Application to electro- and magnetoencephalography , 2019, 1912.05401.

[12]  Lauri Parkkonen,et al.  Optical Co-registration of MRI and On-scalp MEG , 2018, Scientific Reports.

[13]  Robert Oostenveld,et al.  Using a structured-light 3D scanner to improve EEG source modeling with more accurate electrode positions , 2019, Journal of Neuroscience Methods.

[14]  Matti Stenroos,et al.  Measuring MEG closer to the brain: Performance of on-scalp sensor arrays , 2016, NeuroImage.

[15]  Siân E. Robson,et al.  Abnormal visuomotor processing in schizophrenia , 2015, NeuroImage: Clinical.

[16]  Svenja Knappe,et al.  Subpicotesla atomic magnetometry with a microfabricated vapour cell , 2007 .

[17]  Svenja Knappe,et al.  Magnetoencephalography of Epilepsy with a Microfabricated Atomic Magnetrode , 2014, The Journal of Neuroscience.

[18]  M. Romalis,et al.  High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation. , 2002, Physical review letters.

[19]  J. Sarvas Basic mathematical and electromagnetic concepts of the biomagnetic inverse problem. , 1987, Physics in medicine and biology.

[20]  K. D. Singh,et al.  Co-registration of magnetoencephalography with magnetic resonance imaging using bite-bar-based fiducials and surface-matching , 2004, Clinical Neurophysiology.

[21]  V. Shah,et al.  Spin-exchange relaxation-free magnetometry using elliptically polarized light , 2009, 0903.2447.

[22]  Richard Coppola,et al.  Group differences in MEG-ICA derived resting state networks: Application to major depressive disorder , 2015, NeuroImage.

[23]  Matthew J. Brookes,et al.  On the Potential of a New Generation of Magnetometers for MEG: A Beamformer Simulation Study , 2016, PloS one.

[24]  Matthew J. Brookes,et al.  Mapping the topological organisation of beta oscillations in motor cortex using MEG , 2018, NeuroImage.

[25]  Se Robinson,et al.  Functional neuroimaging by Synthetic Aperture Magnetometry (SAM) , 1999 .

[26]  Mark W. Woolrich,et al.  A tool for functional brain imaging with lifespan compliance , 2019, Nature Communications.

[27]  Lauri Parkkonen,et al.  On-scalp MEG system utilizing an actively shielded array of optically-pumped magnetometers , 2018 .

[28]  Joachim Gross,et al.  Good practice for conducting and reporting MEG research , 2013, NeuroImage.

[29]  Matthew J. Brookes,et al.  Cognitive neuroscience using wearable magnetometer arrays: Non-invasive assessment of language function , 2018, NeuroImage.

[30]  Fernando Maestú,et al.  The Role of Magnetoencephalography in the Early Stages of Alzheimer’s Disease , 2018, Front. Neurosci..

[31]  Stephen E. Robinson,et al.  SQUID sensor array configurations for magnetoencephalography applications , 2002 .

[32]  Svenja Knappe,et al.  Chip-scale atomic magnetometer with improved sensitivity by use of the Mx technique , 2007 .

[33]  T. W. Kornack,et al.  A subfemtotesla multichannel atomic magnetometer , 2003, Nature.

[34]  Lauri Parkkonen,et al.  Potential of on-scalp MEG: Robust detection of human visual gamma-band responses , 2019, bioRxiv.

[35]  D. Cohen Magnetoencephalography: Evidence of Magnetic Fields Produced by Alpha-Rhythm Currents , 1968, Science.

[36]  Niall Holmes,et al.  Balanced, bi-planar magnetic field and field gradient coils for field compensation in wearable magnetoencephalography , 2019, Scientific Reports.

[37]  Niall Holmes,et al.  Moving magnetoencephalography towards real-world applications with a wearable system , 2018, Nature.

[38]  Matti Stenroos,et al.  Requirements for Coregistration Accuracy in On-Scalp MEG , 2017, bioRxiv.

[39]  R. Wakai,et al.  A compact, high performance atomic magnetometer for biomedical applications , 2013, Physics in medicine and biology.

[40]  Matthew J. Brookes,et al.  Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalography , 2019, NeuroImage.

[41]  M. Weisend,et al.  Magnetoencephalography with a two color pump-probe fiber-coupled atomic magnetometer. , 2010 .

[42]  Mark W. Woolrich,et al.  Measurement of dynamic task related functional networks using MEG , 2017, NeuroImage.

[43]  Natsuhiko Mizutani,et al.  Human magnetoencephalogram measurements using newly developed compact module of high-sensitivity atomic magnetometer , 2015 .

[44]  M. Sturm,et al.  Minimizing magnetic fields for precision experiments , 2015, 1501.07408.

[45]  R. Bowtell,et al.  Using optically pumped magnetometers to measure magnetoencephalographic signals in the human cerebellum , 2019, The Journal of physiology.

[46]  Mathieu Bourguignon,et al.  Comparing MEG and high-density EEG for intrinsic functional connectivity mapping , 2020, NeuroImage.

[47]  Matthew J. Brookes,et al.  A bi-planar coil system for nulling background magnetic fields in scalp mounted magnetoencephalography , 2018, NeuroImage.

[48]  Svenja Knappe,et al.  Femtotesla atomic magnetometry in a microfabricated vapor cell. , 2010, Optics express.

[49]  Eugen Trinka,et al.  Magnetoencephalography (MEG): Past, current and future perspectives for improved differentiation and treatment of epilepsies , 2017, Seizure.

[50]  Stephen M Smith,et al.  Fast transient networks in spontaneous human brain activity , 2014, eLife.

[51]  Timothy E. J. Behrens,et al.  Human Replay Spontaneously Reorganizes Experience , 2019, Cell.

[52]  Matthew J. Brookes,et al.  Wearable neuroimaging: Combining and contrasting magnetoencephalography and electroencephalography , 2019, NeuroImage.

[53]  D. Hoffman,et al.  Magnetoencephalography with an atomic magnetometer , 2006 .

[54]  Margot J. Taylor,et al.  Low-frequency connectivity is associated with mild traumatic brain injury , 2015, NeuroImage: Clinical.

[55]  Monica Fabiani,et al.  Validation of a method for coregistering scalp recording locations with 3D structural MR images , 2008, Human brain mapping.

[56]  Rebecca J. Theilmann,et al.  Single-subject-based whole-brain MEG slow-wave imaging approach for detecting abnormality in patients with mild traumatic brain injury , 2014, NeuroImage: Clinical.

[57]  Tony W. Wilson,et al.  Circadian modulation of motor-related beta oscillatory responses , 2014, NeuroImage.

[58]  A. C. Maloof,et al.  Ultrahigh sensitivity magnetic field and magnetization measurements with an atomic magnetometer , 2009, 0910.2206.

[59]  Michael P. Weisend,et al.  Multi-sensor magnetoencephalography with atomic magnetometers , 2013, Physics in medicine and biology.

[60]  Matthew J. Brookes,et al.  A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometers , 2017, NeuroImage.

[61]  S E Robinson,et al.  How many channels are needed for MEG? , 2004, Neurology & clinical neurophysiology : NCN.

[62]  T. R. Carter,et al.  A 20-channel magnetoencephalography system based on optically pumped magnetometers. , 2017, Physics in medicine and biology.

[63]  Gareth R. Barnes,et al.  Practical constraints on estimation of source extent with MEG beamformers , 2011, NeuroImage.

[64]  Niall Holmes,et al.  Pragmatic spatial sampling for wearable MEG arrays , 2019, Scientific Reports.

[65]  Matthew J. Brookes,et al.  Optimising experimental design for MEG beamformer imaging , 2008, NeuroImage.

[66]  J. Vrba,et al.  Signal processing in magnetoencephalography. , 2001, Methods.

[67]  L. Trahms,et al.  Magnetoencephalography with a chip-scale atomic magnetometer , 2012, Biomedical optics express.

[68]  Gian Luca Romani,et al.  The impact of improved MEG–MRI co-registration on MEG connectivity analysis , 2019, NeuroImage.

[69]  Lauri Parkkonen,et al.  On-scalp MEG system utilizing an actively shielded array of optically-pumped magnetometers , 2018, NeuroImage.