Robust angle-independent blood velocity estimation based on dual-angle plane wave imaging

Two-dimensional blood velocity estimation has shown potential to solve the angle-dependency of conventional ultrasound flow imaging. Clutter filtering, however, remains a major challenge for large beam-to-flow angles, leading to signal drop-outs and corrupted velocity estimates. This work presents and evaluates a compounding speckle tracking (ST) algorithm to obtain robust angle-independent 2-D blood velocity estimates for all beam-to-flow angles. A dual-angle plane wave imaging setup with full parallel receive beamforming is utilized to achieve high-frame-rate speckle tracking estimates from two scan angles, which may be compounded to obtain velocity estimates of increased robustness. The acquisition also allows direct comparison with vector Doppler (VD) imaging. Absolute velocity bias and root-mean-square (RMS) error of the compounding ST estimations were investigated using simulations of a rotating flow phantom with low velocities ranging from 0 to 20 cm/s. In a challenging region where the estimates were influenced by clutter filtering, the bias and RMS error for the compounding ST estimates were 11% and 2 cm/s, a significant reduction compared with conventional single-angle ST (22% and 4 cm/s) and VD (36% and 6 cm/s). The method was also tested in vivo for vascular and neonatal cardiac imaging. In a carotid artery bifurcation, the obtained blood velocity estimates showed that the compounded ST method was less influenced by clutter filtering than conventional ST and VD methods. In the cardiac case, it was observed that ST velocity estimation is more affected by low signal-to-noise (SNR) than VD. However, with sufficient SNR the in vivo results indicated that a more robust angle-independent blood velocity estimator is obtained using compounded speckle tracking compared with conventional ST and VD methods.

[1]  R. Y. Chiao,et al.  Sparse array imaging with spatially-encoded transmits , 1997, 1997 IEEE Ultrasonics Symposium Proceedings. An International Symposium (Cat. No.97CH36118).

[2]  C. Kasai,et al.  Real-Time Two-Dimensional Blood Flow Imaging Using an Autocorrelation Technique , 1985, IEEE Transactions on Sonics and Ultrasonics.

[3]  Mickael Tanter,et al.  High-contrast ultrafast imaging of the heart , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[4]  J. Jensen,et al.  In-vivo synthetic aperture flow imaging in medical ultrasound , 2003, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[5]  M.E. Aderson,et al.  Multi-dimensional velocity estimation with ultrasound using spatial quadrature , 1998, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[6]  H. Torp,et al.  Simultaneous quantification of flow and tissue velocities based on multi-angle plane wave imaging , 2013, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[7]  M. Crapper,et al.  Flow field visualization of sediment-laden flow using ultrasonic imaging , 2000 .

[8]  D H Evans,et al.  Bias in mean frequency estimation of Doppler signals due to wall clutter filters. , 1995, Ultrasound in medicine & biology.

[9]  M E Anderson,et al.  Speckle tracking for multi-dimensional flow estimation. , 2000, Ultrasonics.

[10]  Billy Y S Yiu,et al.  Vector projectile imaging: time-resolved dynamic visualization of complex flow patterns. , 2014, Ultrasound in medicine & biology.

[11]  Piero Tortoli,et al.  Real-time vector velocity assessment through multigate doppler and plane waves , 2014, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control.

[12]  J Bercoff,et al.  Ultrafast compound doppler imaging: providing full blood flow characterization , 2011, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[13]  Hiroshi Kanai,et al.  Intraventricular blood flow vector and streamline imaging using high frame rate cardiac ultrasound , 2014, 2014 IEEE International Ultrasonics Symposium.

[14]  M. Fink,et al.  Ultrafast compound imaging for 2-D motion vector estimation: application to transient elastography , 2002, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[15]  Che-Chou Shen,et al.  High frame-rate vector flow estimation using speckle tracking with recursive plane-wave compounding , 2010, 2010 IEEE International Ultrasonics Symposium.

[16]  J. Jensen,et al.  A new method for estimation of velocity vectors , 1998, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[17]  K. Boone,et al.  Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study , 1996, Medical and Biological Engineering and Computing.

[18]  Hiroshi Kanai,et al.  Echo speckle imaging of blood particles with high-frame-rate echocardiography , 2014 .

[19]  Ivan Cohen,et al.  Ultrafast compound doppler imaging: A new approach of doppler flow analysis , 2010, 2010 IEEE International Symposium on Biomedical Imaging: From Nano to Macro.

[20]  Damien Garcia,et al.  Doppler vortography: a color Doppler approach to quantification of intraventricular blood flow vortices. , 2014, Ultrasound in medicine & biology.

[21]  Jørgen Arendt Jensen,et al.  Estimation of Velocity Vectors in Synthetic Aperture Ultrasound Imaging , 2006, IEEE Transactions on Medical Imaging.

[22]  K. Kristoffersen,et al.  Clutter filter design for ultrasound color flow imaging , 2002, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[23]  Damien Garcia,et al.  Two-Dimensional Intraventricular Flow Mapping by Digital Processing Conventional Color-Doppler Echocardiography Images , 2010, IEEE Transactions on Medical Imaging.

[24]  Hiroshi Kanai,et al.  High-frame-rate echocardiography using diverging transmit beams and parallel receive beamforming , 2011, Journal of Medical Ultrasonics.

[25]  J. Arendt Paper presented at the 10th Nordic-Baltic Conference on Biomedical Imaging: Field: A Program for Simulating Ultrasound Systems , 1996 .

[26]  J. Jensen,et al.  High frame-rate blood vector velocity imaging using plane waves: Simulations and preliminary experiments , 2008, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.

[27]  Jean Hertzberg,et al.  Development and validation of echo PIV , 2004 .

[28]  R. Daigle,et al.  Estimation and display for Vector Doppler Imaging using planewave transmissions , 2011, 2011 IEEE International Ultrasonics Symposium.

[29]  J. Ophir,et al.  Methods for Estimation of Subsample Time Delays of Digitized Echo Signals , 1995 .

[30]  S. Ohtsuki,et al.  The flow velocity distribution from the doppler information on a plane in three-dimensional flow , 2006, J. Vis..

[31]  Ultrasound Transducers Calculation of Pressure Fields from Arbitrarily Shaped, Apodized, and Excited , 1992 .

[32]  上嶋 徳久,et al.  A new echocardiographic method for identifying vortex flow in the left ventricle : numerical validation , 2009 .

[33]  K. Beach,et al.  Cross-beam vector Doppler ultrasound for angle-independent velocity measurements. , 2000, Ultrasound in medicine & biology.

[34]  M. Fink,et al.  Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography , 2009, IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.