Head-neck disassembly of an uncemented revision stem treated by addition of a proximal spacer.

A 67-year-old woman underwent revision total hip replacement with a long, distally fixed uncemented stem inserted through a transfemoral approach. Initial stability was not achieved and the stem subsided early. Probably due to a conflict with the minor trochanter, the prosthetic head detached from its neck. Instead of a re-revision of the stem, a proximal spacer was inserted to restore stability and to compensate for leg-length discrepancy. This has proved to be a stable and satisfactory solution after three years follow-up.

[1]  H. Chandler,et al.  Revision total hip replacement using the S-ROM femoral component. , 1995, Clinical orthopaedics and related research.

[2]  R. Poss,et al.  The impact of modularity in total hip arthroplasty. , 1995, Clinical orthopaedics and related research.

[3]  E. Salvati,et al.  Complications of femoral and acetabular modularity. , 1995, Clinical orthopaedics and related research.

[4]  M. Mayor,et al.  The tradeoffs associated with modular hip prostheses. , 1995, Clinical orthopaedics and related research.

[5]  Michael Tanzer,et al.  Concerns with modularity in total hip arthroplasty. , 1994, Clinical orthopaedics and related research.

[6]  H. Skinner,et al.  Complications related to modularity of total hip components. , 1993, The Journal of bone and joint surgery. British volume.

[7]  T. Wright,et al.  Dissociation of acetabular components after total hip arthroplasty. Report of four cases. , 1990, The Journal of bone and joint surgery. American volume.

[8]  P. Pellicci,et al.  Disassembly of a modular femoral component during closed reduction of the dislocated femoral component. A case report. , 1990, The Journal of bone and joint surgery. American volume.

[9]  S. Woolson,et al.  Disassembly of a modular femoral prosthesis after dislocation of the femoral component. A case report. , 1990, The Journal of bone and joint surgery. American volume.

[10]  R. Friedman Displacement of an uncemented femoral component after dislocation of a total hip replacement. A case report. , 1989, The Journal of bone and joint surgery. American volume.

[11]  C. Ferenz Polyethylene insert dislocation in a screw-in acetabular cup. A case report. , 1988, Journal of Arthroplasty.