Stability issues of high-energy resolution diode type CdTe nuclear radiation detectors in a long-term operation

Abstract High-energy resolution diode type CdTe detectors were fabricated by growing an n-type epitaxial layer on high resistivity p-like crystal wafers, and their stability issues during a long-term operation were studied. Room temperature stability of the detectors was not good at low operating biases of around 200 V. However, it could be improved significantly by operating them at higher biases under full depletion conditions. On the other hand, detectors exhibited excellent stability by cooling them slightly below room temperature down to 0°C. The effect of this low level of cooling on detector stability was found to be more significant than that of applying high biases at room temperature. By using the detector type presented here, stable operation could be obtained at moderate operating voltages of around 400 V and with a modest degree of cooling.

[1]  M. Martini,et al.  Polarization Phenomena in CdTe Nuclear Radiation Detectors , 1974 .

[2]  J. G. Pasko,et al.  P-I-N CdTe Gamma Ray Detectors by Liquid Phase Epitaxy (LPE) , 1985, IEEE Transactions on Nuclear Science.

[3]  A. Burger,et al.  Defects in CZT crystals and their relationship to gamma-ray detector performance , 2000 .

[4]  T. Schlesinger,et al.  Semiconductors for room temperature nuclear detector applications , 1995 .

[5]  G. Entine,et al.  Time-dependent polarization of CdTe gamma-ray detectors , 1974 .

[6]  Arnold Burger,et al.  Material properties of large-volume cadmium zinc telluride crystals and their relationship to nuclear detector performance , 1997 .

[7]  Elena Aprile,et al.  Evaluation of CdTe detectors with Schottky contacts for imaging applications , 1998 .

[8]  Yasuhiro Tomita,et al.  Performance of a new Schottky CdTe detector for hard x-ray spectroscopy , 1998, Optics + Photonics.

[9]  Kunishiro Mori,et al.  Performance of a new Schottky CdTe detector for hard X-ray spectroscopy , 1997 .

[10]  P. Siffert,et al.  Polarization in Cadmium Telluride Nuclear Radiation Detectors , 1976, IEEE Transactions on Nuclear Science.

[11]  G. Entine,et al.  Advances in CdTe Gamma-Ray Detectors , 1974 .

[12]  Y. Hatanaka,et al.  Fabrication and performance of p–i–n CdTe radiation detectors , 1999 .

[13]  Tadayui Takahashi,et al.  Recent progress in CdTe and CdZnTe detectors , 2001, astro-ph/0107398.

[14]  Bradford H. Parker,et al.  CdZnTe and CdTe detector arrays for hard X-ray and gamma-ray astronomy , 1999 .

[15]  Kazuyuki Satoh,et al.  Growth and characterization of CdTe single crystals for radiation detectors , 1999 .

[16]  M. R. Squillante,et al.  Large, High Resolution CdTe Gamma Ray Sensors , 1986, IEEE Transactions on Nuclear Science.