Exploiting sparsity to accelerate noncontrast MR angiography in the context of parallel imaging

Noncontrast techniques for peripheral MR angiography are receiving renewed interest because of safety concerns about the use of gadolinium in patients with renal insufficiency. One class of techniques involves subtraction of dark‐blood images acquired during fast systolic flow from bright‐blood images obtained during slow diastolic flow. The goal of this work was to determine whether the inherent sparsity of the difference images could be exploited to achieve greater acceleration without loss of image quality in the context of generalized autocalibrating partially parallel acquisition (GRAPPA). It is shown that noise amplification at high acceleration factors can be reduced by performing subtraction on the raw data, before calculation of the GRAPPA weights, rather than on the final magnitude images. Use of the difference data to calculate the GRAPPA weights decreases the geometry factor (g‐factor), because the difference data represent a sparse image set. This demonstrates an inherent property of GRAPPA and does not require the use of compressed sensing. Application of this approach to highly accelerated data from healthy volunteers resulted in similar depiction of large arteries to that obtained with low acceleration and standard reconstruction. However, visualization of very small vessels and arterial branches was compromised. Magn Reson Med, 2012. © 2011 Wiley Periodicals, Inc.

[1]  M J Koelemay,et al.  Magnetic resonance angiography for the evaluation of lower extremity arterial disease: a meta-analysis. , 2001, JAMA.

[2]  P. Boesiger,et al.  SENSE: Sensitivity encoding for fast MRI , 1999, Magnetic resonance in medicine.

[3]  D. Donoho,et al.  Sparse MRI: The application of compressed sensing for rapid MR imaging , 2007, Magnetic resonance in medicine.

[4]  F. Huang,et al.  High‐pass GRAPPA: An image support reduction technique for improved partially parallel imaging , 2008, Magnetic resonance in medicine.

[5]  P. Kellman,et al.  A Sparse TSENSE approach for improved dynamic parallel MRI , 2007 .

[6]  P. Kellman,et al.  Dynamic parallel MRI by generating sparse data : tracking temporal changes , 2006 .

[7]  D. O. Walsh,et al.  Adaptive reconstruction of phased array MR imagery , 2000, Magnetic resonance in medicine.

[8]  Albert P. Chen,et al.  Compressed sensing for resolution enhancement of hyperpolarized 13C flyback 3D-MRSI. , 2008, Journal of magnetic resonance.

[9]  W. Manning,et al.  Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arrays , 1997, Magnetic resonance in medicine.

[10]  M. Miyazaki,et al.  Non‐contrast‐enhanced MR angiography using 3D ECG‐synchronized half‐Fourier fast spin echo , 2000, Journal of magnetic resonance imaging : JMRI.

[11]  Martin Blaimer,et al.  General formulation for quantitative G‐factor calculation in GRAPPA reconstructions , 2009, Magnetic resonance in medicine.

[12]  D. Noll,et al.  Homodyne detection in magnetic resonance imaging. , 1991, IEEE transactions on medical imaging.

[13]  J. Carr,et al.  3D noncontrast MR angiography of the distal lower extremities using flow‐sensitive dephasing (FSD)‐prepared balanced SSFP , 2009, Magnetic resonance in medicine.

[14]  Feng Huang,et al.  k‐t GRAPPA: A k‐space implementation for dynamic MRI with high reduction factor , 2005, Magnetic resonance in medicine.

[15]  V. Runge Gadolinium and nephrogenic systemic fibrosis. , 2009, AJR. American journal of roentgenology.

[16]  Pippa Storey,et al.  Tailoring the flow sensitivity of fast spin‐echo sequences for noncontrast peripheral MR angiography , 2010, Magnetic resonance in medicine.

[17]  John J Sheehan,et al.  Highly accelerated contrast‐enhanced MR angiography: Improved reconstruction accuracy and reduced noise amplification with complex subtraction , 2010, Magnetic resonance in medicine.

[18]  S. T. Nichols,et al.  Quantitative evaluation of several partial fourier reconstruction algorithms used in mri , 1993, Magnetic resonance in medicine.

[19]  Robin M Heidemann,et al.  Generalized autocalibrating partially parallel acquisitions (GRAPPA) , 2002, Magnetic resonance in medicine.