A Wireless Integrated Circuit for 100-Channel Charge-Balanced Neural Stimulation

The authors present the design of an integrated circuit for wireless neural stimulation, along with benchtop and in - vivo experimental results. The chip has the ability to drive 100 individual stimulation electrodes with constant-current pulses of varying amplitude, duration, interphasic delay, and repetition rate. The stimulation is performed by using a biphasic (cathodic and anodic) current source, injecting and retracting charge from the nervous system. Wireless communication and power are delivered over a 2.765-MHz inductive link. Only three off-chip components are needed to operate the stimulator: a 10-nF capacitor to aid in power-supply regulation, a small capacitor (< 100 pF) for tuning the coil to resonance, and a coil for power and command reception. The chip was fabricated in a commercially available 0.6- mum 2P3M BiCMOS process. The chip was able to activate motor fibers to produce muscle twitches via a Utah Slanted Electrode Array implanted in cat sciatic nerve, and to activate sensory fibers to recruit evoked potentials in somatosensory cortex.

[1]  Maurits Ortmanns,et al.  A retina stimulator ASIC with 232 electrodes, custom ESD protection and active charge balancing , 2006, 2006 IEEE International Symposium on Circuits and Systems.

[2]  M. Ortmanns,et al.  A 0.1mm/sup 2/, Digitally Programmable Nerve Stimulation Pad Cell with High-Voltage Capability for a Retinal Implant , 2006, 2006 IEEE International Solid State Circuits Conference - Digest of Technical Papers.

[3]  Maysam Ghovanloo,et al.  Switched-capacitor based implantable low-power wireless microstimulating systems , 2006, 2006 IEEE International Symposium on Circuits and Systems.

[4]  M. Schwarz,et al.  Area saving stimulator cells for multielectrode arrays featuring adaptive waveform generation and monitoring , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[5]  J. Mortimer,et al.  The Effect of Stimulus Parameters on the Recruitment Characteristics of Direct Nerve Stimulation , 1983, IEEE Transactions on Biomedical Engineering.

[6]  Yoon-Kyu Song,et al.  A microelectrode/microelectronic hybrid device for brain implantable neuroprosthesis applications , 2004, IEEE Transactions on Biomedical Engineering.

[7]  J. Thomas Mortimer,et al.  Imbalanced biphasic electrical stimulation: Muscle tissue damage , 2006, Annals of Biomedical Engineering.

[8]  J. Weiland,et al.  A variable range bi-phasic current stimulus driver circuitry for an implantable retinal prosthetic device , 2005, IEEE Journal of Solid-State Circuits.

[9]  L. Wong,et al.  A very low power CMOS mixed-signal IC for implantable pacemaker applications , 2004, 2004 IEEE International Solid-State Circuits Conference (IEEE Cat. No.04CH37519).

[10]  Maurits Ortmanns,et al.  Charge Balancing in Functional Electrical Stimulators: A Comparative Study , 2007, 2007 IEEE International Symposium on Circuits and Systems.

[11]  Andreas Demosthenous,et al.  A Fully Integrated Fail-safe Stimulator Output Stage Dedicated to FES Stimulation , 2007, 2007 IEEE International Symposium on Circuits and Systems.

[12]  M. Ortmanns,et al.  A 232-Channel Epiretinal Stimulator ASIC , 2007, IEEE Journal of Solid-State Circuits.

[13]  Herbert Reichl,et al.  Packaging concepts for neuroprosthetic implants , 2007 .

[14]  K. Wise,et al.  A 64-site multishank CMOS low-profile neural stimulating probe , 1996, IEEE J. Solid State Circuits.

[15]  R. Genov,et al.  256-Channel Neural Recording and Delta Compression Microsystem With 3D Electrodes , 2009, IEEE Journal of Solid-State Circuits.

[16]  Wentai Liu,et al.  A matched biphasic microstimulator for an implantable retinal prosthetic device , 2004, 2004 IEEE International Symposium on Circuits and Systems (IEEE Cat. No.04CH37512).

[17]  G. Cathebras,et al.  An implantable ASIC for neural stimulation , 2004, IEEE International Workshop on Biomedical Circuits and Systems, 2004..

[18]  Sung June Kim,et al.  Design for a Simplified Cochlear Implant System , 2007, IEEE Transactions on Biomedical Engineering.

[19]  Stephen P. DeWeerth,et al.  An Integrated System for Simultaneous, Multichannel Neuronal Stimulation and Recording , 2007, IEEE Transactions on Circuits and Systems I: Regular Papers.

[20]  A. Scheiner,et al.  Imbalanced biphasic electrical stimulation: Muscle tissue damage , 1990, [1990] Proceedings of the Twelfth Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[21]  Reid R. Harrison,et al.  Designing Efficient Inductive Power Links for Implantable Devices , 2007, 2007 IEEE International Symposium on Circuits and Systems.

[22]  G. Geelen,et al.  An inherently linear and compact MOST-only current division technique , 1992 .

[23]  Nigel H. Lovell,et al.  CMOS neurostimulation ASIC with 100 channels, scaleable output, and bidirectional radio-frequency telemetry , 2001, IEEE Transactions on Biomedical Engineering.

[24]  Keith J. Tracey,et al.  A CHARGE-BALANCED CURRENT DRIVER AND ELECTRODE CHARACTERIZATION FOR AN IMPLANTABLE NEURAL STIMULATOR , 2007 .

[25]  R.R. Harrison,et al.  A Low-Power Integrated Circuit for a Wireless 100-Electrode Neural Recording System , 2006, IEEE Journal of Solid-State Circuits.

[26]  Z. Hu,et al.  A 96-channel neural stimulation system for driving AIROF microelectrodes , 2004, The 26th Annual International Conference of the IEEE Engineering in Medicine and Biology Society.

[27]  W. Liu,et al.  A neuro-stimulus chip with telemetry unit for retinal prosthetic device , 2000, IEEE Journal of Solid-State Circuits.

[28]  S.K. Kelly,et al.  A power-efficient voltage-based neural tissue stimulator with energy recovery , 2004, 2004 IEEE International Solid-State Circuits Conference (IEEE Cat. No.04CH37519).

[29]  Karim Abdelhalim,et al.  128-channel fully differential digital neural recording and stimulation interface , 2009, 2009 IEEE International Symposium on Circuits and Systems.

[30]  R. Shepherd,et al.  Electrical stimulation of the auditory nerve: direct current measurement in vivo , 1999, IEEE Transactions on Biomedical Engineering.