Influence of Different Presentations of Oscillometric Data on Automatic Determination of Systolic and Diastolic Pressures

Most non-invasive blood pressure measurements are based on either the auscultatory or the oscillometric technique. In this study, we performed an extensive analysis of the signals, i.e., responses of a microphone implanted in the cuff and pressure changes in the cuff, which can be recorded during such measurements. We applied several methods to separate the cuff deflation from the arterial pressure pulses, as well as to separate the microphone data into an audible part (Korotkoff sounds) and a low frequency part. The oscillometric technique is based on some empirically derived criteria applied to the oscillometric index, which is defined as a certain characteristic physical property of pressure pulses. In addition to the pressure pulses, which are a typical physical property used for the oscillometric index, we also used in this study other properties such as a time derivative and an audible part of data measured by a microphone implanted in the cuff (Korotkoff sounds). We performed a case study of 23 healthy subjects to evaluate the influence of different presentations of the oscillometric index on known height-based and slope-based empirical algorithms for the automatic determination of the systolic and diastolic blood pressures.

[1]  L. Geddes,et al.  Characterization of the oscillometric method for measuring indirect blood pressure , 2006, Annals of Biomedical Engineering.

[2]  Yutaka Imai,et al.  Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults , 2002, Blood pressure monitoring.

[3]  C. Goodman Association for the Advancement of Medical Instrumentation , 1988 .

[4]  M. Ramsey Noninvasive automatic determination of mean arterial pressure , 2006, Medical and Biological Engineering and Computing.

[5]  G. Drzewiecki,et al.  Theory of the oscillometric maximum and the systolic and diastolic detection ratios , 2006, Annals of Biomedical Engineering.

[6]  Ian C. Murray,et al.  Oscillometric Noninvasive Blood Pressure Measurement Devices and Simulators , 2006 .

[7]  Vojko Jazbinsek,et al.  NON-INVASIVE BLOOD PRESSURE MEASUREMENTS: SEPARATION OF THE ARTERIAL PRESSURE OSCILLOMETRIC WAVEFORM FROM THE DEFLATION USING DIGITAL FILTERING , 2005 .

[8]  L. Jackson Digital filters and signal processing , 1985 .

[9]  Vojko Jazbinsek,et al.  AVERAGING OSCILLOMETRIC NON-INVASIVE BLOOD PRESSURE RECORDINGS: TRANSFORMATION INTO THE NORMALISED VIEW , 2005 .

[10]  M. Ursino,et al.  A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement , 1996, IEEE Transactions on Biomedical Engineering.

[11]  Daniel W. Jones,et al.  Recommendations for blood pressure measurement in humans and experimental animals: Part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. , 2005, Hypertension.

[12]  C. Small,et al.  Survey of automated noninvasive blood pressure monitors. , 1994, Journal of clinical engineering.

[13]  A. Beckett,et al.  AKUFO AND IBARAPA. , 1965, Lancet.

[14]  William H. Press,et al.  Numerical recipes in C. The art of scientific computing , 1987 .

[15]  D. Westenskow,et al.  Theoretical analysis of non-invasive oscillometric maximum amplitude algorithm for estimating mean blood pressure , 1997, Medical and Biological Engineering and Computing.

[16]  J M Bland,et al.  Statistical methods for assessing agreement between two methods of clinical measurement , 1986 .

[17]  Alan Murray,et al.  Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements , 2007, Blood pressure monitoring.

[18]  E. O’Brien,et al.  An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. , 1993, Journal of hypertension.