A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

PURPOSE To present a noninvasive technique for directly measuring the CT bow-tie filter attenuation with a linear array x-ray detector. METHODS A scintillator based x-ray detector of 384 pixels, 307 mm active length, and fast data acquisition (model X-Scan 0.8c4-307, Detection Technology, FI-91100 Ii, Finland) was used to simultaneously detect radiation levels across a scan field-of-view. The sampling time was as short as 0.24 ms. To measure the body bow-tie attenuation on a GE Lightspeed Pro 16 CT scanner, the x-ray tube was parked at the 12 o'clock position, and the detector was centered in the scan field at the isocenter height. Two radiation exposures were made with and without the bow-tie in the beam path. Each readout signal was corrected for the detector background offset and signal-level related nonlinear gain, and the ratio of the two exposures gave the bow-tie attenuation. The results were used in the geant4 based simulations of the point doses measured using six thimble chambers placed in a human cadaver with abdomen/pelvis CT scans at 100 or 120 kV, helical pitch at 1.375, constant or variable tube current, and distinct x-ray tube starting angles. RESULTS Absolute attenuation was measured with the body bow-tie scanned at 80-140 kV. For 24 doses measured in six organs of the cadaver, the median or maximum difference between the simulation results and the measurements on the CT scanner was 8.9% or 25.9%, respectively. CONCLUSIONS The described method allows fast and accurate bow-tie filter characterization.

[1]  David G Politte,et al.  Measurement of bow tie profiles in CT scanners using a real-time dosimeter. , 2014, Medical physics.

[2]  Willi A Kalender,et al.  Dose in x-ray computed tomography , 2014, Physics in medicine and biology.

[3]  Bob Liu,et al.  Calculations of two new dose metrics proposed by AAPM Task Group 111 using the measurements with standard CT dosimetry phantoms. , 2013, Medical physics.

[4]  T. Bortfeld,et al.  Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions. , 2000, Physics in medicine and biology.

[5]  David G Politte,et al.  Technical Note: Measurement of bow tie profiles in CT scanners using radiochromic film. , 2015, Medical physics.

[6]  Andrew M Hernandez,et al.  Tungsten anode spectral model using interpolating cubic splines: unfiltered x-ray spectra from 20 kV to 640 kV. , 2014, Medical physics.

[7]  Xun Jia,et al.  Point/counterpoint. GPU technology is the hope for near real-time Monte Carlo dose calculations. , 2015, Medical physics.

[8]  M. Tambasco,et al.  Experimental validation of a kilovoltage x-ray source model for computing imaging dose. , 2014, Medical physics.

[9]  J. Boone,et al.  CT dose index and patient dose: they are not the same thing. , 2011, Radiology.

[10]  M. Kalra,et al.  In vitro dose measurements in a human cadaver with abdomen/pelvis CT scans. , 2014, Medical physics.

[11]  Xinhua Li,et al.  A practical approach to estimate the weighted CT dose index over an infinite integration length , 2011, Physics in medicine and biology.

[12]  John M Boone Method for evaluating bow tie filter angle-dependent attenuation in CT: theory and simulation results. , 2010, Medical physics.

[13]  M. McNitt-Gray,et al.  Development and validation of a measurement-based source model for kilovoltage cone-beam CT Monte Carlo dosimetry simulations. , 2013, Medical physics.

[14]  Cynthia H McCollough,et al.  A method to generate equivalent energy spectra and filtration models based on measurement for multidetector CT Monte Carlo dosimetry simulations. , 2009, Medical physics.

[15]  Mauro Tambasco,et al.  A rapid noninvasive characterization of CT x-ray sources. , 2015, Medical physics.

[16]  J. Boone,et al.  Experimental validation of a method characterizing bow tie filters in CT scanners using a real-time dose probe. , 2011, Medical physics.