Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies.

Tube current modulation governed by x-ray attenuation during CT (computed tomography) acquisition can lead to noise reduction which in turn can be used to achieve patient dose reduction without loss in image quality. The potential of this technique was investigated in simulation studies calculating both noise amplitude levels and noise distribution in CT images. The dependence of noise on the inodulation function, amplitude of modulation, shape and size of the object, and possible phase shift between attenuation and modulation function were examined. Both sinusoidal and attenuation-based control functions were used to modulate tube current. Noise reduction was calculated for both ideal systems and for real systems with limited modulation amplitude. Dose reductions up to 50% can be achieved depending on the phantom geometry and tube current modulation function. Attenuation-based tube current modulation yields substantially higher reduction than fixed-shape modulation functions. Optimal results are obtained when the current is modulated as a function of the square root of attenuation. A modulation amplitude of at least 90% should be available to exploit the potential of these techniques.

[1]  [Dose reduction in computerized tomography with a new scan procedure]. , 1996, Aktuelle Radiologie.

[2]  E Grabbe,et al.  [An anatomically adapted variation of the tube current in CT. Studies on radiation dosage reduction and image quality]. , 1995, RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin.

[3]  I. Kamel,et al.  Radiation dose reduction in CT of the pediatric pelvis. , 1994, Radiology.

[4]  S J Riederer,et al.  Noise Due to Photon Counting Statistics in Computed X‐Ray Tomography , 1977, Journal of computer assisted tomography.

[5]  K S Pentlow,et al.  Radiation dose in CT. , 1992, Radiographics : a review publication of the Radiological Society of North America, Inc.

[6]  B. Bauer,et al.  Erhebungen zur Strahlenexposition der Patienten bei der Röntgendiagnostik , 1995 .

[7]  J. Haaga,et al.  The effect of mAs variation upon computed tomography image quality as evaluated by in vivo and in vitro studies. , 1981, Radiology.